GMC-I Messtechnik GmbH 11
Opening of Equipment / Repair
The equipment may be opened only by authorized service per-
sonnel to ensure the safe and correct operation of the equipment
and to keep the warranty valid. Even original spare parts may be
installed only by authorized service personnel.
In case the equipment was opened by unauthorized personnel,
no warranty regarding personal safety, measurement accuracy,
conformity with applicable safety measures or any consequential
damage is granted by the manufacturer.
2.1 Checklist for High-Voltage Tests
Attention!
!
Measurements may not be performed under moist ambi-
ent conditions, where condensation has occurred or in
explosive atmospheres.
Personnel Safety
Ð Turn the machine under test off and lock it in the off position.
Ð Perform insulation resistance measurement.
Ð Make sure that the system is grounded.
Ð Secure the danger zone with barriers, closing even narrow en-
trance ways.
Ð Set up warning signs such that they are plainly visible.
Ð Set up warning lamps such that they are plainly visible.
Ð Attach the emergency stop button in a plainly visible fashion.
Ð Warn personnel working in the area of possible danger.
Ð Always turn the test instrument off with the key switch and re-
move the key before leaving the work area.
Safety Precautions for the Machine Under Test (recommended)
Ð
Review the circuit diagrams and make a note of all electrical circuits.
Ð Disconnect the neutral conductor from the mains.
Ð Short circuit each electrical circuit to itself.
Ð Disconnect control circuits with overvoltage arresters, if the ar-
resters would be triggered by utilized test voltage.
Ð Disconnect PELV circuits (no high-voltage testing is required
for these circuits).
Ð Test insulation at each circuit with 1000 V.
(If insulation resistance is OK at 1000 V, no failures should
occur during the high-voltage test.)
Ð Disconnect frequency converters.
Ð Caution in TN systems!
The protective conductor is connected to the neutral conduc-
tor in the junction box. High-voltage is thus applied between
the phase conductors and the neutral conductor.
The neutral conductor must be cut if necessary, because it is
not disconnected from the mains by means of fuses.
Setting Up the Test Instrument
High-Voltage Test
Ð
Test all circuits (conductors) against the protective conductor (all
switches in the mains circuit must be turned on, and testing must be
performed upstream and downstream from all relays and contactors).
Ð Remove all shorting devices after testing has been completed.
Testing Without Short-Circuiting
Ð Separately test all conductors from all circuits against the pro-
tective conductor (the machine could be damaged in the case
of arc-over).
Functions Test
Ð The machine must be tested for correct functioning after the
high-voltage test, especially as regards safety functions.
11

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt METRISO PRIME plus wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gossen Metrawatt METRISO PRIME plus

Gossen Metrawatt METRISO PRIME plus Bedienungsanleitung - Deutsch - 68 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info