8
You can move backward through the main menu screens by
pressing the d key and a key simultaneously.
In any submenu the same key combination will move you back to
the main menu screen.
Note Changes that you made in the submenu will not be saved.
1.2.5 Status messages and alarm signals
Your Accu-Chek Spirit Combo insulin pump informs you of the
status of your pump using audible beeps, vibrations and
displayed messages. It is vital that you be able to hear and / or
feel the beeps and vibrations. Be aware that some carrying
pouches may hinder the perception of the signals.
You can turn off either beeps or vibrations, but for your safety
you cannot turn off both at the same time for the initial alarm
signal.
See chapter 8 Warnings and errors for a description of the full
range of warnings and errors.
STOP warning
The STOP warning is triggered when your pump switches from
RUN mode to STOP mode, or when a battery is inserted. The
pump beeps twice and vibrates once every minute. This warns
you that insulin delivery has been interrupted.
man_05504929001_H_en-us.indb 8 01.06.2016 08:14:31