2 x LR03, 1.5 V, AAA
e
izkissen
23210 09/2012
Read the instruction manual carefully before
using this device, especially the safety
instructions, and keep the instruction manual
for future use. Should you give this device to
another person, it is vital that you also pass on
these instructions for use.
GB
Safety instructions
• The unit is intended only for use in the home. Consult your doctor prior to
using the blood pressure monitor if you are concerned about health matters.
Only use the unit for its intended purpose in accordance with the instructions
for use.
Warranty claims become void if the unit is misused.
• Please consult your doctor prior to using the unit if you suffer from illnesses
such as arterial occlusive disease.
The unit may not be used to check the heart rate of a pacemaker.
Pregnant women should take the necessary precautions and pay attention to
their individual circumstances. Please consult your doctor if necessary.
If you experience discomfort during a measurement, such as pain in the
wrist or other complaints, press the button
to release the air
immediately from the cuff. Loosen the cuff and remove it from your wrist.
The unit is unsuitable for children.
Do not allow children to use the unit. Medical products are not toys!
Please ensure that the unit is kept away from the reach of children.
The swallowing of small part like packaging bag, battery, battery cover and so
on may cause the suffocation.
The user must check that the equipment functions safely and see that it is in
proper working condition before being used.
• The device may not be operated in rooms with high levels of radiation or in
the vicinity of high-radiation devices, such as radio transmitters, mobile
telephones or microwaves, as this could cause functional defects or incorrect
measurements.
Do not use the equipment where flammable gas (such as anaesthetic gas,
oxygen or hydrogen) or flammable liquid (such as alcohol) are present.
No modification of this unit/device is allowed.
Please do not attempt to repair the unit yourself in the event of malfunctions,
since this will void all warranty claims. Only have repairs carried out by
authorised service centres.
Protect the unit against moisture. If liquid does manage to penetrate into the
unit, remove the batteries immediately and do not continue to use the unit. In
this case contact your specialist dealer or notify us directly.
Please don’t use a dilution agent, alcohol or petrol to clean the unit.
Please don’t hit heavily or fall down the product from a high place.
If the unit is not going to be used for a long period, please remove the
batteries.
SAFETY NOTES FOR BATTERIES
Do not disassemble batteries!
Replace the new batteries if the unit display a low battery symbol.
• Never leave any low battery in the battery compartment since they may leak
and cause damage to the unit.
Increased risk of leakage, avoid contact with skin, eyes and mucous
membranes! If battery acid comes in contact with any of these parts, rinse
the affected area with copious amounts of fresh water and seek medical
attention immediately!
• If a battery has been swallowed seek medical attention immediately!
• Replace all of the batteries simultaneously!
Only replace with batteries of the same type, never use different types of
batteries together or used batteries with new ones!
• Insert the batteries correctly, observing the polarity!
• Please take off the battery if you won’t use in 3 months.
Remove the batteries from the device if it is not going to be used for an
extended period!
• Keep batteries out of children's reach!
Do not attempt to recharge these batteries!
There is a danger of explosion!
• Do not short circuit! There is a danger of explosion!
• Do not throw into a fire! There is a danger of explosion!
Do not throw used batteries into the household refuse; put them in a
hazardous waste container or take them to a battery collection point, at the
shop where they were purchased!
Unit and LCD display
button (Memory Recall)
Blood Pressure Indicator (green - yellow - orange - red)
LCD screen (display)
START/STOP button
Lid of the Battery Compartment
Wrist cuff
Display of Date/Time
Display of Systolic Pressure
Display of Diastolic Pressure
Pulse symbol
Display of Pulse Rate
Arrhythmia display (arrhythmia cordis)
Change Battery symbol
Memory Location number
Memory-Symbol
Blood Pressure Indicator
User memory 1 / 2
Scope of supply
Please check first of all that the unit is complete.
The following parts are included as standard:
• 1 ecomed blood pressure monitor BW-80E
2 batteries (type AAA, LR03) 1.5 V
• 1
storage box
1 instruction manual
Please contact your supplier if you notice any transport damage on unpacking
the unit.
What is blood pressure?
Blood pressure is the pressure created in the blood vessels by each heart beat.
When the heart contracts (= systole) and pumps blood into the arteries, it brings
about a rise in pressure. The highest value is known as the systolic pressure and
is the first value to be measured. When the cardiac muscle relaxes to take in fresh
blood, the pressure in the arteries also falls. Once the arteries are relaxed, a
second reading is taken – the diastolic pressure.
How is blood pressure measured?
The ecomed BW-80E is a blood pressure unit which is used to measure blood
pressure from the wrist. The measurement is carried out by a microprocessor,
which, via a pressure sensor, measures the vibrations resulting from the inflation
and deflation of the cuff over the artery.
