54/66 Device handbook CAM mobile, 06/2014
C3 Specifications
Measuring ranges 30A / 300A / 3000A AC
Output sensitivity 100mV / 10mV / 1mV per A (AC coupled)
Accuracy (45-65Hz) ± 1% of reading ± 0.1A 30/300A
± 1% of reading ± 1A 3000A
Minimum load 100k for specified accuracy
Frequency range 10Hz to 20kHz (-1dB)
Phase <± 1° (45-65 Hz)
Position sensitivity ± 2% of reading
External field ± 0.2% of range with cable >200mm (8") from the probe
Noise 8mV eff. in 30A range
2mV eff. in 300A / 3000A ranges
Temperature coefficient ± 0.1% of reading /°C
Power supply Provided by CAM mobile
Overload Indicated by a red LED
Enclosure
Material ARNITE T06-200SNF, UL94 V0
Degree of Protection IP40
Dimensions 110 (l) x 65 (w) x 23 (d) mm
Probe
Probe length 610mm (24”), double insulated
Probe diameter 9.9 mm (0,39”)
Output cable 2m (78.7“),probe to integrator
Material Alcryn 2070NC, LATI LATENE 7H2W V0
Degree of Protection IP65
General Characteristics
Operating temp. -20°C up to +80°C
Storage temp. -40°C up to 90°C
Operating humidity 15% up to 85% (non-condensing)
Safety standards EN 61010-1:2001
EN 61010-031:2002
EN 61010-2-032:2002
EMC EN 61326-2-2 :2006
1000 V
RMS
, Category III, 600 V
RMS
, Category IV, Pollution Degree 2 (Probe and integrator)
30V maximum between output and earth (ACP 3005)
ROHS and WEEE compliant
Rated for continuous use