4 General Operation
The test leads are connected to the “+” and “COM” jacks.
Measuring High-Value Resistance with the KS-C Probe (Option)
When measuring electrostatic
discharge capacity for floor
coverings, the shielded cable
should also be connected to
the COM and SHIELD jacks (KS-C
accessory set, “cable set con-
sisting of measurement cable
and high-resistance measure-
ment cable for measurements
in the GΩ range”, see diagram). Be sure to observe color coding.
For measuring with the Test Probe for Remote Triggering (Option Z550A)
see section 14.5.1.
4.1 Switching On, Monitoring and Switching Off
If battery voltage falls below the allowable limit value (U < 8 V) the
instrument cannot be switched on, or it is immediately switched
off.
Measurement cannot be started in the resistance measuring
ranges in the event of external voltage.
The instrument only switches itself off automatically after comple-
tion of an automatic measuring sequence, and after the predeter-
mined on-time has expired (APOFF parameter in SETUP switch
position, see page 12). On-time is reset to its original value as
defined in the setup menu, as soon as any key or the rotary func-
tion switch is activated.
If the instrument is switched off automatically with the rotary
switch in any position other than OFF, it can be reactivated by
pressing the stby OFF key.
The instrument can be switched off manually by turning the rotary
switch to the OFF position.
Optical Indicators
1
Prerequisite: The LiLEd parameter is set to on (see page 13).
2
Function testing should be executed regularly (see following section on testing
LEDs).
+
C
O
M
S
H
I
E
L
D
KS-C
(Z541F)
LED Status Function – Cause
Limit value indication
1
Measured insulation resistance does not violate the limit
value.
– Measured low-resistance Rlo does not violate the limit
value.
Limit value indication
1
Measured insulation resistance has fallen short of the se-
lected limit value.
Measured low-resistance Rlo does has exceeded the per-
missible limit value.
External voltage when on or off (not M550N) and during discharge
2
Dangerous voltage of greater than 50 V is present at the mea-
surement inputs:
Initialization of the (insulation) resistance and
low-resistance measurements is disabled.
The discharging cycle has not yet been completed,
e.g. residual voltage at capacitive devices under test
Test/Measuring voltage with device switched on
Dangerous voltage of greater than 50 V is present at the mea-
surement inputs.
LCD Status Functions
Display
External voltage when off, and during discharge
Dangerous voltage of greater than 50 V is present at the
measurement inputs:
Initialization of the insulation resistance and
low-resistance measurements is disabled.
Test voltage applied during insulation measurement
The discharging cycle has not yet been completed,
e.g. residual voltage at capacitive devices under test
Limit
Green
Limit
Red
!
Red
14

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt METRISO INTRO wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gossen Metrawatt METRISO INTRO

Gossen Metrawatt METRISO INTRO Bedienungsanleitung - Deutsch - 6 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info