4 GMC-I Messtechnik GmbH
3 Performing Measurements
3.1 Switching the Instrument On and Off, Energy-Saving
Mode, Automatic Shutdown
The instrument is switched on by pressing and holding the START
key. Briefly press the START key twice in order to switch the instru-
ment off, during which no voltage may be applied to the test
probes! The instrument is switched to the standby mode after
several seconds (reduced brightness), if none of the keys has
been pressed and no voltage is applied to the test probes. The
instrument is switched out of the standby mode (i.e. back to full
brightness) by pressing any key or applying voltage to the test
probes. The instrument is shutdown automatically if it has
remained inactive for about 1 minute, i.e. if no keys have been
pressed and no voltage has been applied to the test probes dur-
ing this time.
3.2 Instructions and Principles with Validity for
All Measurements
The desired functions or parameters are selected with the
R
ISO/USPD, FUNC, UTEST and UTEST keys. Measurement is
triggered by pressing the START key. All selected functions or
parameters remain valid until they are changed.
If (interference) voltage of greater than 10 V is applied to the
test probes, this is indicated at the display in the UIN field
along with the measured value, as well as the “!” symbol. If
this is the case, the START key is disabled.
Bild 3.1 (Interference) Voltage Display
If battery voltage is too low (only the red field is lit up in the
battery symbol), measurement cannot be started. After press-
ing the START key, the depleted battery symbol appears for
about 1 second (see figure below). Replace the batteries as
described in section 5.1 on page 7.
When measuring high insulation resistance, the measurement
cables should be laid unobstructed in the room, or placed
upon a material with good insulating properties.
Contact the device under test with the test probe before trig-
gering measurement with the START key, thus causing any
possible interference voltage to appear at the display.
Do not remove the test probes prematurely during the course
of the measurement, because this may lead to distorted mea-
surement results.
After insulation measurement has been completed, any remaining
residual voltage is displayed as U
IN , which may result from cable
capacitance. Contact to the device under test must be main-
tained for as long as a capacitive DUT is being discharged via the
test instrument’s internal resistor. The falling voltage value can be
observed directly at the U
IN display. Do not disconnect the DUT
until voltage U
IN is less than 25 V!
3.3 Measuring Functions
Attention!
Each time before measuring insulation resistance or surge
protection, make sure that the device under test is voltage-
free!
Do not touch the device under test or the test probes during
measurement, and for a short time thereafter as long the DUT
is being discharged.
The DUT may be charged with as much as 1000 V during
measurement, for which reason contact to the DUT must be
maintained after measurement has been completed until it is
fully discharged! Residual voltage appears at the display along
with the “!” warning symbol.
Do not remove the test probes until voltage has dropped to a
safe level and the “!” warning symbol has been cleared from
the display.
All consuming devices must be disconnected and the
switches must be turned on when measuring insulation resis-
tance between conductors.
3.3.1 Voltage Measurement
Connect the test instrument to the device under test.
Figure 3.4: Connection Example for Voltage Measurement
If a voltage of greater than about 10 V is present in the mea-
sured electrical circuit, its value is displayed in the U
IN field and
the “~” symbol appears for AC or polarity is displayed for DC.
The “+” symbol lights up if the positive pole is connected to
the fixed test probe, and “–” appears if the negative pole is
connected to the fixed test probe. “!” appears at the same
time as a warning. The START key is disabled!
Figure 3.2: Low Battery
Voltage
Figure 3.3: Low Battery
Voltage: Display
After Pressing the
START Key
Figure 3.5: Display of (Inter-
ference) Voltage
(Riso function)
Figure 3.6: Voltage
Measurement
(U
SPD function)
!
4

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt METRALINE ISO-CHECK wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gossen Metrawatt METRALINE ISO-CHECK

Gossen Metrawatt METRALINE ISO-CHECK Bedienungsanleitung - Holländisch - 8 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info