15
dimensions: length = 210 mm, tube length = 120 mm, tube diameter =
3 mm, V4A (P.no. 044121)
Note on replaceable parts:
- The BENNING MM 7-1 contains fuses for overload protection:
One fuse rated 11 A rapid-acting (1000 V) 20 kA, D = 10 mm, L = 38.1 mm
(P.no. 10016656) and one fuse rated 440 mA rapid-acting (1000 V), D =
10 mm, L = 34.9 mm (P.no. 10016655).
-
The BENNING MM 7-1 is powered by one 9-V block battery (IEC 6 LR 61 “micro”).
- The above mentioned safety cable (tested spare part) are approved in
accordance with CAT III 1000 V/ CAT IV 600 V and for a current up 10 A.
4. Description of unit
See fig. 1: Front panel
The display and operating elements shown in fig. 1 are as follows:
Digital display, for the measurement reading, bar graph and display for
overrange indication,
Polarity display,
Battery display,
RANGE key, switchover between automatic and manual measuring range,
VoltSensor key, for determining AC voltage to earth,
MIN/MAX key, storage of highest and lowest measurement value or peak
value,
Smart HOLD key,
Button (yellow), display illumination,
Function key (blue), for direct voltage/ current (DC) or alternating voltage/
current (AC), resistance or capacity measurement, continuity or diode test-
ing, frequency measurement, temperature measurement in °C or °F,
Rotating switch, for selecting measuring function.
Socket (positive
1
) for V, Ω, Hz, ,
COM socket, joint socket for measurement of current, voltage, resistance,
frequency, temperature, capacity, continuity and diode testing,
Socket (positive), for mA range, for currents up to 600 mA,
Socket (positive), for 10-A range, for currents up to 10 A,
Protective rubber holster
1
) The automatic polarity display for DC current and voltage refers to this.
5. General data
5.1 General data on multimeter
5.1.1 The digital display is designed as a 4 digit liquid-crystal indicator with 14 mm
digit height and decimal point. The highest value displayed is 6000.
5.1.2 The bar graph display consists of 62 segments.
5.1.3 The polarity indication
functions automatically. Only a polarity con-
trary to the socket definition is indicated, as “-”.
5.1.4 The range overload will be displayed with “OL” or “-OL” and sometimes
with an acoustic signal.
Attention: no display or warning by complete overload.
5.1.5 The “RANGE” key
can be used to change over to the manual meas-
uring ranges and to hide “AUTO” on the display at the same time. By
pressing the key for approx. 2 seconds, the automatic range selection is
activated (“AUTO” on the display).
5.1.6 Volt sensor key
: The voltage indicator function is intended for local-
izing AC voltages to earth (see 8.9).
5.1.7 The “MIN/MAX” key function records and stores the highest and low-
est measurements automatically. By pressing the button several times,
the following values are shown: Display “MAX/MIN” shows the current
measurement value, “MAX” shows the highest value stored and “MIN”
the lowest value stored. The “HOLD” key serves to interrupt the “MIN/
MAX” function. When the key is pressed for 2 sec. the unit switches
back to normal mode. Press the “MIN/MAX” key
for 2 seconds to
switch over to the “PEAK” function (peak value storage). The “PEAK”
function detects and stores the positive and negative peak/ crest
value (> 1 ms) in the mV, V AC/ DC and mA, A AC/ DC function. The
automatic range selection is deactivated in the MIN/ MAX and PEAK
function.
5.1.8 “Smart HOLD” - storage of measurement reading: When the “Smart
HOLD” key is pressed, the measurement reading is stored in the
memory. The symbol “HOLD” appears in the display at the same time.
If the measured value increases by 50 digit above the saved value the
change in measured value is shown by a blinking display and by a
signal tone. (changes of measured values between AC and DC voltage/
current will not be recognized).
22

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Benning MM 7.1 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info