598220
66
Zoom out
Zoom in
Vorherige Seite
1/476
Nächste Seite
Front Panel Operation
3-34 2001-900-01 Rev. K/ August 2010
A procedure to measure in-circuit current follows:
1. Select the in-circuit current measurement mode from the
CONFIGURE DCI menu and place the instrument in
the DCI function. Note that the 4W annunciator lights to
indicate this is a 4-wire measurement.
2. Connect a set of Kelvin test probes, such as Keithley
Model 5805 or 5806, to the Model 2001 INPUT HI and
LO terminals and SENSE HI and LO terminals.
3. For R
TRACE
<50m, or where the conductors are physi-
cally hot, rel out (zero correct) any thermal offsets that
are present before measuring in-circuit current. With
power to the test circuit removed, place the probes on
the desired trace and enable rel from the DCI function.
4. Turn on the power to the test circuit and read the in-cir-
cuit current calculation. The speed of this measurement
is four readings per second at 1PLC. Its range is fixed at
12A. The default filter for in-circuit current is a moving
average of ten readings. Additional filtering may be
needed at low voltage levels.
COUPLING
This parameter selects the input coupling for the ACI
function. When AC coupling is selected, a DC blocking
capacitor is placed in series with the AC measurement
circuit. (Note that the current shunt resistors are always DC
coupled to the inputs.) This removes the DC component
from the RMS and average ACI measurement.
When AC+DC coupling is selected, the blocking capacitor is
removed. Subsequent RMS or average ACI measurements
will reflect both the AC and DC components of the signal.
You can set the ACI coupling as follows:
1. From the CONFIGURE ACI menu, select COUPLING
and press ENTER. The following menu is displayed:
SET AC COUPLING
AC AC+DC
2. Highlight the desired coupling and press ENTER.
Note that the coupling settings for ACI and ACV are
discrete. Thus, setting the coupling of ACI has no effect on
the coupling of ACV.
AC-TYPE
This parameter selects the measurement type for the ACI
function. The Model 2001 directly measures RMS and
average AC current. You can program the ACI type
parameter as follows:
1. From the CONFIGURE ACI menu, select AC-TYPE
and press ENTER. The following menu is displayed:
SET ACI MEASUREMENT
RMS AVERAGE
2. Highlight the desired measurement type and press EN-
TER.
With this parameter selected, the instrument performs
RMS AC current measurements.
When this item is selected, the signal path in
the instrument bypasses the RMS converter, so the average
ACI measurement is just the filtered output of a full wave
rectifier.
Multiple displays
The multiple displays for AC current that show multiple
functions are shown in Figure 3-13. There are no
multifunction displays for DC current.
66

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Keithley 2001 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info