FOREHEAD THERMOMETER
INTRODUCTION
Thank you for purchasing this Fore-
head thermometer. This thermo-
meter is designed with advance
infrared and ambient temperature
compensation technology for accu-
rate and fast temperature measure-
ments. It also has a wide range
mode to scan temperature real time.
ON / OFF
Power on:
Press the “ON/MEN” button.
Power off:
Device will automatically power off
if left idle for more than 1 minute to
extend battery life. In next power on,
it will get back to the default setting
of “HEAD” temperature function
MODE SELECTION
FOREHEAD Mode:
The preset mode
of the thermometer
is forehead. The
thermometer is
ready for use after
the “Head” icon
stops ashing and two beeps are
sounded.
SCAN Mode:
While the unit is
ON, hold dow
n the
“ON/MEN” button,
and press the
“Meas.” button,
then the icon
will replace the “HEAD” icon in the
LCD display.
Fever Alarm:
If the thermometer detects a tempe-
rature >37.5°C (or 99.5°F) under
forehead mode, a long beeping
sound followed by three short beep-
ing sounds to warn the user for
potential fever.
Measuring temperature under SCAN
mode:
When you press the “Meas” button,
you will get the temperature
immediately about 0.6 sec. If you
hold down the scan button, the
reading of measurement will be
continuous updated.
(This mode shows the
actual, unadjusted
temperatures, which is
different with the forehead mode.)
ATTENTION: the BC-12 is NOT
waterproof. The measuring of, for
example, the bath-water must be
done by pointing the thermometer
to the surface of the bath-water and
press the MEAS button.
Memory Function:
Press the “ON/MEN”
button to see the
temperature stored.
This thermometer provides 25 sets
memory for the last measurements.
** Please be noted that only the
“HEAD” function’s reading will be
recorded. (When the thermometer
is turned off, the las
t measurement
data is saved into memory.)
CHANGE °C / °F
To change the LCD from °C to °F:
When the unit is OFF, hold down the
“Meas.” button, then press and hold
the “ON/MEN” button until the LCD
shows ‘°F’, then release the “Meas.”
button. Use the same process to
change the LCD display from °F to
°C.
CLEANING AND STORAGE
The Probe is the most delicate part
of the thermometer. Use with care
when cleaning the probe to avoid
damage.
a. Allow the lens to fully dry for at
least 30 minutes.
b. Keep the unit dry and away from
any liquids and direct sunlight.
c. Please use the alcohol swab or
the soft cotton moistened with the
95% alcohol to clean the probe
lens.
d. Clean the probe and probe lens
after each use to ensure an
accurate reading and avoid cross
contamination.
IMPORTANT NOTES
1. Make sure the Probe is clean and
avoid the smothered with Rubber
boat (Protector).
2. Before the measurement, please
stay in a stable environment for
5 mins and avoid the exercise,
bath for 30mins.
3. Remember to keep the temple
area clean and away from sweat,
cosmetics and scar while taking
temperature.
4. During the measurement, please
away from the direct sunshine and
wind.
5. The BC-12 is NOT waterproof. Do
not plunge into any liquid.
CHANGING THE BATTERY
It is supplied with one lithium cell
(CR2032x1pcs).
1. Open the battery compartment
by turning the cover 45° clockwise
2. Remove the battery safety strip
with a sharp-pointed object.
3. Insert a new battery under the
metal hook and press down until
you hear a click sound
4. Close the battery cover.
ENVIRONMENT
Exhausted batteries may
never be disposed of in
household waste but must
be handed in at your local
depot for chemical waste.
At the end of the product
lifecycle, you should not
throw this product into the
normal household garbage
but bring the product to a
collection point for the recycling of
electrical and electronic equipments.
DISPLAY READOUT
Device can not be powered
on to the ready stage.
Change with a new battery.
Battery is low and no more
measurements are possible.
Replace the battery.
Device stabilization in
process. Wait until ‘HEAD’
stops ashing.
error-1: Measurement before
device stabilization. Wait until
‘HEAD’ stops ashing.
error-2: The device showing a
rapid ambient temperature
change.
error-3: The ambient
temperature is not within the
range between 10°C and
40°C (50°F~104°F).
error-4 ~ error-9:
The system is not
functioning
properly. Unload the battery, wait for 1
minute and repower it. If the message
reappears, contact the retailer for service.
Temperature
taken is higher
than 42.2°C
(108.0°F) for forehead mode or 80°C
(176°F) for scan mode.
Temperature
taken is lower
than 34°C
(93.2°F) for forehead mode or –20°C
(-4°F) for scan mode.
SPECIFICATIONS
Range:
34°C ~ 42,2°C (forehead)
-20°C ~ 80°C (scan)
Accuracy:
0,3°C (forehead)
4% (scan)
Operating Range: 10~40°C
Storage conditions: -20°C ~ 50°C
Battery: CR2032, 3V, lithium
Lifetime battery: measure 40 hours
continuously
DECLARATION OF CONFORMITY
The declaration
of conformity is
available on the website:
GUARANTEE
You have a guarantee of 12 MONTHS after
the date of purchase of this Thermometer. We
guarantee during this period the free repair of
defects caused by material- and construction
faults. This at the importers discretion. It is only
possible to claim to the guarantee when you
enclose your purchase ticket with the BC-12.
HOW TO ACT: If you notice a defect, consult
the manual rst. If this manual gives you no
de nite answer, consult the Alecto Helpline on
telephonenumber +31 (0)73 6411 355.
THE GUARANTEE EXPIRES: When used
inexpert, leaking and faulty placed batteries, use
of unoriginal parts or accessories, negligence
and defects caused by re, ood, stroke of
lightning and natural disasters. With unquali ed
modi cations or repairs by third persons. With
incorrect transport without a suitable packing. If
the device is not handed in with the correct lled in
guarantee card and the purchase ticket.
NOT UNDER THIS GUARANTEE:
batteries.
All further responsibilities, in particular subsequent
damages, are excluded.
USERS MANUAL
HOW TO USE
Measuring the FOREHEAD tempe-
rature:
Always make sure the probe lens is
clean without any damage, and the
forehead is clean.
Pres the “SCAN” button
and gently scan around
the temple area, you can
select the left side of the
left temples or the right side of the
right temples.
Measurement has been completed
after a long beep is heard. (The
time of measurement might be
between 5~8sec (up to 30sec). That
depend on when the thermometer
get the highest temperature of the
forehead.)
Note: Forehead temperature is
displayed in Oral Mode. This mode
converts the forehead temperature
to display its “Oral equivalent.”
Service
Help
+31 (0) 73 6411 355
ver 1.2
display
press
to measure
the
temperature
On/Off
button
Provission
for a strap
BC-12 set to
forehead-mode
BC-12 set to
scan-mode
1

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Alecto BC-12 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Alecto BC-12

Alecto BC-12 Bedienungsanleitung - Holländisch, Französisch - 2 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info