If you have any queries, please contact the appropriate
local authorities.
10. Troubleshooting
Problem/
Question
Possible Cause/Remedy
The atomiser
produces
little or no
aerosol.
1. Too much or too little medicine in
the atomiser.
Minimum: 0.5 ml, Maximum: 8 ml.
2. Check that the nozzles are not
blocked If necessary, clean the
nozzle (e.g. by rinsing). Use the
atomiser again.
CAUTION: Only poke implements
through the fine holes from the
underside of the nozzle.
3. The atomiser is not held in an
upright position.
4. The medicine solution is unsuit-
able for nebulising (e.g. too thick).
The medicine solution should be
specified by the doctor.
Output is too
low.
Hose kinked, filter blocked, too
much inhalation solution.
Which
medicines are
suitable for
inhalation?
Consult your doctor.
Essentially all medicines that are
suitable and approved for device-
based inhalation can be inhaled.
Some inhala-
tion solution
remains in the
atomiser.
This is normal and occurs for techni-
cal reasons. Stop inhalation if you
hear a notable difference in the noise
of the atomiser.
What special
steps should
be taken for
babies and
children?
1. In babies, the mask should cover
the mouth and nose to guarantee
effective inhalation.
2. In children, the mask should also
cover the mouth and nose. Nebu-
lisation next to a sleeping person
is not suitable because insufficient
medicine reaches the lungs.
Note: Children should only use the
device with help and under supervi-
sion of an adult. Never leave a child
alone with the nebuliser.
Inhalation with
the mask takes
longer.
This is for technical reasons. Less
medicine is inhaled per breath
through the mask holes than using
the mouthpiece. The aerosol is
mixed with ambient air through the
holes in the mask.
Problem/
Question
Possible Cause/Remedy
Why do I need
to replace the
atomiser at
regular inter-
vals?
There are two reasons for this:
1. To guarantee a therapeutically
effective particle spectrum, the
nozzle holes should not exceed a
certain diameter. Due to the ther-
mal and mechanical stresses, the
plastic is subjected to a certain
amount of wear.
The atomiser [9] is
particularly sensitive. This can also
cause changes to the droplet com-
position of the aerosol, which has a
direct effect on the effectiveness of
the treatment.
2. You are also recommended to
replace the atomiser on a regular
basis for hygiene reasons.
Does each per-
son need their
own atomiser?
This is absolutely necessary for
hygiene reasons.
11. Technical specifications
Dimensions
(WxHxD) 300 x 180 x 100 mm
Weight 1.9 kg
Operating
pressure approx. 1.2 bar
Atomiser fill
volume max. 8 ml
Medicine flow > 0.2 ml/min.
Sound pressure < 60 dBA
(acc. to DIN EN 13544-1 section 26)
Mains
connection 230 V ~, 50 Hz; 180 VA
Housing material ABS
Operating
conditions
Temperature: +10 °C – +40 °C
Relative humidity: 10% – 95%
Storage and
transport
conditions
Temperature: -20 °C – +70 °C
Relative humidity: 10% – 95%
Atmospheric pressure: 500 – 1060 hPa
Subject to technical modifications.
Particle size diagram
All measurements were obtained with a sodium chlo-
ride solution using a laser diffraction method.
Cumulative distribution [–]
Particle size [μm]
17

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Beurer IH 21 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Beurer IH 21

Beurer IH 21 Bedienungsanleitung - Holländisch - 12 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info