Instruction Manual
Wrist T ype
Blood Pressure Monitor
Model BPM280XL
This manual and product are not meant to be a substitute for advice provided by
your doctor or other medical professional. You are not to use the information
contained herein, or this product for diagnosing or treating a health problem or
prescribing any medication. If you have or suspect that you have a medical
problem, promptly consult your healthcare provider.
The device is designed for non-invasive measurements and monitoring of adults’
arterial blood pressure. It measures human's systolic blood pressure, diastolic blood
pressure and heart rate on wrist by oscillometric measurement method.
The device is recommended for use by people over the age of 18 and is for home
use and not clinical use.
Blood Pressure Monitor Intended Use
*Attention!
To operate the device properly and get the best result from it, please read the
Instruction Manual thoroughly before you begin measuring and keep it carefully
so that you can access and consult it at any time.
Medical Disclaimer
1-1. What is blood pressure ?
1. About Blood Pressure
Having one's blood pressure measured by a doctor in a hospital or a clinic, and a
group health checks, tends to stimulate nervousness in the subject and may even
create high blood pressure, blood pressure also varies in accordance with a variety of
conditions and so judgment is not possible on the basis of a single measurement.
The blood pressure measured first thing in the morning after getting up, before taking
any food and with the subject still, is known as the fundamental blood pressure. In
practice it is rather difficult to record the fundamental blood pressure, but to come as
near as possible to measuring the blood pressure in an environment that is close to
this is why it is useful to take the measurement at home.
Blood pressure is a measurement of the force of blood flowing against the walls of
the arteries. Arterial blood pressure is constantly changing during the course of the
cardiac cycle. The highest pressure in the cycle is called the systolic blood pressure
and the lowest is the diastolic blood pressure. Both of pressure readings, the systolic
and the diastolic, are necessary to enable a physician to evaluate the status of a patient's
blood pressure. Many factors such as physical activity, anxiety or the time of day can
influence your blood pressure. Blood pressure is typically low in the morning and increases
from the afternoon to the evening. It is lower in the summer and higher in the winter.
1-2. Why is it useful to measure blood pressure at home?
A. WHO blood pressure classifi cations
Standards for assessment of high or low
blood pressure without regard to age, have
been established by the World Health
Organization (WHO), as shown in the chart.
Optimal
Normal
High normal
Mild hypertension
Moderate hypertension
Severe hypertension
120 130
85
80
95
100
90
105
110
140 150 160 170 180
* Figure No.1
B. Variations in blood pressure
* Figure No.2
Individual blood pressures vary greatly
both on a daily and a seasonal basis.
These variations are even more pronounced in hypertense patients.
(Hypertense: means a person who has high blood pressure symptom.)
Normally the blood pressure rises while at work
and is at its lowest during sleeping period. The
graph below illustrated the variations in blood
pressure over a whole day with measurement
taken every five minutes. (The thick line
represents sleep. The rise in blood pressure at 4
PM <A in the graph> and 12 PM <B in the
graph> correspond to an attack of pain.)
● Please rest for at least 5 ~ 10 minutes before taking the measurement.
● To allow your blood vessels to return to the condition prior to taking the measurement, please wait at least 3 ~ 5 minutes
between measurements. You may need to adjust the wait time according to your personal physiological situation.
● If you have one of the circulatory problems as arteriosclerosis, diabetes, liver disease, kidney disease, severe
hypertension, peripheral circulation......, please consult your doctor or healthcare professional before using the devices.
● Blood pressure measurements should be interpreted by a physician or trained health professional who is familiar with
one's medical history. By using the unit regularly and recording the results for the physician to interpret, one can keep
physician informed of the continuing trends in one's blood pressure.
● Wind the cuff snugly around your wrist and the cuff must be at the same level as your heart.
● Do not move the unit during measurement, or the proper measurement will not be achieved.
● Take measurement in a relaxed position without talking.
● Do not wind the cuff over jacket or sweater sleeve, or measurement cannot be done.
