24 GMC-I Messtechnik GmbH
10.3 High-Voltage Test (Feature F02 or SECULIFE ST HV)
Only safety class I and II devices can be tested, which can be connected
to the test socket.
The high-voltage test is performed with direct voltage. In order to comply
with requirements for alternating voltage, testing is performed with 1.5-
fold direct voltage. This multiplier is applied automatically during testing. A
selected nominal voltage of 3.5 kV thus results in a DC output voltage of
5.25 kV.
Output voltage is measured for the entire duration of the test, and its min-
imum value is determined. The minimum voltage value is indicated as the
test result. If this value is less than the pre-selected test voltage, the test is
failed.
The conversion factor must be taken into consideration for testing and
calibration of the SECULIFE ST
HV.
The device is designed such that special measures in accordance with
DIN VDE 0104 (high-voltage test) need not be observed.
This is accomplished by means of the following characteristics:
1. Continuous short-circuit current is less than 3 mA (DC).
2. Discharge energy (at 5.25 kV) is less than 350 mJ.
In order to comply with high-voltage test regulations despite the minimal
continuous short-circuit current, the charging capacitors are connected to
the test socket (L, N) with relatively low-value protective resistors. This
results in a peak short-circuit current value of approximately 5 A (at 5 kV),
which causes plainly audible and visible sparking.
Connection
High-voltage testing can only be performed via the test socket. The pro-
tective conductor at the socket is grounded during the high-voltage test.
Safety class I:
The test is conducted bet-
ween the short-circuited
mains poles L-N and the
protective conductor PE of
the test socket.
If safety class II is recog-
nized, the test is conducted
between the short-circuited
conductors L-N of the test
socket and the probe (con-
nection 4-5).
Caution: High-Voltage!
Touc h neither the test socket nor the device under test during
voltage testing!
A high-voltage of up to 5.5 kV is present at the test socket output!
A current may flow over your body, and although it does not
reach life endangering levels, the resulting shock is plainly
discernible.
Previous testing of the protective conductor is absolutely essential
for safety class I devices because if the protective conductor is in-
terrupted, the high-voltage test does not charge all dielectric ma-
terial and the scope of testing is thus insufficient.
Connection
õ Insert the mains plug from the device under test into the test socket.
õ Safety class II: Connect the probe to jacks 4 and 5.
Attention!
!
Make sure that the application parts (18) are not connected dur-
ing high-voltage testing!
Individual Test
õ Set the rotary selector switch to the Iso/HV position.
õ Select the U-HV menu with the key and acknowledge with the
key.
õ If the device under test has not yet been switched on, a message to
this effect appears at the display. Nominal voltage is then selected.
DUT
PE
L
N
SC I
V
=
0.5 ... 6 kV
DUT
L
N
SC II
V
=
0.5 ... 6 kV
Probe
24

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info