Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The AV500 is intended for use in the electromagnetic environment specied below. The customer or the user of the AV500 should assure that it is used in such an environment.
Immunity TestIEC 60601 test levelCompliance levelElectromagnetic environment – guidance
Electrostatic discharge (ESD) IEC
61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete, or ceramic tile. If oors are covered with synthetic
material, the relative humidity should be at least 30 %.
Electrical fast transient/burst IEC
61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
±2kV for power supply
Lines
±1kV for input/output
lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1kV line(s) to line(s)
±2kV line(s) to earth
±1kV line(s) to line(s)
±2kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
<5% UT >95% dip for
0.5 cycle
40% UT 60% dips
70% UT 30% dip
Mains power quality should be that of a typical commercial or hospital environment. If
the user of the AV500 requires continued operation during power mains interruptions, it
is recommended that the AV500 be powered from an uninterruptible power supply or a
battery.
Power frequency (50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m3 A/mPower frequency magnetic elds should be at levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The AV500 is intended for use in the electromagnetic environment specied below. The customer or the user of the AV500 should assure that it is used in such an environment.
Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 VrmsPortable and mobile RF communications equipment should be used no closer to any part of the AV500,
including cables, than the recommended separation distance calculated from the equation applicable to the
frequency of the transmitter. Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m). Field strengths from xed RF
transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each
frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures, objects, and people.
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be
considered. If the measured eld strength in the location in which the AV500 is used exceeds the applicable RF compliance level above the AV500 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the AV500. Over the frequency range 150 kHz to 80 MHz,
Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.
Art des Missbrauchs:
Forenregeln
Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:
Lesen Sie zuerst die Anleitung;
Schauen Sie nach, ob die Frage bereits gestellt wurde;
Stellen Sie die Frage so deutlich wie nur einigermaßen möglich;
Erwähnen Sie was Sie bereits versucht haben um das Problem zu lösen;
Ist Ihr Problem von einem Besucher gelöst dann lassen Sie ihn / sie wissen in diesem Forum;
Falls Sie reagieren möchten, so verwenden Sie bitte das Antworten- Formular;
Da ihre Frage für alle Besucher sichtbar ist, sollten Sie lieber keine persönliche Daten erwähnen.
Neu registrieren
Registrieren auf E - Mails für AccuVein AV500 wenn:
neue Frage gestellt werden
neue Handbücher vorhanden sind
Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.
Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.
Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.
Ihre Frage wurde zu diesem Forum hinzugefügt
Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.