17
12. SW1 press switch (picture 21)
12.1 ON/OFF switch
The SW1 switch acts as the on/off switch if this function is
activated (DIP switch 6 is down).
Press the SW1 switch for three (3) seconds in order to
activate or deactivate the control unit. When the control unit is
switched on, you will hear a beep and the blue LED light will start to flash. When the
device is switched off, you will hear a "beep-beep-boop" (high-high-low) sound. The blue
LED light will go off.
12.2 Acknowledgement switch
The notification sound (if activated) can be muted by pressing the SW1 switch shortly. The
device will make a "beep-beep" sound.
NOTE! If you press the switch for too long, you might accidentally switch off the device.
The notification sound will also stop when the fast-paced movements stop.
12.3 Bed exit alarm bypass switch
The SW1 switch acts as the bypass switch for bed exit alarm if this function is activated
and person wants to exit the bed without causing an alarm. Note! This feature will not work
if any bed exit delay is being used!
Before leaving the bed press SW1 shortly to hear sound “beep beep”. You now have 20
seconds to leave the bed without causing a bed exit alarm. If you do not leave the bed
within 20 seconds the device will re-activate.
12.4 Bed exit alarm “presence / absence” sensitivity switch
SW1 switch can be used to launch automatic calibration for presence / absence sensitivity.
See chapter 16. Calibrating the “presence / absence” sensitivity.