590775
68
Zoom out
Zoom in
Vorherige Seite
1/72
Nächste Seite
68 GMC-I Messtechnik GmbH
22 Maintenance - Recalibration
22.1 Housing Maintenance
No special maintenance is required for the housing. Keep outside
surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the
use of cleansers, abrasives or solvents.
22.2 Recalibration
The respective measuring task and the stress to which your measuring in-
strument is subjected affect the ageing of the components and may result
in deviations from the guaranteed accuracy.
If high measuring accuracy is required and the instrument is frequently
used in field applications, combined with transport stress and great tem-
perature fluctuations, we recommend a relatively short calibration interval
of 1 year. If your measuring instrument is mainly used in the laboratory and
indoors without being exposed to any major climatic or mechanical stress,
a calibration interval of 2-3 years is usually sufficient.
During recalibration* in an accredited calibration laboratory
(DIN EN ISO/IEC 17025) the deviations of your instrument in relation to
traceable standards are measured and documented. The deviations de-
termined in the process are used for correction of the readings during
subsequent application.
We are pleased to perform DAkkS or factory calibrations for you in our cal-
ibration laboratory. Please visit our website at www.gossenmetrawatt.com
( Company DAkkS Calibration Center or FAQs Calibration
questions and answers).
By having your measuring instrument calibrated regularly, you fulfill the re-
quirements of a quality management system per DIN EN ISO 9001.
Standards DIN VDE 0701-0702 and IEC 63353 (VDE 0751) stipulate that
only measuring instruments which are regularly tested and calibrated may
be used for testing.
* Verification of specifications or adjustment services are not part of the calibration.
For products from our factory, however, any necessary adjustment is frequently
performed and the observance of the relevant specification is confirmed.
22.3 Safety Checks
Check the safety of your test instrument on a regular basis. We recom-
mend the same test intervals as for recalibration.
The SECUTEST... is designed as a totally insulated instrument in accor-
dance with standards IEC 61010 and VDE 0404. The protective conduc-
tor is only used for measurement purposes and is therefore not accessible
in idle state. A protective conductor test at the test socket can be per-
formed as follows:
Ð Connect the SECUTEST... to a multiple distribution box.
Ð Conduct a contact current measurement for permanently installed
DUTs (the test socket may not be connected to any load).
Ð Measure the protective conductor resistance between the neighbour-
ing socket of the multiple distribution box and the test socket.
Ð The measured value may not exceed 0.3 Ω.
Insulation resistance between LN and PE in the SECUTEST... equals ap-
prox. 150 kΩ for metrological reasons.
This must be taken into account when performing the safety checks, i.e.
the protective conductor current measurement must produce a value
below 3.5 mA, instead of insulation resistance measurement (if the equiv-
alent leakage current measurement method has been chosen, the value
must remain below 7 mA).
Apart from this, the SECUTEST... features 3 exposed conductive parts for
which contact current measurement must produce a value below 0.5 mA:
RS232 interface
metallized start key
protective conductor clip at the test socket.
68

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info