58 GMC-I Messtechnik GmbH
16 Storing in (P)SI Module (Accessory) and Database Operations
(Feature KB01)
16.1 Storing Measurement Data in the (P)SI Module
Upon completion of a measurement – message „Test passed/failed“ is
displayed – you can file the measurement data in the memory of the (P)SI
module.
õ Press the STORE key at the (P)SI module.
A text entry field is displayed.
õ You may now enter your comments on the measurement
and/or an ident. no.
õ Press the STORE key once more in order to save the measurement data
and your comments.
The message: „Data are being stored“ is shown on the display.
A detailed description is included in the operating instructions for your
(P)SI module in the chapter „Display, print and store protocol“.
16.2 Database Operations
16.2.1 Storing Test Results to the Test Instrument
If no (P)SI module has been connected, up to 125 reports can be stored
to the test instrument (without function test values and without comments
on DUT). The reports can be viewed as required at the instrument and can
be printed out, for example with the help of a terminal program, see chap-
ter 18.
The reports are sorted by time and date and are displayed with the ID
number. If no ID number was assigned, date and time are automatically
saved instead.
As an alternative, consecutive numbering can be selected.
16.2.2 Uploading Report Templates into Test Instrument, Reading Out From the
Test Instrument, Editing at the PC and Re-Saving to the Test Instrument
The „Test result“ menu allows for storage of up to 4 report templates in the
test instrument. Report templates are uploaded from a file at a PC to the
test instrument (“Load file”). Templates can also be uploaded from the test
instrument to a PC (“Templates from Secutest”), and then edited and re-
saved.
After completion of a test (“Passed” or “Failed” appears at the instru-
ment’s LCD), the test results are read out to one of these report templates
in the form of report data (depending upon the selector switch position or
test regulation) via the RS 232 interface. The report menu in the test in-
strument is activated with the and keys to this end. In order to read
out the report data at a PC, it is necessary to connect the SECUSTORE stor-
age adapter with the RS232 interface. The report can, for example, be
displayed with the PC analysis program WinProfi (as from version 3.06) or
ETC (as from version 1.22).
16.2.3Reading Out and Saving Test Results / Test Data from the (P)SI Module
Test results saved to the test instrument can be displayed, edited, printed
(only PSI module) or saved after reading them into a PC using the desired
report template.
Data can be processed immediately after completion of the test, or from
the database (Feature KB01). Data can be saved with or without a report
template (e.g. for further processing with PS3).
Test results saved to the (P)SI module can also be read out, printed (only
PSI module), saved or processed by means of a report template.
Easy method of printing out reports:
Activate the „test result“ menu in the update and options installation pro-
gram. When the test result is displayed in the test instrument select the
print function (press key „Cursor up“, then set the cursor on „print“ and
press ENTER).
17 Recognition of Probe to Protective Conductor
(Feature KD01)
The protective conductor measurement is expanded to include the
function: “automatic recognition of measuring point change”.
During protective conductor measurement, the instrument recognizes
whether or not the probe is in contact with the protective conductor, and
indicates these two possible conditions by means of acoustic signals.
This function is helpful if several protective conductor connections need to
be tested. The function can be activated in the “test sequence setup”
menu with “auto measuring point” (see “changes to the switch position
menu” above).
58

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info