Blood pressure classification WHO
These values have been specified by the World Health Organisation (WHO)
without making any allowance for age.
Low blood pressure systolic <100 diastolic <60
Normal blood pressure (green display area
/
)
systolic 100 - 139 diastolic 60 - 89
Forms of high blood pressure / hypertension
Mild hypertension
(yellow display area
/
)
systolic 140 – 159 diastolic 90 – 99
Moderate hypertension (orange display area
/
)
systolic 160 – 179 diastolic 100 – 109
Severe hypertension (red display area
/
)
systolic >_180 diastolic >_110
WARNING
Blood pressure that is too low represents just as great a health
risk as blood pressure that is too high! Fits of dizziness may lead
to dangerous situations arising (e.g. on stairs or in traffic)!
Influencing and evaluating readings
Measure your blood pressure several times, then record and compare the
results. Do not draw any conclusions from a single reading.
Your blood pressure readings should always be evaluated by a doctor who is
also familiar with your personal medical history. When using the unit regularly
and recording the values for your doctor, you should visit the doctor from time
to time to keep him updated.
When taking readings, remember that the daily values are influenced by
several factors. Smoking, consumption of alcohol, drugs and physical exertion
influence the measured values in various ways.
• Measure your blood pressure before meals.
• Before taking readings, allow yourself at least 5-10 minutes rest.
If the systolic and diastolic readings seem unusual (too high or too low) on
several occasions, despite correct use of the unit, please inform your doctor.
This also applies to the rare occasions when an irregular or very weak pulse
prevents you from taking readings.
Starting up
Insert / replace batteries
You must insert the batteries provided before you can use your unit. The lid of
the battery compartment
is located on the right hand side of the unit. Open
it pushing the cover, remove it and insert the 2 AAA LR03 type 1.5 V batteries
supplied. Ensure correct polarity when inserting the batteries (as marked on
the battery compartment cover). Close the battery compartment. Replace the
batteries if the change battery symbol
appears on the display
or if nothing
appears on the display after the unit has been switched on.
Settings
1. Start to set:
Press and hold button
when power off, 0 or 0. 0 will be displayed, then
the setting begins.
2. Unit setting:
Continue to above step. The unit (mmHg/kPa) will be changed when press
button
each time. Press button
when you confirm the unit, then it
will enter into the user setting mode.
3. User setting:
Continue to above step,the screen will display or , press button
, it
will be changed between user and user , press button
when you
confirm the user, then it will enter into the year setting mode.
4. Year setting:
The input for setting the year then starts flashing. Keep pressing the button
until the required year appears. Press button
when you confirm the
year, then it will enter into the month and date setting mode.
5. Month and date setting:
The input for setting the month then starts flashing. Keep pressing the
button
until the required month appears. Press button
when you
confirm the month, then it will set the date.
As with the month setting, keep pressing the button
until the required
day appears. Press button
when you confirm the date, then it will switch
to the time setting mode.
6. Time setting:
The input for setting the hour then starts flashing. Keep pressing the button
until the required hour appears. Press button
when you confirm the
hour.
The input for setting the minute then starts flashing. As with the hour setting,
keep pressing the button
until the required minute appears. Press
button
when you confirm the minute, which will complete the setting mode.
CL appears in the display. You can quit set-up mode by pressing button
.
If the batteries are changed the information is lost and the date and time must
be set again.
Fitting the cuff
1. Attach the clean cuff to the left wrist which is free from clothing with the palm
of the hand facing upwards (fig.1).
2. There should be a gap of approx. 1 cm between the cuff and the palm of the
hand (fig.2).
3. Press the Velcro tape firmly around the wrist ensuring that it is not so tight that
it could falsify the measured result (fig.3).
Correct position for taking readings
• Carry out the measurement whilst seated.
• Relax your arm and place it comfortably, on a table for instance.
Raise the wrist so that the cuff of the monitor is at heart level (fig.4,
position a = too high, b = correct, c = too low).
• Keep still during the measurement: not move and do not talk as this may alter
the measurements.
Taking a blood pressure measurement
After the cuff has been appropriately positioned,the measurement can begin:
1. Switch the unit on by pressing the button
.
2. If the button
is pressed, all display characters are shown (display test).
This test can be used to check that the display is indicating properly and in full.
3. The unit is ready for measurement and the number 0 flashes for 2 seconds. It
automatically inflates the cuff slowly in order to measure your blood pressure.
The rising pressure in the cuff is shown on the display.
