1) Measuring Contact Current Without Tripping the RCD
Measuring Method
The instrument uses a measuring current of only 1/3 nominal
residual current for the determination of contact voltage U
IN
which occurs at nominal residual current. This prevents tripping of
the RCCB.
This measuring method is especially advantageous, because
contact voltage can be measured quickly and easily at any electri-
cal outlet without tripping the RCCB.
The usual, complex measuring method involving testing for the
proper functioning of the RCD at a given point, and subsequent
substantiation that all other systems components requiring pro-
tection are reliably connected at low resistance values to the
selected measuring point via the PE conductor, is made unneces-
sary.
N-PE Reversal Test
Additional testing is conducted in order to
determine whether or not N and PE are
reversed. The pop-up window shown at
the right appears in the event of reversal.
Attention!
!
Execute a data backup before starting measurement and
switch off all consumers in order to prevent the loss of
data in data processing systems.
Start Measurement
Amongst other values, contact voltage U
IN
and calculated earth-
ing resistance R
E
appear at the display panel.
Note
The measured earthing resistance value R
E
is acquired
with very little current. More accurate results can be
obtained with the selector switch in the R
E
position.
The DC + function can be selected here for sys-
tems with RCCBs.
Unintentional Tripping of the RCD due to Bias Current within the System
If bias currents should occur, they can be measured with the help
of a current clamp transformer as described in section 13.1 on
page 50. The RCCB may be tripped during the contact voltage
test if extremely large bias currents are present within the system,
or if a test current was selected which is too great for the RCCB.
After contact voltage has been measured, testing can be per-
formed to determine whether or not the RCCB is tripped within
the selected time limits at nominal residual current.
Unintentional Tripping of the RCD due to Leakage Current in the Measur-
ing Circuit
Measurement of contact voltage with 30% nominal residual cur-
rent does not normally trip an RCCB. However, the trip limit may
be exceeded as a result of leakage current in the measuring cir-
cuit, e.g. due to interconnected power consumers with EMC cir-
cuit, e.g. frequency converters and PCs.
2) Tripping Test after the Measurement of Contact Voltage
Press the
I
N
key.
The tripping test need
only be performed at
one measuring point for
each RCCB.
If the RCCB is tripped at nominal residual current,
the MAINS/NETZ LED blinks red (line voltage disconnected) and
time to trip t
a
and earthing resistance R
E
appear at the display
panel.
If the RCCB is not tripped at nominal residual current,
the RCD/FI LED lights up red.
Contact Voltage Too High
If contact voltage U
IN
, which has been measured with 1/3 nomi-
nal residual current I
N
and extrapolated to I
N
, is > 50 V (> 25 V),
the U
L
/R
L
LED lights up red.
If contact voltage U
IN
exceeds 50 V (25 V) during the measuring
sequence, safety shut-down occurs.
Note
Safety Shut-down: At up to 70 V, a safety shut-down is
tripped within 3 seconds in accordance with IEC 61010.
Contact voltages of up to 70 V are displayed. If contact voltage is
greater than 70 V, U
IN
> 70 V is displayed.
Limit Values for Allowable, Continuous Contact Voltage
The limit for allowable, continuous contact voltage is U
L
=50V for
alternating voltages (international agreement). Lower values have
been established for special applications (e.g. medical applica-
tions: U
L
=25V).
Attention!
!
If contact voltage is too high, or if the RCCB is not
tripped, the system must be repaired (e.g. earthing resis-
tance is too high, defective RCCB etc.)!
3-Phase Connections
For proper RCD testing at three-phase connections, the tripping
test must be conducted for one of the three phase conductors
(L1, L2 and L3).
Inductive Power Consumers
Voltage peaks may occur within the measuring circuit if inductive
consumers are shut down during an RCCB trip test. If this is the
case, the test instrument may display the following message: No
measured value (– – – ). If this message appears, switch all power
consumers off before performing the trip test. In extreme cases,
one of the fuses in the test instrument may blow, and/or the test
instrument may be damaged.
19

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln
1

Forum

gossen-metrawatt-profitest--mxtra

Suche zurücksetzen

  • 1) Kann ich die Profitest Mxtra Oberfläche auf den Monitor darstellen..?
    2) Wie wird da eine Bluetooth Tastaur angeschlossen..? kann ich dann über das Pach Feld das Messgerät einstellen und die Funktionen wählen...? Eingereicht am 28-11-2022 16:37

    Antworten Frage melden
  • Wenn ich den Adapter nr 4 Prüfspitze L-PE zur des Schleifenwiderstandes kommt immer das symbol stopp mit 2 mal xx und so einer art sicherungssymbol . was bedeutet das ? Eingereicht am 14-10-2017 09:33

    Antworten Frage melden

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt PROFITEST MXTRA wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gossen Metrawatt PROFITEST MXTRA

Gossen Metrawatt PROFITEST MXTRA Bedienungsanleitung - Holländisch - 92 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info