Basic use and operation
Setup
The AV500 needs to be placed in the charging cradle before
rst use. AccuVein recommends fully charging the AV500 at
that time. Connect the charging cradle to the power supply.
The AV500 comes with power supply adaptors to ensure outlet
compatibility. If necessary, secure a dierent power adaptor
to the power supply before plugging it into an outlet. The
AV500 is compatible with all voltages from 100
240VAC.
Remove the plastic protective cover from the AV500 display
screen before use.
Intended Use
The AccuVein AV500 is a portable, hand-held instrument
that helps medical professionals to locate certain supercial
veins. The AV500 is intended to be used as a supplement
to appropriate medical training and experience. The AV500
should not be used as the sole method for locating veins,
and should be used only by a qualied medical professional,
who should do so either prior to palpation to help identify
the location of a vein, or afterwards to conrm or refute the
perceived location of a vein. When using the AV500, medical
practitioners should always follow the appropriate medical
protocols and practices as required by their medical facility,
as well as exercise sound medical judgment.
When used properly, the AV500 enables users to locate certain
supercial veins in connection with medical procedures, such
as venipuncture.
The AV500 can be used whenever the determination of vein
location is appropriate such as hospitals and clinics.
Product Description
The AV500 operates by using infrared light to detect veins
beneath the skin, then projecting the position of the veins
on the skin surface directly above the veins. Qualied medi-
cal personnel can observe the vasculature as displayed to
assist them in nding a vein of the right size and position
for venipuncture and other medical procedures requiring
the location of supercial veins. No training is required to
operate the AV500.
The AV500 only shows supercial vasculature. The maximum
depth that veins are displayed varies by patient. In addition,
some patients veins or a portion of their veins might not be
displayed well or at all. Causes for less than optimal or lack
of vein display include, but are not limited to, vein depth,
skin conditions (e.g., eczema, tattoos), hair, scarring or other
highly contoured skin surface, and adipose (fatty) tissue.
When held directly overhead, the AV500 accurately locates
the center of a vein. Increasing the displacement from directly
overhead results in an oset in the displayed vein position.
Width of displayed vein may dier from the actual width
depending on patient to patient dierences and vein width.
The center line of the vein is accurate when the device is
being used correctly and should always be used as the target
when performing venipuncture or other medical procedures.
The AV500 requires no routine or preventative maintenance.
The AV500 is portable, internally powered by battery and
approved for continuous operation. The AV500 is considered
a Type B applied part.
The technology in the AV500 is patented. Visit www.accuvein.com
for the most current information.
5
5

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für AccuVein AV500 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von AccuVein AV500

AccuVein AV500 Bedienungsanleitung - Deutsch - 16 seiten

AccuVein AV500 Bedienungsanleitung - Holländisch - 16 seiten

AccuVein AV500 Bedienungsanleitung - Französisch - 16 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info