14
9. Connectors and cables
Connector symbols can be found at the bottom of the
control unit. (picture 3)
X1 Connector for an external power supply. Only
use Globtek inc. power supply (see chapter Technical Specifications for details) which can
be obtained as an original accessory from Abilia.
X2 AUX Connector to connect the device to a nurse call system, a personal
emergency phone or an external wireless transmitter. The connector may only be
connected to a system safety voltage input with max. voltage below 25 V(AC) / 60 V (DC),
where both poles have been separated from the electrical network. Max. load current
100 mA.
X3 - Sensor connector. Only use Emfit Bed sensor.
Connect the sensor (picture 13), the power supply delivered as an accessory (picture 14)
and any connector cables (picture 15) according to the pictures.
X2 (AUX) Connector Pin Order (picture 16)
From left to right: