EMC
Guidance and manufacturer's declaration-electromagnetic emissions
T able 1
The Infrared forehead thermometer is intended for use in the electromagnetic environment specified below . The customer or
the user of the Infrared forehead thermometer should assure that it is used in such an environment
The Infrared forehead thermometer uses RF energy only for its internal
function.Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The Infrared forehead thermometer is suitable for use in all
establishments other than domestic and those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic purposes.
Guidance and manufacturer's declaration-electromagnetic emissions
T able 2
The Infrared forehead thermometer is intended for use in the electromagnetic environment specified below . The customer or
the user of the Infrared forehead thermometer should assure that it is used in such an environment
Guidance and manufacturer's declaration-electromagnetic emissions
T able 3
The Infrared forehead thermometer is intended for use in the electromagnetic environment specified below . The customer or
the user of the Infrared forehead thermometer should assure that it is used in such an environment
NOTE UT is the a.c. mians voltage prior to application of the test level.
The Infrared forehead thermometer is intended for use in the electromagnetic environment specified below . The customer or
the user of the Infrared forehead thermometer should assure that it is used in such an environment
Recommended separation distances between portable and mobile RF communications equipment and the
T able 4
Guidance and manufacturer's declaration-electromagnetic emissions
T able 5
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter , where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer .
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent
actual modulation, it would be worst case.
The MANUF ACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT , and
using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum
separation distances for higher IMMUNITY TEST LEVEL .S shall be calculated using the following equation:
The MANUF ACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT , and
using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum
separation distances for higher IMMUNITY TEST LEVEL .S shall be calculated using the following equation:
Where P is the maximum power in W , d is the minimum separation distance in m, and E is the IMMUNITY TEST LEVEL in
V/m.
The Infrared forehead thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Infrared forehead thermometer can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Infrared forehead thermometer as recommended below , according to the maximum output power of the communications
equipment.
Portable and mobile RF communications equipment should be
used no closer to any part of the Infrared Non-contact
Thermometer , including cables, than the recommended
separation distance calculated from the equation appropriate
for the frequency of the transmitter . Recommended separation
distances:
Where, P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer , d is the recommended
separation distance in meters (m) Field strengths
from fixed RF transmitters, as determined by an
electromagnetic site survey , should be less than
the compliance level in each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
80MHz to
800MHz:
800MHz to
2.7GHz:
10V/m, 80% Am
at 1kHz
10V/m, 80% Am
at 1kHz
150KHz to
80MHz:3Vrms
6Vrms (in ISM
and amateur
radio bands)80%
Am at 1kHz
150KHz to
80MHz:3Vrms
6Vrms (in ISM
and amateur
radio bands)80%
Am at 1kHz
Memory-clear Of Measurements
Exceptional Situation
Care And Maintenance
Specification
Clinical Measurement Accuracy And Safety V erification:
If you are sure that you want to remove all stored memo-
ries permanently . Long press the MEM Button for 10
seconds until CLr appears when power off, CLr will flashes
for 5 times to clear all the memories along with 5 short
beeps.
●Keep the unit in the storage case when not in use.
●Clean the unit with a soft dry cloth. Do not use any abrasive or volatile
cleaners.
●Never immerse the unit or any of its' component in water .
Symbol Correction
In Body Mode, Measured temperature is above Measuring range
of 43°C/109.4°F , it is shown on the LCD with red LED.
In Body Mode, Measured temperature is below the Measuring
range of 32°C/89.6°F , it is shown on the LCD with red LED
In Object Mode, Measured temperature is above Measuring range
100.0°C/199.9°F or environmental temperature is above system
Operating range 40°C/104.0°F , it is shown on the LCD with red
LED.
In Object Mode, Measured temperature is below Measuring range
0.0°C/32.0°F , or environmental temperature is be low system
operating range 5°C/41.0°F , it is shown on the LCD with red LED.
Low battery , r eplace all the worn batteries with new ones.
Thermometer system fails or affected by electric magnetic field
Please contact the distributor if you can not solve the problem, do not disassemble the
unit by yourself!
Care for the main unit
Maintenance
Do not clean the unit with naphtha, thinner
or gasoline etc.
Remove the batteries if the unit will notbe
used in 3 months or longer .
Note: We will not be responsible for any quality problem if you do not care and maintain the
product as instructed.
