If the thermometer shows «RET» (without the battery symbol) on its display, the
anti-theft time has elapsed and the thermometer must be returned to the station
for unlocking.
If the anti-theft function is not enabled, the display will show «off» when in stand-
by mode.
Care and cleaning
Braun ThermoScan probe covers are intended for single use only!
To ensure accurate readings, always use Braun ThermoScan disposable probe
covers with the PRO 4000 thermometer. Damaged, perforated, or soiled probe
covers can lead to inaccurate temperature readings. Additionally, the accuracy of
your temperature readings can be affected by damage to the probe window, or
the presence of dirt or cerumen on the probe window.
To assure a high level of accuracy, it is very important to keep this window clean,
dry and undamaged. Fingerprints, cerumen, dust and other soiling compounds
reduce transparency of the window and result in lower temperature measure-
ments. If the thermometer is accidentally placed in the ear without a probe cover
installed, the window must be cleaned immediately. Additionally, the window
and/or probe tip area should never be modied, changed, or adjusted. These
changes will affect calibration and accuracy of the thermometer.
To clean the window, gently wipe its surface with a cotton swab slightly moist-
ened with isopropyl alcohol and immediately wipe dry with a clean cotton swab.
After cleaning, allow at least 5 minutes drying time before taking temperatures.
Make sure probe window is clean and dry. Avoid touching the window except
when cleaning is required. If the window is damaged, the thermometer must be
returned to Welch Allyn for service.
Note: Do not use any chemical other than isopropyl alcohol to clean the probe
window.
Thermometer
Use a soft cloth slightly moistened with isopropyl alcohol to clean the thermome-
ter display and exterior. As needed, clean the thermometer and probe with a 70%
isopropyl alcohol solution or a 10% chlorine bleach solution or a nonstaining
disinfectant such as CaviCide
®
or its equivalent. Do not use abrasive cleaners.
Never submerge the thermometer in water or any other liquid. Store thermometer
and probe covers in a dry location (the thermometer is not protected against in-
gress of water), free from dust and contamination and away from direct sunlight.
The ambient temperature at the storage location should remain fairly constant
and within the range of 50 °F to 104 °F (10 °C to 40 °C).
Trouble-shooting
Situation Solution
Error message
No probe cover is attached. Attach new, clean probe cover.
(POS = position error)
The infrared monitor cannot nd a
temperature equilibrium and allows
no measurement.
Clear the display by changing
the probe cover.
Ensure that the positioning of the
probe is correct and remains stable.
(ashing)
Ambient temperature is not within the
allowed operating range 50–104 °F
(10–40 °C) or changing too rapidly.
Allow the thermometer to remain for 30
minutes in a room where the tempera-
ture is between 50 and 104 °F / 10 and
40 °C.
Temperature taken is not within typical
human temperature range.
HI will be displayed when temperature
is higher than 108 °F (42.2 °C)
LO will be displayed when temperature
is lower than 68 °F (20 °C)
Make sure a new, clean probe cover is
attached and thermometer is properly
inserted. Then, take a new temperature
System error
If error persists,
If error still persists,
Wait 1 minute until the thermometer
turns off automatically, then turn on
again.
... reset the thermometer by removing
the batteries and putting them back
in.
... contact Welch Allyn.
10 11
6

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln
1

Forum

braun-pro-4000

Suche zurücksetzen

  • Temperaturmessergebnis, muß etwas vom Messergebnis abgezogen werden?
    Eingereicht am 25-12-2018 14:55

    Antworten Frage melden

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Braun PRO 4000 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Braun PRO 4000

Braun PRO 4000 Bedienungsanleitung - Holländisch, Dänisch, Polnisch, Schwedisch, Norwegisch, Finnisch - 69 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info