Periodically check the condition of the filter. Correct
filter maintenance prolongs the life of the aerosol.
The frequency with which filter should be replaced
depends on the conditions in which the unit is used.
However, it is recommended that filters be replaced
after 20 working hours or in the event they are vis-
ibly dirty. The filter is located on the lower part of the
aerosol unit. To replace the filter:
Open the filter cover cap, using a small pointed ob-
ject if necessary
· Remove the filter to be replaced from its housing;
use a sharp-pointed object for assistance if necessary;
· Carefully insert the new filter;
· Return the filter cover cap to its original position
TECHNICAL SPECIFICATIONS
Nominal voltage: 230 V
Frequency: 50 Hz
Power: 130 VA
Fuse: T 1.6 AL 250V ~
Maximum flow: 15 ± 2 l/min.
Operational flow: 7 ± 1 l/min.
Max pressure: 2 +- 0.4 atm
Operating pressure: 0,6 ± 0,1 atm
Protection against liquids: IPX0
Weight: 1210 g
Device conditions of use: 20’ ON - 40’ OFF
MMAD: 3,2 ± 0,2 µm
Breathable fraction: 72,2 % ± 2%
Phial capacity: 5 ml
Noise level at 50cm*: 63 dB(A)
nebulisation speed: 0,25 ± 0,04 ml/min.
residual volume not nebulised: 0,8 ± 0,05 g
% < 5µm: 72,2% ± 2%
% between 2 and 5 microns 51,4%
*The data relative to noise were measured with a new
unit. Values may vary with use.
If the product is used and maintained in accordance
with the instructions for use, it should have a life
expectancy of at least 1000 operating cycles. In the
event of breakage or damage, dispose of according to
instructions.
The information on the unit’s capabilities provided
by the manufacturer in accordance with EN 13544-
1 may not apply to suspended or viscous medicines.
EN 60601-1 Electrical-medical devices -Part 1: gen-
eral safety standards
EN 60601-1-2 Medical electrical equipment - Part
1-2: General requirements for basic safety and essen-
tial performance - Collateral standard: Electromag-
netic compatibility - Requirements and tests
EN 13544-1 Respiratory therapy units - Part 1: nebuli-
sation systems and related components
SYMBOLS KEY
WARNING!
Consult the Instructions for Use
Type BF applied part
Double insulation
Compliant with MDD Directive 93/42 EEC
+ 2007/47/EC
Alternate current
Storage and transport conditions
Humidity: 15%/85%. Store in a cool, dry
place
Temperature: -10°C/+55°C
Environmental Conditions of Use:
Humidity: 15%/85%
Store in a cool, dry place
Temperature: 10°C/40°C
The medical device requires special precautions re-
garding EMC and must be installed and used accord-
ing to the information provided in the accompanying
documents.
WARRANTY
Chicco is covered by a guarantee should faulty ma-
terials or manufacturing defects be discovered. The
guarantee period is 2 years. Any faulty parts shall
be replaced free of cost only if the unit has not been
tampered with and all directions of use attached to
the unit have been strictly followed. Any tampering
with unit/repairs carried out by personnel not author-
ised by Artsana SpA result in the immediate voiding
of all conditions of guarantee. In case of faulty or
unsatisfactory operation, please contact the Artsana
S.p.A dealer or distributor.
AEROSOLS
AEROSOLS
AEROSOLS
AEROSOLS
0068
12

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Chicco Aerosol met Zuiger wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info