BENNING CM 1-1/ 1-2
11
against solid impurities of a diameter > 2.5 mm
0 – second index: no protection against water,
- Operating temperature and relative air humidity:
For operating temperatures from 0 °C to 30 °C: relative air humidity less than
80 %,
For operating temperatures from 31 °C to 40 °C: relative air humidity less than
75 %,
For operating temperatures from 41 °C to 50 °C: relative air humidity less than
45 %,
- Storage temperature: The BENNING CM 1-1/ 1-2 can be stored at tempera-
tures between - 20 °C and + 60 °C (air humidity of 0 to 80 %). During stor-
age, the battery should be removed.
7. Electricalspecications
Note: The measuring accuracy is specied as the sum of:
- a relative part of the measured value and
- a number of digits (i.e. counting Stepps of the last digit).
This measuring accuracy applies for temperatures from 18 °C to 28 °C and a
relative air humidity less than 80 %.
7.1 DC voltage ranges ( BENNING CM 1-2)
The input resistance is 10 MΩ.
Measuring
range
Resolution Measuring accuracy
Overload
protection
200 V 0.1 V ± (1.0 % of the measured value + 2 digits) 600 V
eff
600 V 1 V ± (1.0 % of the measured value + 2 digits) 600 V
eff
7.2 AC voltage ranges ( BENNING CM 1-2)
The input resistance is 10 MΩ in parallel 100 pF.
Measuring
range
Resolution
Measuring accuracy *
1
within the frequency
range 50 Hz - 500 Hz
Overload
protection
200 V 0.1 V ± (1.5 % of the measured value + 5 digits) 600 V
eff
600 V 1 V ± (1.5 % of the measured value + 5 digits) 600 V
eff
7.3 AC current ranges
BENNING CM 1-1
Measuring
range
Resolution
Measuring accuracy *
1
within
the frequency range 50 Hz - 60 Hz
Overload
protection
20 A 0.01 A ± (3.0 % of the measured value + 5 digits) 600 A
eff
200 A 0.1 A ± (2.0 % of the measured value + 5 digits) 600 A
eff
400 A 1 A ± (2.0 % of the measured value + 5 digits) 600 A
eff
BENNING CM 1-2
Measuring
range
Resolution
Measuring accuracy *
1
*
2
within
the frequency range 50 Hz - 60 Hz
Overload
protection
40 A 0.1 A ± (1.9 % of the measured value + 10 digits) 600 A
eff
200 A 0,1 A ± (1.9 % of the measured value + 5 digits) 600 A
eff
400 A 1 A ± (1.9 % of the measured value + 5 digits) 600 A
eff
*
1
The measured value is obtained by mean value rectication and is displayed
as r.m.s. value. Its calibration is adapted to a sinusoidal curve.
*
2
The stated accuracy is specied for conductors that are centrally clamped
by means of the current clamp
L
(see gure 4 AC current measurement).
For conductors that are not centrally clamped, an additional error of 1,5 % of
the displayed value has to be considered.
7.4 Resistance measuring range and acoustic continuity test
( BENNING CM 1-2)
Overload protection: 600 V
eff
Measuring
range
Resolution Measuring accuracy
Max. open-
circuit voltage
200 0.1 ± (1,0 % of the measured value + 5 digits) 1.3 V
2 kΩ 1 ± (0.7 % of the measured value + 2 digits) 1.3 V
17

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Benning CM1.1 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info