Note: The pole orientation of the batteries must be as shown
in the in Figure, otherwise the device may be damaged.
Please insert or remove the batteries in the correct order , other-
wise the devices tab could be damaged. Please remove the
batteries if the oximeter is no longer in use. If the battery status
indicator lights up, replace the batteries in time.
Warning: Do not dispose the battery in normal domestic
waste! Keep out of reach of young children. Do not dispose
of the battery in a re. It may explode.
The heart
rate-bar ashes
according to the
displayed pulse
rate.
Install two AAA batteries into the
battery compartment. Match the
plus (+) and minus (-) signs in the
compartment. If the polarities are
not matched, damage may be
caused to the oximeter .
Slide the battery door cover hori-
zontally along the arrow shown as
the picture.
HOW TO REPLACE THE BA TTERY
DISCRIPTION OF DISPLA Y
1.
2.
verwendet werden. Der Einsatz während ein er sportlichen Betätigung wird nicht empfohlen.)
Einlegen der Batteri en
1. Schieben Sie die Abdeckung des Pulsoximeters herunter wie die Abbildung zeigt
2. Die beiden AA A-Batterien mit der richtigen Polausrichtung
in das Batteriefach einlegen.
3. Die Batterieabdeckung, wie gezeigt, in Pfeilrichtung
waagerecht daraufschieben.
Achtung: Die Polausrichtung d
er Batterien muss wie in der
Abbildung gekennzeichnet, erfolgen, andernfalls kann das
Gerät bes chädigt werden.
Die Batterien bitte in der richtigen Reihenfolge einleg en oder entfernen, sonst könnte die
Gerätelasche beschädigt werden.
Die Batterien bitte entnehmen, wenn das Oximeter längere Zeit nicht benutzt wird.
Wenn der Batterieanzeigestatus leuchtet, rechtzeitig die Batterien auswechseln.
Halten Sie die Batterie n von Kleinkindern fern. Die Batterien nicht ins Feuer werfen,
sie könnte explodieren.
Beschreibung d es Displays
Der Pulsfrequenz-Balken blink
t entsprechend der angezeigten Pulsr ate.
Anwendung
1 Zwei AAA-Batterien in das Batteriefach einlegen, bevor die Abdeckung angebracht wird.
2 Das Gerät am hinteren T eil zusammendrücken, um
den vorderen T eil zur Aufnahme des Fingers zu öffnen.
3 Wenn sich Ihr Finger im Oximeter befindet, muss
die Nageloberfläche nach o ben zeigen.
4 Nach Einlegen des Fingers den hinter en T eil wieder los lassen.
5 Den Schaltknopf an der Frontseite einmal drücken.
6 Während das Oximeter arbe itet, sollten Sie Ihren Finger nicht bewegen u nd sich selbst
ruhig verhalten.
7 Auf der Anzeig e die entsprechenden Messwerte ablesen.
Nach dem Einschalten wechselt das Oximeter jedes Mal, wenn Sie den Einschaltknopf
drücken, zu einem anderen Anzeigemodus. Es erscheinen folgende 6 Anzeigemodi:
1. 2. 3.
4. 5. 6.
Drücken Sie den Einschaltknopf längere Z eit (länger als eine Sekunde), um die Helli gkeit des
Displays einzustellen. Durch Halten des Einschaltknopfes ändert sich stufenweise die
Helligkeit des Oximeters. Es gibt 10 Helligkeitsstufen. Stufe 4 ist die S tandardhelligkeit.
Das Pulsoximeter wird mit einer Schutzhülle geliefert, um es vor Schmutz und Stoß zu
schützen.
Bewertung der gemessenen Werte
Bei einem gesunden Patienten liegt die Sau erstoffsättigung zwischen 96 und 98 %. Werte
unterhalb von 95 % können schon e inen Krankheitswert anzeigen und eine medizinisc he
Behandlung erforderl ich machen. Wenn Ihr Sauerstoff-Sättigungswert unter 95 % liegt,
konsultieren Sie bitte Ihren Arzt.
Normale Pulswerte hängen vom Alter und dem Fitness-Level einer Person ab.
Im Folgenden ein paar Referenzwerte für eine normale Pul srate:
• Kinder unter 1 Jahr: 100 - 160 bpm
• Kinder zwischen 1 und 10 Jahren: 70 - 12 0 bpm
• Personen über 10 Jahre: 60 - 100 bpm
• Trainierte Personen/ Athleten: 40 - 60 bpm
Spezifikation
1. Model-Nummer: GT -300C2 03
2. Anzeigetyp: 1-farbige OLED-Anzeige
3. SpO
2
:
Displaybereich:0% bis 99 %
Messbereich: 70% bis 99 %
Genauigkeit: 7 0% bis 99%: ± 3%; 0% bis 69
% nicht definiert
Auflö sung : 1%
4. Pulsfrequenz:
Displaybreich: 0 bis 254 bpm
Messbereich: 30 bis 235 bpm
Genauigkeit: 30 bis 99bpm, ±2bpm; 100 bis 235 bpm, ±2%
Auflösung 1 bpm
Pulsintensität: Pulsfrequenz-Balken
5.
Wellenlänge Strahlungsleistung
RED 660± 2nm 1.8mW
IR 940± 10nm 2.0mW
HINWEIS: Die Informationen über Wellenlängenbereiche kann besonders für Ärzte
Pulsfrequenz-Balken
SpO
2
/ Sauerstoff-Sättigung
Pulsfrequenz
Ein/Aus-
Schalter
Batterieanzeigestatus
grafische Darstellung von SpO
2
!
