590776
85
Zoom out
Zoom in
Vorherige Seite
1/96
Nächste Seite
GMC-I Messtechnik GmbH 85
1)
U > 230 V with 2 or 3-pole adapter only
2
1
·
/ 2
·
IN > 300 mA and 5
·
IN > 500 mA and If > 300 mA only up to U
N
230 V !
3
The transformation ratio selected at the clamp (1 ... 1000 mV/A) must be set in the “Type” menu with the rotary switch in the “SENSOR” position.
4
Where R
Eselective
/R
Etotal
< 100
Special Function MPRO, MXTRA
5
Signal frequency without interference signal
6
PRO-RE (Z501S) adapter cable for test plug, for connecting earth probes (E-Set 3/4)
7
PRO-RE/2 (Z502T) adapter cable for test plug, for connecting the generator clamp
(E-CLIP2)
8
Generator clamp: E-CLIP2 (Z591B)
9
Clamp meter: Z3512A (Z225A)
10
Where RE.sel/RE < 10 or clamp current > 500 µA
11
Where RE.H/RE 100 and RE.E/RE 100
12
Where d = 20 m
13
Where d = 2 m
14
Where Z
L-PE
< 0.5 , I
k
> U
N
/0.5 is indicated
15
)Only where RANGE = 20 k
16
Only where RANGE = 50 k or AUTO
Key: D = digits, rdg. = measured value (reading)
R
ISO
R
ISO
, R
E ISO
1 ... 999 k
1.00 ... 9.99 M
10.0 ... 49.9 M
1 k
10 k
100 k
I
K
= 1.5 mA 50 k... 500 M
U
N
= 50 V
I
N
= 1 mA
k range
±(5% v.M.+10D)
Mrange
(5% rdg. + 1 d)
k
range
(3% rdg.+10d)
M
range
(3% rdg. + 1 d)
●●
1 ... 999 k
1.00 ... 9.99 M
10.0 ... 99.9 M
1 k
10 k
100 k
U
N
= 100 V
I
N
= 1 mA
1 ... 999 k
1.00 ... 9.99 M
10.0 ... 99.9 M
100 ... 200 M
1 k
10 k
100 k
1 M
U
N
= 250 V
I
N
= 1 mA
1 ... 999 k
1.00 ... 9.99 M
10.0 ... 99.9 M
100 ... 500 M
1 k
10 k
100 k
1 M
U
N
= 500 V
U
N
= 1000 V
I
N
= 1 mA
U
10 ... 999 V–
1.00 ... 1.19 kV
1 V
10 V
10 ... 1.19 kV (3% rdg. + 1 d)
(1.5% rdg. + 1 d)
R
LO
R
LO
0.01 ... 9.99
10.0 ... 199.9
10 m
100 m
I
m
200 mA
I
m
< 200 mA
0.1 ... 5.99
6.0 ... 100
U
0
= 4.5 V (4% rdg. + 2 d) (2% rdg. + 2 d)
SEN-
SOR
I
L/Amp
0 ... 99.9 mA 0.1 mA
5 ... 1000 mA
3
(10% rdg.+8d) (4% rdg.+7d)
100 ... 999 mA 1 mA (10% rdg.+3d) (4% rdg. + 2 d)
0 ... 99.9 A 0.1 A
5 ... 150 A
3
(8% rdg. + 2 d) (3% rdg. + 2 d)
100 ... 150 A 1 A (8% rdg. + 1 d) (3% rdg. + 1 d)
0 ... 99.9 mA 0.1 mA
5 ... 1000 mA
3
0.05 ... 10 A
3
0.5 ... 100 A
3
5 ... 1000 A
3
(7% rdg.+8d) (4% rdg.+7d)
100 ... 999 mA 1 mA (5% rdg.+3d) (2% rdg. + 2 d)
1.0 ... 9.99 A 0.01 A (4% rdg. + 2 d) (2% rdg. + 2 d)
10.0 ... 99.9 A 0.1 A (4% rdg. + 2 d) (2% rdg. + 2 d)
100 ... 999 A 1 A (4% rdg. + 1 d) (2% rdg. + 1 d)
1.00 ... 1.02 kA 0.01 kA (4% rdg. + 1 d) (2% rdg. + 1 d)