● Keep in mind that blood pressure naturally varies from time to time throughout the day and also is affected by lots of
different factors such as smoking, alcohol consumption, medication and physical activity.
● People with a condition that causes circulatory problems (diabetes, kidney disease, arteriosclerosis or poor peripheral
circulation) may get lower readings with this monitor than with a blood pressure monitor that is used on the upper arm.
Please consult your physician to determine if your wrist blood pressure accurately reflects your actual blood pressure.
● Blood pressure measurements taken with the unit are equivalent to those obtained by a trained observer using the
cuff / stethoscope auscultation method and are within the accuracy limits prescribed by the American National Standard
for Electronic or Automated Sphygmomanometers.
2. Important information before use of the unit
● Wait 30 ~ 45 minutes before measurement if you have just consumed caffeinated beverages or smoked cigarettes.
● We recommend you to use the same wrist (preferably the left wrist) and to measure around the same time each day.
4-2. Irregular Heartbeat Detector (IHB)
*NoteΜ
● Any irregular heartbeat detected will not be stored in the memory.
The appearance of this symbol, , signifies that a certain pulse irregularity was detected
during the measurement. Talking, moving, shaking or an irregular pulse during the measurement
can result in the appearance of this icon. Usually this is not a cause for concern, however if the
symbol appears often, we recommend you seek medical advice. Please note that the device will
not replace a true cardiac examination, but serves to detect pulse irregularities at an early stage.
As a safeguard, we recommend that if you have arrhythmias such as atrial or ventricular premature
beats and atrial fibrillation or any other special conditions you should check with your physician
before using your BPM.
The hearbeat rhythm that is more than or less than 25% from the average heart beat intervals is
usually defined as an irregular heartbeat rhythm.In order to filter the unstable status of user and
avoid affecting the detection of heart rate from any movement, shaking or talking in the begining of
measurement, the method of averaging heart beat intervals of subject device is calculated with the
three proper heart beat pulses detected in the begining of measurement and that is different from
a strict mathematical averaging of all recorded intervals.At least 3 beats with at least 25% difference
from the average heart beat interval will generate the IHB symbol on the LCD.
*Attention!
● Batteries are hazardous waste. Do not dispose them together with the household garbage.
Please take the used batteries to the recycling collection point according to your local regulations.
● Keep the battery away from small children.
● To prolong the battery life and prevent damage caused by leakage, remove the batteries from the
device if the device is not to be used for a long period.
● Memories (if any) will not be deleted during battery replacement.
● After replacing the batteries, reset the date and time.
LCD DISPLAY
START/STOP
BUTTON
RECALL MEMORY
BUTTON
WRIST CUFF
CARRYING CASE
ALKALINE
BATTERIES
UM-4 (AAA) 1.5V x 2
* Figure No.3
3-1. Name of the parts
3. Device Description
UP & DOWN BUTTON
FUNCTION
BUTTON
3-2. Description of display symbols
* Figure No.4
4. Main feature:
4-1. Personalized Target Limit Function
The device allows you to pre-set your personalized values of blood pressure which should be
recommended or decided by your physician or medical professional. If the Systolic or Diastolic
pressure of the measurement is greater than your pre-set personalized level, the digit of the result
on the screen will flash to alert. If the Systolic or Diastolic pressure of the measurement is equal to or
less than your pre-set personalized levels, or if you did not set the personalized levels, the SMILE
FACE symbol will appear and flash on screen after each measurement.
4-3. 3D Positioning Technology
Keeping wrist at heart level while measuring is extremely important to get accurate blood pressure
values. The device features a Wrist Position Sensor which guides you to adjust your hand by the
WRIST DIRECTION INDICATOR symbols shown on screen and helps to determine the optimal
height of the wrist for measurement.
HOUR: MINUTE
SYSTOLIC PRESSURE
IRREGULAR HEART BEAT SYMBOL
LOW BATTERY SYMBOL
DIASTOLIC PRESSURE
HEART RATE
PULSE SYMBOL BLOOD PRESSURE UNIT
SELF-DIAGNOSTIC
SMILE FACE
MONTH/ DATE
ERROR SYMBOL
MEMORY SET
WRIST DIRECTION
INDICATOR
*Note!
