See picture 7-4.
Picture 7 Picture 7-1 Picture 7-2 Picture 7-3 Pictur e 7-4
c. Pr ess “MEM” button again, LCD will display the average value of all the
results which is measured fr om 18 o’clock to 20 o’clock in last 7 days in
the current user memory zone. See picture 7-5. If no r esult stored from 18
o’clock to 20 o’clock in last 7 days, LCD will display “0” for blood pressure
and pulse rate. See picture 7-6.
Picture 7-5 Picture 7-6 Picture 7-7 Picture 7-8 Pictur e 7-9
d. Press “MEM” button again, the most recent result will be displayed with date
and time stamp. See picture 7-7. Irregular heartbeat symbol (if any) and blood
pressure classication indicator will blink at the same time. If the monitor has
no result stored in the curr ent user memory zone, the LCD will display “0” for
blood pressure and pulse rate. See pictur e 7-8.
e. Press “MEM” button again to review the next result. See pictur e 7-9. In this
way , repeatedly pressing the “MEM” button displays the respective r esults
measured previously .
f. When reviewing the results, the monitor will turn off automatically after 1 min-
ute of no operation. Y ou can also press the “ST ART/STOP” button to turn off
the monitor manually .
Note: When the monitor displaying the measurement, the classication color
indicator can be shown different color accor ding to the systolic pressure and
diastolic pressure. Refer to the “ASSESSING HIGH BLOOD PRESSURE FOR
ADUL TS” section.
8. Deleting measurements from the memory
When any result is displaying, keeping on pressing but-
ton “MEM” for three seconds, all results will be deleted.
Press the button “ST ART/STOP”, the monitor will turn off.
9. Assessing high blood pressure for adults
The following guidelines for assessing high blood pressure (without r egard to
age or gender) have been established by the W orld Health Organization (WHO).
Please note that other factors (e.g. diabetes, obesity , smoking, etc.) need to be
taken into consideration. Consult with your physician for accurate assessment,
and never change your treatment by yourself.
Classication of blood pressure for adults
MAINTENANCE
1. Do not drop this monitor or subject it to strong impact.
2. Avoid high temperature and solarization. Do not immerse the monitor in
water as this will result in damage to the monitor .
Note: It is not intended to provide a basis of any type of rush toward emer gency
conditions/diagnosis based on the color scheme and that the color scheme is
meant only to discriminate between the different levels of blood pr essure.
10. T echnical alarm description
The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with no delay if
the determined blood pressure (systolic or diastolic) is outside the rated range
specied in part SPECIFICACIONS. In this case, you should consult a physician
or check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory
and cannot be adjusted or inactivated. This alarm condition is assigned as low
priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on
LCD will disappear automatically after about 8 seconds.
12. T roubleshooting (2)
ELECTROMAGNETIC COMP A TIBILITY INFORMA TION
T able 1
Emission
12 It is recommended the cuff should be disinfected 2 times every week if need-
ed (For example, in hospital or clinic). Wipe the inner side (the side contacting
the skin) of the cuff with a soft cloth squeezed after being moistened with
Ethyl alcohol (75-90%), then dry the cuff by airing.
EXPLANA TION OF SYMBOLS ON UNIT
Finally the blood pressure and pulse rate will be calculated and displayed
on the LCD screen. The blood pressur e classication indicator and Irregular
heartbeat symbol (if any) will blink on the screen. The result will be automati-
cally stored in the monitor .
d. After measurement, the monitor will turn off automatically after 1 minute of no
operation.
e. During measurement, you can press the “ST ART/STOP” button to turn off the
monitor manually .
Note: Please consult a health care professional for interpr etation of pressure
measurements.
7. Displaying stored results
a.
In StandBy Mode, press “MEM” button, the monitor will display sign of current
group. The amount of results in curr ent user memory zone will be displayed. See
picture 7. Press “ST ART/STOP” button to switch group,pr ess “MEM” to conrm
current group. Then LCD will display the average value of all results in the current
user memory zone. See picture 7-1. If no result stor ed in the current user memory
zone, LCD will display “0” for blood pressure and pulse rate. See pictur e 7-2.
b. Press “MEM” button, LCD will display the average value of all the r esults
which is measured from 5 o’clock to 9 o’clock in last 7 days in the curr ent
user memory zone. See picture 7-3. If no r esult stored from 5 o’clock to 9
o’clock in last 7 days, LCD will display “0” for blood pressure and pulse rate.
M32901-GB-Rev-2-02.20
21
22
23
24
20
19
27
28
29
30
26
25
33
34
35
36
32
31
32901 / KD-5923
PROBLEM POSSIBLE CAUSE SOLUTION
LCD shows
battery symbol
Low Battery Change the batteries
LCD shows “Er 0” Pr essure system is unstable
before measurement
Don’t move and try
again
LCD shows “Er 1” Fail to detect systolic pr essure
LCD shows “Er 2” Fail to detect diastolic
pressure
LCD shows “Er 3” Pneumatic system blocked or
cuff is too tight during ination
Apply the cuff correctly
and try again.
If the monitor is still ab-
normal, please contact
the local distributor or
the factory .
