56 GMC-I Messtechnik GmbH
15.13 Testing in Accordance with EN 60601 (Feature KA01)
The following leakage and auxiliary currents can be measured in accor-
dance with the above mentioned standard in the operating mode, as well
as under normal and single-fault conditions:
Protective conductor measurement R
PE
,
test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
Earth leakage current I
PE
Housing leakage current I
HL
Patient leakage current I
PL
(with nominal voltage at application part)
Patient auxiliary current I
PA
Leakage current is converted to reference voltage (see limit values chapter
8 on page 18). Reference voltage must be adapted to the supply voltage
range.
Check connection parameters and start test.
To
Test Socket This is the default setting. For other connection types, please
refer to chapter 7 on page 17.
Class If the DUT has been connected to the test socket, safety
class recognition is performed (SC I or SC II). In all other
cases, or if it is not clear whether or not all exposed, conduc-
tive parts are connected to one another or to the protective
conductor, the safety class can be selected manually.
Type The DUT type can be selected from a list.
If the limit values of the 3rd issue are to be taken into account,
select a DUT type ending with ... 3rd under this parameter.
Cond. Various test conditions can be activated here, including
insulation resistance measurement.
App PrtSee below and on page 54.
ID-No. See parameters database in chapter 15.2 on page 37.
Setup Refer to chapter 15.2 on page 37 regarding setup of the
measuring sequence.
Configure Application Parts
Entry is made here indicating whether or not the application parts are to
be tested. Jacks A through K (for connection of cables or probes) can
also be assigned to groups (application parts) for mutual testing.
Selecting Preset Test Combinations
õ Use the cursor to select test combinations with groups of 1, 2, 5 or 10
application parts and acknowledge with the key.
The groups are assigned automatically after the application parts have
been selected.
Selecting User Defined Test Combinations
õ Select the respective application part from the JACK column with
the cursor and acknowledge with the key. You can set up a user
defined group for each application part which includes 1 to 10
application parts with the cursor keys in the GR column (GRoup).
Acknowledge your settings with the key.
As long as at least one group number has been entered, the application
parts test is preset in the “To Socket” window.
The application part type with the strictest limit value takes precedence
in the initial window. All groups are set to this type. Various types can be
assigned to the groups with “direct print” (adjustable option in the setup
menu).
If testing is not to be performed with application parts, the group assign-
ment must be cancelled with the “clear” function.
The TYPE is filled in automatically if the safety class has already been
entered to the “To Socket” window.
Flowchart, see next page.
56

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECUTEST SIII wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info