Select connection, select test standard EN 61010, classify DUT (SC I, II or III), setup/sequence: R-INS-/HV test X/– (with/without)
Visula inspection
OK?
SC I: RPE
OK?
Short-circuit
at DUT?
Test
OK?
No
Yes
Display results, save / print report
Yes
No
Yes
No
No
Test
OK?
Framed with dashed line:
The test is only run if it has been
activated in the Setup menu under
Sequence...
.
Switch line voltage to test socket, start function test
Start
function test
* If it is not clear whether or not all
exposed, conductive parts are
connected to one another or to the
protective conductor, testing can be
performed in the manual mode.
Switch DUT on
Apply high voltage
Test R-INS ?
Yes
R-INS OK?
Test HV ?
Yes
HV OK?
IHL
NC/SFC
No
No
Yes/No
Yes/No
Safety Class I only: Contact all exposed, conductive parts with the
probe. The test can be repeated as often as desired for various
protective conductor parts with the manual sequence*.
Select connection, select test standard EN 61010, classify DUT (SC I, II or III), setup/sequence: R-INS-/HV test X/– (with/without)
Visula inspection
OK?
SC I: R
PE
OK?
Short-circuit
at DUT?
Test
OK?
No
Yes
Yes
No
Yes
No
No
Test
OK?
Framed with dashed line:
The test is only run if it has been
activated in the Setup menu under
Sequence...
.
Switch line voltage to test socket, start function test
Start
function test
Switch DUT on
Apply high voltage
Test R
INS
?
Yes
R
INS
OK?
Test HV ?
Yes
HV OK?
I
HL
NC/SFC
No
No
Yes/No
Yes/No
Safety Class I only: Contact all exposed, conductive parts with the
probe. The test can be repeated as often as desired for various
protective conductor parts with the manual sequence*.