Technical Specifications
Technical Specifications: IQAir® Cleanroom H13 NE
Power requirements
Energy consumption, 6 fan speeds*
Dimensions
Weight, system incl. filters
Air delivery, incl. filters, 6 fan speeds*
Sound pressure/power level, 6 fan speeds**
Fan motor
Control panel
Air intake
Air outlet
Color of main housing / locking arms
Housing material
Operating temperature / humidity
Electrical safety certification
Medical device registration
Total system efficiency
Air delivery (certified)
Leak tested
Surface area: 75 sq.ft. (7 m
2
)***
Surface area: 81 sq. ft. (7.5 m
2
)***
Display languages
Intelligent filter life monitor
Advanced timer
Advanced fan speed selection
Supplied accessories
Optional accessories
105.01.00.38 IQ TS CR H13 NE 120V USA 130723All technical specifications are subject to change without prior notice.
General Specifications
100-120 V, 50/60 Hz
27, 53, 92, 121, 154, 215 Watt; Standby: <1 Watt
H 32" x W 15" x D 16" (H 82 x W 38 x D 41 cm)
46 lbs. (21 kg)
45, 90, 150, 200, 250, 360 cfm (75, 150, 260, 340, 430, 610 m
3
/h)
L
P
25, 36, 44, 50, 54, 59 dB(A); L
W
35, 46, 54, 60, 64, 69 dB(A)
centrifugal, backward curved, with thermal protector, non-stop use approved
4-key touch-pad with 16 character 2-line LCD display
dual arches at base of system
320° EvenFlow™ diffuser
light gray / green
non off-gassing, impact-resistant, UV-stabilized ABS
41° to 104° F (5° to 40°C) / 5% to 95% non-condensing
CSA rated for continuous operation under applicable UL and CSA standards, BSMI
Efficiency: ≥ 55% at ≥ 0.3 microns (EN 779 class F8)
Average filter life: approx. 2 ½ years (based on average daily usage of 10h on speed 3)****
U.S. FDA-registered as a "Class II Medical Device"; Registration No.: 2087558
Performance
≥ 99.97% for particles ≥ 0.3 microns (indiv. tested at highest fan speed), ≥ 99.5% at ≥ 0.003 microns
yes (individually tested with all filters installed)
yes
EN 1822 classification
HEPA class H13 - MPPS efficiency: ≥ 99.95% @ 0.22 microns at airflow rate ≤ 112 cfm (190 m
3
/h)
HEPA class H12 - MPPS efficiency: ≥ 99.50% @ 0.16 microns at airflow rate ≤ 280 cfm (475 m
3
/h)
CR HyperHEPA® Filter H13 (XL) A
Purpose: control of fine & ultra-fine particulate matter
Media: antimicrobial glass microfiber, medical-grade HyperHEPA® filter, non-offgasing separators
Efficiency: ≥ 99.97% at ≥ 0.3 microns (EN 1822 class H13)
Average filter life: approx. 4 ½ years (based on average daily usage of 10h on speed 3)****
Filter Configuration
Pre-filter
CR PreMax™ Filter F8 (XL) A
Purpose: control of coarse and fine particulate matter; protection of subsequent filters
Media: antimicrobial glass microfiber, mini-pleated for high-capacity, non off-gassing separators
FlexVac & VM FlexVac mobile & wall-mounted source capture kits, InFlow & OutFlow
positive/negative pressure duct kits, Mobility 56 kit, PF40 coarse dust pre-filter,
VMF wall-mount bracket.
* per fan speed measured at 120V, 60Hz; tolerance: ± 10% (± 6 cfm); ** tolerance: ± 3 dB(A); *** tolerance: ± 10% **** filter life can vary based on actual usage and pollution levels
Features
4 user-selectable languages: English, French, German, Spanish
yes (monitors actual usage of each individual filter), 3 filter life status LEDs
yes (allows programming of operating hours and weekdays)
yes (allows programming of different fan speeds for 2 different time periods)
remote control (incl. battery), power cord with plug, anti-tampering toolkit, certificate of performance
Main particle filter
26

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für IQAir Cleanroom H13 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info