0 ... 99.9 mA 0.1 mA
1 V/A 30 ... 1000 mA
3
U
N
= 120/230/
400 V
f
N
= 50/60 Hz
(7% rdg.+100d)
(4% rdg.+100d)
100 ... 999 mA 1 mA (6% rdg.+12d) (3% rdg.+12d)
1.0 ... 9.99 A 0.01 A 100 mV/A 0.3 ... 10 A
3
(6% rdg.+12d) (3% rdg.+12d)
10.0 ... 99.9 A 0.1 A 10 mV/A 3 ... 100 A
3
(5% rdg.+11d) (2% rdg.+11d)
Uez
0.0 ... 99.0 mV 0.1 mV
400 k 1000 mV
(3% rdg. + 2 d) (2% rdg. + 2 d)
100 ... 999 mV 1 mV (3% rdg. + 1 d) (2% rdg. + 1 d)
Func-
tion
Measured
Quantity
Display Range
Reso-
lution
Input
Impedance /
Test Cu rrent
Measuring
Range
Nominal
Values
Measuring
Uncertainty
Intrinsic
Uncertainty
Connections
Plug
Insert
1
2-Pole
Adapter
3-Pole
Adapter
Probe
Clamp
WZ12C Z3512A
MFLEX
P300
Func-
tion
Measured
Quantity
Display Range
Reso-
lution
Test Current/
Signal
Frequen
cy
5
Measuring Range
Measuring
Uncertainty
Intrinsic
Uncertainty
Connections
Adapter for Test Plug
Current Clamps
PRO-RE PRO-RE/2 Z3512A Z591B
RE
BAT
RE, 3-pole
0.00 ... 9.99
10.0 ... 99.9
100 ... 999
1.00 ... 9.99 k
10.0 ... 50.0 k
0.01
0.1
1
0.01 k
0.1 k
16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
1.00 ... 19.9
5.0 ... 199
50 ... 1.99 k
0.50k ... 19.9k
0.50k ... 49.9k
(10% v.M.+10D)
+ 1
(3% v.M.+5D)
+ 0,5
6
RE, 4-pole (10% rdg.+10d) (3% rdg.+5d)
RE, 4-pole
Selective
With clamp meter
0.00 ... 9.99
10.0 ... 99.9
100 ... 999
1.00 ... 9.99 k
10.0 ... 19.9 k
15
10.0 ... 49.9 k
16
0.01
0.1
1
0.01 k
0.1 k
0.1 k
16 mA/128 Hz
16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16mA/128 Hz
1.00 ... 9.99
10.0 ... 200
(15% rdg.+10d)
(20% rdg.+10d)
10
(10% rdg.+10d)
(15% rdg.+10d)
69
Soil resistivity
(p)
0.0 ... 9.9 m
100 ... 999 m
1.00 ... 9.99 km
0.1 m
1 m
0.01 k
m
16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16mA/128 Hz
100 m ... 9.99 km
12
500 m ... 9.99 km
12
5.00 km ... 9.99 km
13
5.00 km ... 9.99 km
13
5.00 km ... 9.99 km
13
(20% rdg.+10d)
11
(12% rdg.+10d)
11
6
Probe distance
d (p)
0.1 ... 999 m
RE, 2 clamps
0.00 ... 9.99
10.0 ... 99.9
100 ... 999
1.00 ... 1.99 k
0.01
0.1
1
0.01 k
30 V / 128 Hz
0.10 ... 9.99
10.0 ... 99.9
(10% rdg.+5d)
(20% rdg.+5d)
(5% rdg.+5d)
(12% rdg.+5d)
798
85

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt PROFITEST MPRO wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gossen Metrawatt PROFITEST MPRO

Gossen Metrawatt PROFITEST MPRO Bedienungsanleitung - Holländisch - 92 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info