571678
60
Zoom out
Zoom in
Vorherige Seite
1/72
Nächste Seite
60 GMC-I Messtechnik GmbH
19 Characteristic Values
Please refer to chapter 1.2 on page 7 to check as to which of the following measurements is required for the relevant regulation
1
As of 25 mA: shutdown by residual current measurement within 100 ms
2
Exception earth leakage current: only 0.000 3.100 mA
3
Measuring circuit is highly resistive, indication at display
4
Measurement with AC test current (Feature G00 or G01) is not possible at the
sockets (1 through 3).
Feature G01: > 25 A; Short-circuit current is less than 25 A if the SC5 special cable
is used.
5
Test duration max. 40 s, protection against overheating: measurement cannot be
restarted until a waiting period of 1 minute has elapsed.
6
Calculated value
7
AC and DC are measured for patient leakage current and patient auxiliary current.
8
The data are only valid for the values displayed at the test instrument. Data which
are transmitted via the RS232 interface may deviate.
9
only applies to earth leakage current or mains at application part
Key: of rdg. = of reading (measured value), D = digit
Measured Quantity
Meesuring Range/
Nominal Range of
Use
Resolu-
tion
Nominal
Voltage
U
N
Open-
Circuit
Voltage
U
0
Nominal
Current
I
N
Short-
Circuit
Current
I
K
Internal
Resis-
tance
R
I
Reference
Resistance
R
REF
Measuring
Uncertainty
8
Intrinsic
Uncertainty
8
Overload Capacity
Value Time
Device Protective
Conductor Resis-
tance R
PE
0.000 ... 2.100 1m
4.5 ... 9 V
DC
>200mA
DC
——
(5% rdg.+10 digits)
> 10 d
(2.5% rdg.+ 5 digits)
> 10 digits
253 V cont.
2.11 ... 31.00 10 m
0.000 ... 2.100 1m —<6VAC
>10A
AC
4
>5 s
no protection
5
Insulation Resistance
R
INS
0.050 ... 1.500 M 1k
50 ...
500 V DC
1.0 • U
N
...
1.5 • U
N
>1mA <10mA
(5% rdg.+10 digits
)
(2.5 % rdg.+5 digits)
> 10 digits
253 V cont.1.01 ... 10.00 M 10 k
10.1 ... 310.0 M 100 k
(10% rdg.+10 digits)
(10 % rdg.+10digits)
Equivalent Leakage
Current I
EL
0.00 ... 21.00 mA 10 A
230 V
– 20/
+10 %
—<3.5mA>72k
2k
(5% rdg.+10 digits)
(2.5 % rdg.+5 digits)
> 10 digits
253 V cont.
20.1 ... 120.0 mA 100
A
Contact Current
(Absence of Voltage)
I
probe
0 ... 3.500 mA 1 A— 2k
(5% rdg.+10 digits)
(2.5 % rdg.+5 digits)
> 10 digits
253 V
cont.
2
Residual Current I
RC
between L and N
0.000 ... 3.100 mA
3.00 ... 31.00 mA
1
A
10
A
——
(10% rdg.+10 digits)
> 10 digits
(5 % rdg.+5 digits)
> 10 digits
11
Equivalent Device
and/or Patient
Leakage Current
I
EDL and/or
I
EPL
0.0 ... 310.0 A0.1A
230 V
– 20/
+10 %
—<3.5mA>72k
1k
50
(5% rdg.+10 digits)
(2.5 % rdg.+5 digits)
> 10 digits
253 V
cont.
1 3
0.000 ... 2.100 mA 1 A
2.101 ... 21.00 mA 10
A
20.1 ... 120.0 mA 100
A
Leakage Current I
L
2
0.0 ... 310.0 A 100 nA
110 % of
highest line
voltage
6
——1k
(5% rdg.+10 digits)
(2.5 % rdg.+5 digit)
> 10 digit
253 V
cont.
1 3
All Leakage Current
I
L
7
0.210 ... 3.600 mA 1 A
3.10 ... > 15.00 mA
9
10 A
60

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECUTEST SIII wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info