!
EN
Install two AAA batteries according to the Battery Installation
instructions.
Squeeze the device at the rear to open the front part to
accommodate the nger .
If your nger is in the oximeter , your nail must be facing up.
After inserting the nger release the
rear section again.
Press the power switch in the front of
the device once.
While the oximeter is measuring, you should not move your
nger and keep yourself quiet.
The display will read the corresponding measured values.
1.
2.
3.
4.
5.
6.
7.
HOW TO OPERA TE
verwendet werden. Der Einsatz während einer sportlichen Betätigung wird nicht empfohlen.)
Einlegen der Batterien
1. Schieben Sie die Abdeckung des Pulsoximeters herunter wie die Abbildung zeigt
2. Die beiden AAA-Batterien mit der richtigen Polausrichtung
in das Batteriefach einlegen.
3. Die Batterieabdeckung, wie gezeigt, in Pfeilrichtung
waagerecht daraufschieben.
Achtung: Die Polausrichtung d
er Batterien muss wie in der
Abbildung gekennzeichnet, erfolgen, an dernfalls kann das
Gerät beschädigt werden.
Die Batterien bitte in der richtigen Reihenfolge einlegen oder entfernen, sonst könnte die
Gerätelasche beschädigt werden.
Die Batterien bitte entnehmen, wenn das Oximeter längere Zeit nicht benut zt wird.
Wenn der Batterieanzeigestatus leuchtet, rechtzeitig die Batterien auswechseln.
Halten Sie die Batterien von K leinkindern fern. Die Batterien nicht ins Feuer werfen,
sie könnte explodieren.
Beschreibung des Displ ays
Der Pulsfrequenz-Balken blink
t entsprechend der angezeigten Pulsrate.
Anwendung
1 Zwei AAA-Batterien in das Batteriefach einle gen, bevor die Abdeck ung angebracht wird.
2 Das Gerät am hinteren T eil zusammendrücken, um
den vorderen T eil zur Aufnahme des Fingers zu öffnen.
3 Wenn sich Ihr Finger im Oximeter befindet, muss
die Nageloberfläche nach oben zeigen.
4 Nach Einlegen des Fingers den hinteren T eil wieder los lassen.
5 Den Schaltknopf an der Frontseite einmal drücken.
6 Während das Oximeter arbei tet, sollten Sie Ihren Finger nicht bewegen und sich selbst
ruhig verhalten.
7 Auf der Anzeige die entsprechenden Mess werte ablesen.
Nach dem Einschalten wechselt d as Oximeter jedes Mal, wenn Sie den Einschaltknopf
drücken, zu einem anderen Anzeigemodus. Es erscheinen folgende 6 Anzeigemodi:
1. 2. 3.
4. 5. 6.
Drücken Sie den
Einschaltknopf längere Zeit (länger als eine Sek unde), um die Helligkeit des
Displays einzustellen. Durch Halten des Einschaltknopfes ändert sich stufenweise die
Helligkeit des Oximeters. Es gibt 10 Helligkeitsstufen. Stufe 4 ist die S tandardhel ligkeit.
Das Pulsoximeter wird mit einer Schutzhülle geliefert, um es vor Schmutz und Stoß zu
schützen.
Bewertung der gemessenen Werte
Bei einem gesunden Patienten liegt die Sauerstoffsättigung zwischen 96 und 98 %. Werte
unterhalb von 95 % können schon einen Krankhe itswert anzeigen und eine medizinische
Behandlung erforderlich mac hen. Wenn Ihr Sauerstoff-Sättigungswert unter 95 % liegt,
konsultieren Sie bitte Ihren Arzt.
Normale Pulswerte hängen vom Alter und dem Fitness-Level einer Person ab.
Im Folgenden ein paar Referenzwerte für eine normale Pulsrate:
• Kinder unter 1 Jahr: 100 - 1 60 bpm
• Kinder z wischen 1 und 10 Jahren: 70 - 120 bpm
• Personen über 10 Jahre: 60 - 100 bpm
• Trainierte Personen/ Athleten: 40 - 60 bpm
Spezifikation
1. Model-Nummer: GT-300C203
2. Anzeigetyp: 1-farbige OLED-Anzeige
3. SpO
2
:
Displaybereich:0% bis 99%
Messbereich: 70% bis 99 %
Genauigkeit: 70% bis 99%: ± 3%; 0% bis 69
% nicht definiert
Auflö sung : 1%
4. Pulsfrequenz:
Displaybreich: 0 bis 254 bpm
Messbereich: 30 bis 235 bpm
Genauigkeit: 30 bis 99bpm, ±2bpm; 100 bis 235bpm, ±2%
Auflösung 1 bpm
Pulsintensität: Pulsfrequenz-Balken
5.
Wellenlänge Strahlungsleistung
RED 660± 2nm 1.8mW
IR 940± 10nm 2.0mW
HINWEIS: Die Informationen über Wellenlängenbereiche kann besonders für Ärzt e
Pulsfrequenz-Balken
SpO
2
/ Sauerstoff-Sättigung
Pulsfrequenz
Ein/Aus-
Schalter
Batterieanzeigestatus
grafische Darstellung von SpO
2