The product has passed clinical trials. The measured results of the infrared
forehead thermometer was compared with the measured re sults of mercury
thermometers, the deviation averageΔt
b
= 0.01 1°C not exceeding 0.3°C; the
clinical repeatability of the infrared forehead thermometer SR =0.100°C, not
exceeding 0.3°C. The measured results up to the laboratory standard and the
clinical standard.Therefore, the deviation average and the clinical repeatability of
the infrared forehead thermometer are complied with the regulatory requirement
ISO 80601-2-56. The conclusions are drawn from the clinical trials, the accuracy
and safety are compliedwith the regulatory requirement.
W arranty Information
▲The unit is guaranteed to be free of defects in workmanship and materials
under normal use for a period of 1 Y ear from the date of purchase.
▲For repair under this warranty . Our authorized service agent must be advised
of the fault wi thin the period of the warranty . This warranty only covers parts and
labor service under normal operations. Any defect resulting
from natural causes, eg. flood, hurricane etc, is not covered in this guarantee.
This guarantee also do es not cover damage incurred by use of the unit notin
accordance with the instructions, accidental damage, or being tampered or
serviced by unauthorized service agents.
▲The fllowing will be excluded from this warranty-if the thermometer has been
misused, abused, or neglect in following the manual's instructions on purpose
and unauthorized repair or modifications.
▲The device requires no calibration.
▲The device is not repairable and contains no user serviceable parts.
Memory-recall of measurements
This Non-contact forehead thermometer automatically stores 20 sets measure-
ments value, the oldest record will be replaced by the la test measurement value
when more than 20 sets.
Read memory record
Press the MEM Button to enter the memory mode when power off. Each time you
press the MEM Button, a number (from 1 to 20) will be displayed along with the
symbol M , 1 second later , the measurement will be shown, as well as subse-
quent measurements can be display one after the other by pressing the MEM
Button each time.
Note:
The reading number "1" is the latest measurement, “20” is the oldest measure-
ment. When the oldest measurement is read, pressing the MEM But ton again will
read the reading "1" again.
Note:
●As the object scan temperature is likely to be affected by the surroundings, the reading shall
be taken as a reference only .
●If the probe is placed at an angle close the object scan, the reading will be affected by
surrounding temperature.
As a Object Thermometer
1). Press the ST ART/STOP Button, all symbols appear on
the display , you can hear 1 short beep, then unit °C/°F on
the display start to flash.
2). Press the SET Button to select Object Mode, the
temperature unit flashes.
3). Move the probe close to the object, make sure the probe
is flat and take measurement with a distance between
1~3cm. Press the ST ART/ST OP Button, The meter will t ake
measurement.
4). The measurement will be finished in 1 second, when it
has been completed, you can hear 1 short beep, in the
meantime, the reading will appears on the display with
green LED blinks.
EC
RE P
Manufacturer
Shenzhen Urion Technology Co.,Ltd.
Floor 4-6th of Building D,Jiale Science&Technology Industrial
Zone,No.3,ChuangWei Road,Heshuikou Community,MaTian Street,
GuangMing New District,518106 ShenZhen,PEOPLE'S REPUBLIC OF CHINA
Tel:(86 )-755-29231308 E-mail:urion@urion.com.cn
MADE IN CHINA
Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg,Germany
Tel:+49-40-251317 E-mail:shholding@hotmail.com
REV:1.1
Manual shutdown
1 short beep when power on and start measurement;
1 long beep when measurement reading is below 37.5°C/99.5°F;
10 short beeps when measurement reading is greater than or equal to
37.5°C/99.5°F;
3 short beeps when system fails
In 60±20s In 8s
2pcs AAA alkaline battery
Approx.75g (batteries not included)
L151mm*W53mm*H41mm
Could be used for 300 times for normal condition
Instruction manual
Body mode 10.0°C~40.0°C/50.0°F~104.0°F
Object mode 5°C ~40.0°C/41.0°F~104.0°F
T emperature: -20°C~+ 50°C/-4.0°F~+122.0°F Humidity: 15% ~95%RH
avoid crash, sun burn or rain during transportation
five years Expected service life
20 sets memory of measurement values
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy . T o
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the Infrared Non-contact Thermometer is used exceeds the applicable
RF compliance level above, the Infrared Non-contact Thermometer should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary , such as re-orienting or relocating the Infrared
Non-contact Thermometer .
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Store the unitin a clean and dry location. not subject the unit
to extreme hot or cold temperature, humidity and direct
sunlight.
Do not use the unit under the influence of electromagnetic
interference (nearby cellphones, microwave etc.)