SAFETY INFORMATION
Guidance and manufacturers declaration – electromagnetic immunity
The MD-24 and MD-27 is intended for use in the electromagnetic environment specied below. The
customer or the user of The MD-24 and MD-27 should assure that it is used in such an environment.
Not Life-supporting Medical Equipment.
Immunity test IEC/EN60601
test level
Compliance level Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC/EN61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete
or ceramic tile. If oors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient / burst
IEC/EN61000-4-4
±2kV for power
supply lines
±1kV for input/output
lines
±2kV for power
supply lines
±1kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC/EN61000-4-5
±1kV line(s) to line(s)
±2kV line(s) to earth
±1kV line(s) to line(s)
±2kV line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage uctuations /
icker emissions
IEC/EN61000-3-3
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in
UT) for 5 cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 5 sec.
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in
UT) for 5 cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip
in UT) for 5 sec.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
MD-24 and MD- 27 requires
continued operation during
power mains interruptions, it is
recommended that the MD-24
and MD-27 be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60Hz) magnetic
eld
IEC/EN61000-4-8
3A/m 3A/m
Power frequency magnetic elds
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the MD-24,
MD-27 requires continued operation during power mains interruptions, it is recommended that the MD-24
MD-27 be powered from an uninterruptible power supply or a battery.
ESD declaration statement
There are icker disturbance on the screen during the test, but auto recover after the test. This permissive
loss of performance is specied by the manufacturer, and this phenomena will be put as a clear statement in
the Users Manual to avoid misunderstanding.
DIP declaration statement
The EUT power off during the test, but self- recover after the test. This permissive loss of performance is
specied by the manufacturer, and this phenomena will be put as a clear statement in the User’s Manual to
avoid misunderstanding.
7

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Ag neovo MD-24 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Ag neovo MD-24

Ag neovo MD-24 Bedienungsanleitung - Deutsch - 39 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info