A.6 Error margins
The following error margins are valid for the CYCLE models of the CARDIO LINE 400 MED
series in compliance with the standard DIN VDE 750-238:
1. The read-out error of the wattage p must not exceed ± 5% of the displayed value.
However, the read-out error is not required to fall below ± 3W.
2. The read-out error of the rotational speed n is set to a maximum of ± 2 min-1 above
40 min-1.
3. The measuring device to assess the output calculated on the base of deceleration
torque and the rotational speed of the treadle ergometer must not exceed an error
margin of 1 %.
The characteristic curve family of the deceleration torque control‘s work capacity can be
assessed from the following figure:
The work capacity is shown on the display as follows:
As shown right to the rotational
speed (CYCLE 450/457/457 SPO
2
)
As shown right/left to the rotational
speed (CYCLE 400/407)
Devices in compliance with
VDE 750-238 EN 957
none 5% range 10% range
Points 10% range 10% range
Arrows Deviation > 10% Deviation > 10%
Rotational speed
40 50 60 70 80 90 100
100 200 300 400
Power