4. The unit inflates the cuff until sufficient pressure has built up for a
measurement. Then the unit slowly releases air from the cuff and carries
out the measurement. When the device detects the signal, the heart symbol
on the display starts to flash.
5. When the measurement is finished,the cuff is deflated completely. The
systolic and diastolic blood pressure and the pulse value with the pulse symbol
appear on the display
. It will read in kPa as unit when you choose
kPa. The blood pressure indicator
flashes next to the relevant coloured bar
depending on the WHO blood pressure classification. If the unit has detected
an irregular heartbeat, the arrhythmia display
also flashes.
WARNING
Do not take any therapeutic measures on the basis of a self
measurement. Never alter the dose of a medicine prescribed
by a doctor.
6. The readings are automatically saved in the selected memory ( or ). Up to
90 measured values with date and time can be stored in each memory.
7. The measurement readings remain on the display. The unit switches off
automatically after approx. 3 minutes if no buttons are pressed. The unit may
also be switched off using the button
.
Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for any reason
(eg. the patient feels unwell) the button
can be pressed at any time. The
device immediately decrease the cuff pressure automatically.
Displays of stored values
This unit features 2 separate memories, each with a capacity of 90 memory slots.
Results are automatically stored in the memory selected by the user. Press the
button
when power off, to call up the measured values stored. All values
average appear on the display. Pressing the button
again, the last
measurement value will be shown. Repeatedly pressing the button
displays the respective values measured previously. If you have reached the final
entry and do not press a button, the unit automatically switches itself off after
approximately 3 min. in memory recall mode. Memory recall mode can be exited
at any time by pressing the button
, which will also switch the unit off. The
oldest value is deleted if 90 measured values have been stored in the memory
and a new value is saved.
Memory - clear of measurements
If you are sure that you want to permanently remove all stored memories, press
and hold the button
to enter into setting mode when power off. Press
button
again for 7 times until CL appears. Press the button
, CL will
flash for 3 times to clear all the memories. After this press button
, M and
no” will be shown on the display which mean that no memory in store.
Error messages and error remedying
Error indicators
The following symbol will appear on the display when measuring abnormal.
Symbol Cause Correction
E-1 Weak signal or Wrap the cuff properly.
pressure change Remeasure with correct way.
suddenly
E-2 External strong When near cell phone or other high radiant
disturbance device, the measurement will be failed.
Keep quiet and no chatting when measure.
E-3 It appears error Wrap the cuff properly.
during the process Remeasure.
of inflating
E-5 Abnormal blood Repeat the measurement after relax for
pressure 30 mins. If get unusual readings for 3 times,
please contact your doctor.
Low battery The batteries are too low or empty.
Replace all batteries with new 1.5 V
type AAA LR03 batteries.
Trouble removal
Problem Check Cause and solutions
No power Check the battery power Replace new one
Check the polarity position Installation for proper placement
of the batteries polarities
Err and stop Check whether move the arm Keep the wrist and body
working when inflate peaceful
Check if chatting when Keep quiet when measure
measured
Cuff leak Check whether the cuff Wrap the cuff tightly
wrap too loose
Please contact the distributer if you can’t solve the problem. Do not disassemble
the unit by yourself.
Cleaning and care
Remove the batteries before cleaning the unit. Clean the unit and the cuff
using a soft cloth lightly moistened with a mild soapy solution. Never use
abrasive cleaning agents, alcohol, naphta, thinner or gasoline etc.. Never
immerse the unit or any component in water. Be cautiious not to get any
moisture in the main unit.
Do not use the unit again until it is completely dry.
Only inflate the cuff when it is in position around the wrist.
Do not expose the unit to direct sunlight; protect it against dirt and moisture.
Do not subject the unit to extreme hot or cold temperature. Keep the unit in the
storage case when no use. Store the unit in a clean and dry location.
Disposal
This product must not be disposed of together with domestic waste.
All users are obliged to hand in all electrical or electronic devices,
regardless of whether or not they contain toxic substances, at a
municipal or commercial collection point so that they can be disposed
of in an environmentally acceptable manner.
Remove the batteries before disposing of the device/unit. Do not dispose of
old batteries with your household waste, but at a battery collection station at a
recycling site or in a shop.
Consult your local authority or your supplier for information about disposal.
Guidelines / Standards
This blood pressure monitor meets the requirements of the EU standard for
non-invasive blood pressure monitors. It is certified in accordance with
EC Guidelines and carries the CE symbol (conformity symbol) “CE 0297”.
The blood pressure monitor corresponds to European standards EN 1060-1 and
EN 1060-3.
The specifications of EU Guideline “93/42/EEC of the Council Directive dated 14
June 1993 concerning medical devices” are met and likewise those of EU
Guideline.
