Select connection, select test standard VDE 0701-0702, classify DUT (SC I, II or III), extension cable X/– (with/without)
Visual inspection
OK?
SC I: R
PE
OK?
Short-circuit
at DUT?
Yes/No
No
Display results; save/print report
Yes
No
No
Test
OK?
Framed with dashed line:
The test is only run if it has been
activated in the initial program window
of in the Setup menu under Sequence ...
Switch line voltage to test socket
Contact all exposed, conductive parts with the probe
Start
function test
Switch DUT on
R
INS
OK?
Yes
ΔI and/or I
C
with MLV O.K.: Yes
I
L
*
OK?
Yes
* • Equivalent leakage current I
EA
• Residual current ΔI and
contact current (direct) I
B
with MLV
OK?
Enter manufacturer’s limit values
Yes
start function test
No (I
EL
not OK)
No
No (ΔI or I
C
not OK)
dev.
heating power?
No
No
Test without insulation measurement,
if there are reservations against
an insulation measurement
Protective conductor:
not available, not connected,
has no contact
I
L
O.K.: Yes