● The 3D positioning function is optional. If it is switched off, the device will not show the symbols to
indicate the optimal position of the wrist while measuring.
● User can press “START/STOP” button to exit setting mode and go back to standby mode at any step.
*Note!
● When taking repeated measurements, make sure to select the same memory
zone number before pressing “START/STOP” button, so the measurement is
registered to the appropriate memory.
● Do not move or talk during measurement.
*Note!
● To check the pre-set personalized levels, press and hold "MODE" button.
*Note!
● If there is no data in the selected memory zone, nothing (except month, date, time
and memory zone) will appear on the display.
● If there is one data in the selected memory zone, the measurement data will
appear on the display.
● If there are 2 data in the selected memory zone, the average of the 2 data will
appear on the display.
D. 3D positioning function during measurementǺ
a. While measuring, slowly move your wrist
according to the Arrow symbols on screen
to adjust the wrist angle.
b. When the wrist is in the right position, all Arrow symbols will disappear and PULSE
symbol will flash twice on screen. Then the measurement will automatically begin.
A. Press "▲" or "▼" button to select a particular memory zone (1, 2 or 3).
B. Press "START/STOP" button to start measurement. The monitor will automatically
inflate to the level that is right for you.
C. After inflation of the cuff, the pressure will slowly decrease and when a pulse is
detected, the SYMBOL will start flashing.
D. When the measurement is finished, the systolic pressure, diastolic pressure, heart rate
will be displayed for 1 minute.
E. If the Systolic or Diastolic pressure of the result is greater than your pre-set
personalized level, the digit of Systolic or Diastolic pressure will flash. If the Systolic or
Diastolic pressure of the result is equal to or less than your pre-set personalized level,
the SMILE FACE symbol will flash.
F. The device will switch to standby mode without any operation after 1 minute.
A. Press "MODE" button, the screen displays “3d on” (the device is default with the
position sensor function on).
B. Press "▲" or "▼" button to switch on/off.
C. Press "MODE" button to go back to standby mode.
8-6. Storing, recalling, and erasing measurement data:
8-5. Steps to take blood pressure measurement:
A. Storing data:
After each measurement, the systolic and diastolic pressure, heart rate with the time and
date will be automatically stored. The monitor features a 3 zone memory capability. Each
memory zone holds the last 40 measurements, replacing the oldest data with new one.
B. Recalling data:
C. Erasing data:
a.Press
"▲" and "▼"
button to select the memory zone.
b. Press
"
MEMORY
"
button to enter MEMORY MODE. The average of last 3 measurement
readings will first appear on the display.
c. Press
"
MEMORY
"
button. The latest measurement data will appear with the number as well
as the recording date and time. Data are recalled from the latest measurement to the oldest.
d. Press
"
MEMORY
"
button to read the next measurement data.
e. To stop reading the memories, press the
"
START/STOP
"
button.
a. Press "▲" or "▼" button to select the memory zone (1, 2 or 3), then press "MEMORY" button.
b. Press and hold "MODE" and "
▲
" buttons simultaneously, all stored data,
including the average reading, in the selected memory zone will be erased.
c. To confirm the data has been erased, press "MEMORY" button, no data
(except month, date, time and memory zone number) should appear.
9. Maintenance
9-1. Clean the device body, case and the cuff carefully with a clean, soft and slightly dampened cloth and
then wipe them dry with another piece of dry cloth.
9-2. Do not use thinner, alcohol, petrol (gasoline) or any strong cleansers to clean.
9-3. Do not wash the cuff or make it wet, and never use chemical cleaner on it.
9-4. Do not in any way twist the cuff.
9-5. Make sure no water will enter the main unit.
9-6. When the unit is not to be used for a long time, remove the batteries. (Leaking of battery liquid can
cause damage.)
9-7. Do not attempt to disassemble or change any parts of the monitor, including cuff.
9-8. If any suggestion or service is requested please consult your service station.
