10 GMC-I Messtechnik GmbH
7Self-Test
Because it is so easy to perform, the self-test should
be conducted before each use of the AT3-III E.
Please be careful to ensure that the case of the
AT3-III E is not being contacted with devices
under PE or earth potential during the self-test.
Performing the Protective Conductor Test
Ð
Insert the mains plug from the
SECUTEST.../
SECULIFE ST
into the earthing contact outlet (12)
at the AT3-III E.
ÐThe earthing contact plug (13) and the SE-
CUTEST port (14) of the AT3-III E may not be
plugged into the SECUTEST.../SECULIFE ST.
ÐConnect the AT3-III E to mains power.
ÐSet the SECUTEST.../SECULIFE ST to the
protective conductor test individual measure-
ment (RSL/RPE).
ÐContact protective conductor potential at the
mains system which is supplying power to the
AT3-III E with the test probe at the end of the
probe cable from the SECUTEST.../
SECULIFE ST, e.g. at the earthing contact of
an earthing contact outlet. If an excessively
high value or interruption is indicated, the pro-
tective conductor is interrupted.
The AT3-III E test adapter must be disconnected
from the mains, and the error in the electrical
system or the test adapter must be corrected. The
AT3-III E is equipped with a “Trigger IΔn” test key in
order to allow for testing the included IΔn safety
shutdown function at any time.
Performing the IΔn Self-Test
ÐRemove the device under test
(device or extension cable).
ÐSet the SECUTEST.../SECULIFE ST to individ-
ual measurement.
ÐSECUTEST.../SECULIFE ST (as of version
3.1): Select the “DI current” sub-menu (resid-
ual current).
ÐSignal lamps L1, L2 and L3 indicate the pres-
ence of line voltage.
ÐActivate the “Trigger IΔn” key (11).
ÐThe AT3-III E disconnects mains power from
the outlets (1 – 4).
ÐSignal lamps L1, L2 and L3 go out.
ÐThe function LED (10) blinks.
Disconnect the AT3-III E from the mains to re-
set. Reconnect the AT3-III E to the mains.
The test adapter is once again ready for use
after a brief self-test.
Note!
If mains power is not switched during this
test, the fuses in the AT3-III E may have
blown. If the AT3-III E still does not function
correctly after replacing the fuses, it must be
removed from service and repaired.
Measurement is no longer possible!
8 Technical Data
Residual Current Measuring Function
Measuring Range 0 20 mA
Transformation Ratio 1 V per 10 mA
Intrinsic Uncertainty ±(5%
of rdg.
+ 0.05 mA)
Nominal Ranges of Use
Line Voltage
L1/L2/L3/N 207 253 V AC
Frequency 49 51 Hz
Temperature 0 °C +40 °C
Line Voltage
Waveshape sinusoidal
Reference Conditions
Ambient Temperature +23 °C ±2K
Relative Humidity 50% ±5%
Line Voltage 230 V/400 V ±10%
Frequency
of Measured Quantity 50 Hz ±0.2%
Ambient Conditions
Operating Temperatures 10 + 40 °C
Storage Temperatures 25 + 60 °C
Humidity max. 75%, no
condensation allowed
Elevation up to 2000 m
Power Supply
Nominal Line Voltage
3~230/400 V/50 Hz/CAT II
Connection only permissible with overload
protection device In = 16 A I2 1.45 In
Electrical Safety
Standard
DIN VDE 0404-1/-2:2002-5
IEC/EN 61010-1:2010
Contamination degree 2
Safety Class I
Residual Current
Shutdown (4-pole) at approx. 18 mA
Device Fuses F0315 L250V
5 x 20 T32mA L 250 V
DIN EN 60127-2
10

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt AT3-III E wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info