Electromagnetic compatibility: (see attached leaflet)
Technical specifications
Name and model : ecomed Blood pressure monitor BW-80E
Display system : Digital display
Memory slots : 2 x 90 measurement values
Measuring method : Oscillometric
Voltage supply : 3 V=, 2 x 1.5V battery AAA LR03
Measuring range blood pressure : 0 – 299 mmHg
Measuring range - pulse : 40 – 199 beats/min.
Maximum measuring deviation
of static pressure : ± 3 mmHg
Maximum measuring deviation
of pulse values : ± 5 % of the value
Pressure generation : Automatic with pump
Air release : Automatic
Autom. switch-off : After 3 min.
Operating conditions : +5 °C to +40 °C,
15 to 85 % max. relative humidity
Storage conditions : -20 °C to +55 °C,
10 to 85 % max. relative humidity
Dimensions (main unit) : approx. 66,5 x 63,5 x 55 (L x W x H)
Weight : approx. 92 g without batteries
Item number : 23210
EAN number : 40 15588 23210 7
In accordance with our policy of continual product improvement, we
reserve the right to make technical and visual changes without notice.
Warranty and repair terms
Please contact your supplier or the service centre in case of a claim under the
warranty. If you have to return the unit, please enclose a copy of your receipt and
state what the defect is.
The following warranty terms apply:
1. The warranty period for ecomed products is two years from date of purchase.
In case of a warranty claim, the date of purchase has to be proven by means
of the sales receipt or invoice.
2. Defects in material or workmanship will be removed free of charge within the
warranty period.
3. Repairs under warranty do not extend the warranty period either for the unit
or for the replacement parts.
4. The following is excluded under the warranty:
a. All damage which has arisen due to improper treatment, e.g.
nonobservance of the user instructions.
b. All damage which is due to repairs or tampering by the customer or
unauthorised third parties.
c. Damage which has arisen during transport from the manufacturer to the
consumer or during transport to the service centre.
d. Accessories which are subject to normal wear and tear as batteries, cuff etc..
5. Liability for direct or indirect consequential losses caused by the unit are
excluded even if the damage to the unit is accepted as a warranty claim.
MEDISANA AG, 41468 NEUSS, GERMANY.
ECOMED is a registered trademark of MEDISANA AG.
For service, accessories and spare parts, please contact:
MEDHEALTH SUPPLIES LTD.
Unit 6, Dock offices
Surrey Quays Road
London
SE16 2XU
Tel.: + 44 / 207 - 237 88 99
Fax: + 44 / 207 - 231 13 00
eMail: info@medhealthcare.co.uk
Internet: www.medisana.co.uk
Explanation of symbols
IMPORTANT
Follow the instructions for use!
Non-observance of these instructions
can result in serious injury or damage
to the device.
WARNING
These warning notes must be observed
to prevent any injury to the user.
CAUTION
These notes must be observed to prevent
any damage to the device.
NOTE
These notes give you useful additional
information on the installation or
operation.
Device classification: type B applied part
LOT number
Manufacturer
Date of manufacture
Proper use
The device is intended for measuring the blood
pressure from the wrist of an adult.
Contraindication
The device is not suitable for measuring the
blood pressure of children. Ask your doctor
before using it on older children.
The device is not suitable for persons with
arrhythmia, diabetes, cardiovascular problems
or who have had a stroke.
Common factors of wrong measurement
Please keep quiet for 5-10 minutes and avoid
eating, drinking alcohol, smoking, exercising
and bathing before taking measurement. All
these factors will influence the measurement
result.
Remove watch or other ornaments from the
measured wrist.
Always measure on the same wrist (normally left).
Take measurement regularly at the same time of
every day, as blood pressure changes even during
the day.
All efforts by the patient to support their arm can
increase blood pressure.
Make sure you are in a comfortable, relax position
and do not activate any of the muscles in the
measurement arm during measurement. Use a
cushion for support if necessary.
If the wrist lies lower or higher than the heart, a
false reading will be obtained.
A loose or open cuff causes false reading.
With repeated measurements , blood accumulates
in the wrist which can lead to false reading.
Consecutive blood pressure measurements should
be repeated after 1 minute pause or after the
arm has been held up in order to allow the
accumulated blood to flow away.
GB Unit and LCD Display
Blood pressure monitor
BW-80E
Instruction Manual
Please read carefully!
23210_BW-80E_16spr_10_09_12.qxd:Handgelenk-Blutdruckgerät 10.09.2012 21:03 Uhr Seite 2
2

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln
1

Forum

ecomed-bw-80e

Suche zurücksetzen

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Ecomed BW-80E wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info