9-9. Do not place the item directly under the sunlight, in high temperature, or humid and dusty places.
9-10. Do not store in extremely low (less than - 20 °C/ -4 °F) or high (more than 70 °C/ 158 °F) temperature.
Display mark Corrective action
Battery symbol
Measurement in progress.
Do not move or talk.
Replace all two batteries with new ones.
Insert the batteries in correct polarities.
Pulse symbol
Flashes on the LCD when pulse is
detected during measurement.
Condition/Cause
Appears when the battery voltage is
excessively low.
Measuring
error symbol
Press "START/STOP" button again and re-measure.
Or check cuff if wrapped properly around the wrist
according to instructions. Check palm if exerting effort.
Check if there is moving during measurement. Check if
posture is correct.
Appears when the accurate blood pressure
could not be obtained.
IHB symbol
Repeat the measurement.
Relax for at least 5 minutes and sit comfortably
and quietly before you restart a measurement.
Appears for 1 minute when the user was
talking, moving or shaking or an irregular
heart beat was detected during measurement.
1 0. Display Messages/T r oubleshooting
Model Number
Method of Measurement
Range of Measurement
Accuracy
Inflation
Exhaust
Display
Dimension
Unit Weight
Sets of Memory
Cuff Size
Operating Conditions
Storage Conditions
Power Supply
Battery Life
Accessories
BPM280XL
Oscillometric
Pressure 0 ~ 300 mmHg,
Pulse 40 ~ 199 beats/minute
Pressure +/- 3 mmHg, Pulse +/- 5%
Automatic inflation (Air Pump)
Automatic exhaust valve
Liquid Crystal Display
6.7 x 3.4 x 8.5cm (L x W x H)
Approx. 148 g (including battery)
3 databases of 40 memory each = total 120 sets
Wrist circumference approx. 135 ~ 195 mm
+10 °C to + 40 °C, less than 85% R.H.
- 20 °C to + 70 °C, less than 85% R.H.
AAA (1.5 V) Alkaline battery x 2
Approx. 250 times measurement
Carrying case, instruction manual, 2pcs”AAA” (1.5 V) battery
These specications are subject to change without notice for purpose of improvement.
11. Specifi cations
0197
This blood pressure monitor complies with the EC Directive and bears the CE
mark “CE0197”. This blood pressure monitor also complies with mainly
following standards,(included but not limited)
Safety standard:
EN 60601-1 Medical electrical equipment part 1: General requirements for safety
EMC standard:
EN 60601-1-2 Medical electrical equipment part 1-2: General requirements for safety- Collateral
standard: Electromagnetic compatibility- Requirements and tests
Performance standards:
EN 1060-1 Non-invasive sphygmomanometers - General requirements
EN 1060-3 Non-invasive sphygmomanometers - Supplementary requirements for
electromechanical blood pressure measuring systems.
EN 1060-4 Non-invasive sphygmomanometers - Test procedures to determine the overall
system accuracy of automated non-invasive sphygmomanometers.
Note
● To ensure precise measurement reading,
recalibration of the device is recommended
after 2 years from the date of purchase.
Shipping plus handling cost and recalibration
service fee shall be charged accordingly.
Important /Note! Read the operating
instructions.
Consult instructions thoroughly before
use.
Classification:
- Internally powered equipment
- BF type applied part
- IPX0
- Not suitable for use in presence of
flammable anaesthetic mixture with
air or with Oxygen or nitrous oxide
- Continuous operation with short-time
loading
To avoid inaccurate results caused by
electromagnetic interference between
electrical and electronic equipments, do
not use the device near a cell phone or
microwave oven.
Discard the used product to the recycling
collection point according to local
regulations.
Authorized Representative in the
European Community
EMERGO EUROPE
Molenstraat 15, 2513 BH,
The Hague, The Netherlands
TEL : +31-70-3458570
FAX : +31-70-3467299
Manufacturer:
HEALTH & LIFE Co., Ltd.