LCD shows “Er 4” Pneumatic system leakage
or cuff is too loose during
ination
LCD shows “Er 5” Cuf f pressure above
300mmHg
Measure again after ve
minutes. If the monitor
is still abnormal, please
contact the local
distributor or the factory
LCD shows “Er 6” Mor e than 3 minutes with cuff
pressure above 15 mmHg
LCD shows “Er 7” Inner memory err or
LCD shows “Er 8” Device parameter checking
error
LCD shows “Er A” Pressur e sensor parameter
error
No response when
you press button or
load battery
Incorrect operation or strong
electromagnetic interference
T ake out batteries for
ve minutes, and then
reinstall all batteries
T able 2
Enclosure Port
T able 3
Proximity elds from RF wir eless communications equipment
11. T roubleshooting (1)
PROBLEM POSSIBLE CAUSE SOLUTION
LCD Display shows
abnormal result
The cuff position was
not correct or it was not
properly tightened
Apply the cuff correctly
and try again
Body posture was not
correct during testing
Review the “BODY POSTURE
DURING MEASUREMENT”
sections of the instructions
and re-test
Speaking, arm or body
movement, angry , excited
or nervous during testing
Re-test when calm and
without speaking or moving
during the test
Irregular heartbeat
(arrhythmia)
It is inappropriate for people
with serious arrhythmia
to use this Electronic
Sphygmomanometer
Disposal: The product must not be disposed of along with other domestic
waste. The users must dispose of this equipment by bringing it to a specic
recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
WHO/ISH Denitions and classication of blood pressure levels
80 85 90 100 110
120
130
140
160
180
Severe Hypertension
Moderate Hypertension
Mild Hypertension
High-normal BP
Normal BP
Diastolic
(mmHg)
Systolic
(mmHg)
BLOOD PRESSURE
CLASSIFICA TION
SBP
mmHg
DBP
mmHg
Color
Indicator
Optimal
<120 <80 green
Normal
120-129 80-84 green
High-Normal
130-139 85-89 green
Grade 1 Hypertension 140-159 90-99 yellow
Grade 2 Hypertension 160-179 100-109 orange
Grade 3 Hypertension ≥180 ≥110 red
Follow instructions for use
Caution: read instructions (warnings) carefully
T ype BF applied part
WEEE disposal
0197
Medical Device complies with Directive
93/42/EEC
Manufacturer
Date of manufacture
Authorized representative in the Eur opean
community
Serial number
IP20 Covering Protection rate
Keep in a cool, dry place
Keep away from sunlight
Product code
Lot number
ANDON HEAL TH CO., L TD.
No. 3 JinPing Street, Y a An Road, Nankai District,
Tianjin 300190, China
Made in China
iHealthLabs Europe SAS
36 Rue de Ponthieu,
75008, Paris, France
01 97
IP20
Phenomenon Compliance Electromagnetic
environment
RF emissions CISPR 11 Group 1,
Class B
Home healthcare
environment
Harmonic distortion IEC 61000-3-2
Class A
Home healthcare
environment
V oltage uctuations
and icker
IEC 61000-3-3
Compliance
Home healthcare
environment
Phenomenon Basic EMC
standard
Immunity test levels
Home Healthcare
Environment
Electrostatic Dischar ge IEC 61000-4-2 ±8 kV contact
±2kV , ±4kV , ±8kV , ±15kV air
Radiated RF EM eld IEC 61000-4-3 10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity elds from RF
wireless communications
equipment
IEC 61000-4-3 Refer to table 3
Rated power frequency
magnetic elds
IEC 61000-4-8 30A/m
50Hz or 60Hz
T est frequency
(MHz)
Band (MHz) Immunity test levels
Professional healthcare facility
environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470 FM, ±5kHz deviation, 1kHz sine,
28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745
780
810 800-960 Pulse modulation 18Hz, 28V/m
870
930
1720 1700-1990 Pulse modulation 217Hz, 28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
3. If this monitor is stored near freezing, allow it to acclimate to room tempera-
ture before use.
4. Do not attempt to disassemble this monitor .
5. It is r ecommended the performance should be checked every 2 years or after
repair . Please contact the service center .
6. Clean the monitor with a dry , soft cloth or a soft cloth squeezed well after
moistened with water , diluted disinfectant alcohol, or diluted detergent.
7. No component can be maintained by user in the monitor . The circuit dia-
grams, component part lists, descriptions, calibration instructions, or other
information which will assist the user’ s appropriately qualied technical per -
sonnel to repair those parts of equipment which are designated r epairably
can be supplied.
8. The monitor can maintain the safety and performance characteristics for a
minimum of 10,000 measurements or thr ee years, and the cuff integrity is
maintained after 1,000 open–close cycles of the closure.
9. The monitor r equires 6 hours to warm from the minimum storage tempera-
ture between uses until the monitor is ready for its INTENDED USE when the
ambient temperature is 20°C.
10. The monitor requir es 6 hours to cool from the maximum storage temperature
between uses until the monitor is ready for its INTENDED USE when the
ambient temperature is 20°C.
11. Not servicing/maintenance while the monitor is in use.