9F, No. 186, Jian Yi Road,
Chung Ho City 235, Taipei, Taiwan
Distributor:
Sabre Nederland b.v.
Ellermanstraat 23
1099 BX Amsterdam the Netherlands
MODE
START
ST
OP
MEMORY
* Figure No.6
6. Attaching Pressure Cuff
6-1. Do not place the pressure cuff over a jacket or sweater
sleeve. Wrap the pressure cuff around the bare wrist with
the monitor facing you.
6-2. Wrap the cuff snugly. Do not make it too tight.
Fold the remaining part of the cuff back out of the way.
6-3. Leave approximately 10 mm (Approx. 0.4 inch) between
the cuff and the bottom of your hand palm.
*Note!
● Do not use this device if your wrist has any wound or injury.
● Do not wrap the cuff around any body part other than your wrist.
0.4” (10 mm)
135 ~ 195 mm
8-3. Activating Personalized Target Limit Function and Setting Personalized Levels:
8-4. Activating 3D Positioning Technology Function:
*Note!
● Systolic pressure range for setting personalized levels is between 70 and 250 mmHg.
● Diastolic pressure range for setting personalized levels is between 40 and 115 mmHg.
● The default value for Systolic pressure is 135 mmHg and 85 mmHg for Diastolic pressure.
A. Press "MODE" button (" - - " will flash for 1st systolic pre-set value).
Press "▲" or "▼" button to set/adjust the personalized systolic pressure.
B. Press "MODE" button (" - - " will flash for 1st diastolic pre-set value).
Press "▲" or "▼" button to set/adjust the personalized diastolic pressure.
C. Press "MODE" button (" - - " will flash for 2nd systolic pre-set value).
Press "▲" or "▼" button to set/adjust the personalized systolic pressure.
D. Press "MODE" button (" - - " will flash for 2nd diastolic pre-set value).
Press "▲" or "▼" button to set/adjust the personalized diastolic pressure.
E. Press "MODE" button (" - - " will flash for 3rd systolic pre-set value).
Press "▲" or "▼" button to set/adjust the personalized systolic pressure.
F. Press "MODE" button (" - - " will flash for 3rd diastolic pre-set value).
Press "▲" or "▼" button to set/adjust the personalized diastolic pressure.
G. After setting the personalized levels, the device will return to Standby Mode and is
ready for a measurement by pressing START/STOP button.
* Figure No.7
7-1. To get accurate measurements, make sure the cuff is at the same height as the heart.
Having the cuff higher or lower may cause inaccurate results.
7-2. Measurement while sitting down:
A. Place your elbow on a table or other object (such as the carrying case).
B. Use the armrest to position the arm so that the pressure cuff is at the same
height as the heart.
C. Relax your hand with the palm facing up.
7. Correct Measurement Position
8-1. Inserting batteries in position:
8. How to Measure
A. Open battery cover.
B. Refer to the picture in positioning batteries. (Figure No.5)
C. All segments will appear on the display in 3 seconds.
8-2. Setting Date and Time:
A. Press "MODE" button ("month" starts flashing). Press "▲" or "▼" button to set the correct month.
B. Press "MODE" button again ("date" starts flashing). Press "▲" or "▼" button to set the correct date.
C. Press "MODE" button again ("hour" starts flashing). Press "▲" or "▼" button to set the correct hour
in 12-hour format.
D. Press "MODE" button again ("minute" starts flashing). Press "▲" or "▼" button to set the correct
minute.
E. Press "MODE" button again to switch to normal time.
When the low battery symbol appears on the display, please change the batteries.
Replace all the batteries with new ones and do not mix new and old batteries. Do not mix alkaline,
standard (carbon-zinc) or rechargeable (cadmium)
batteries either. It may shorten the battery life or cause
the device to malfunction. Remove the battery cover at
the bottom of the unit and insert 2 pieces AAA alkaline
batteries into the battery compartment. Make sure the
polarities "+" and "–" ends are properly positioned.
5. Inserting/ Changing Batteries
* Figure No.5
Move wrist upward
Move wrist right
Move wrist left
Move wrist downward