MI 3152(H) EurotestXC (2,5 kV) General description
9
1.1.5 Notes related to measurement functions
Insulation resistance
The measuring range is decreased if using Plug commander.
If a voltage of higher than 30 V (AC or DC) is detected between test terminals, the
measurement will not be performed.
Diagnostic test
If any insulation resistance values (RISO(15 s) or RISO(60 s )) are over-ranged the DAR
factor is not calculated. The result field is blank: DAR:_____!
If any insulation resistance values (RISO(60 s) or RISO(10 min)) are over-ranged the PI
factor is not calculated. The result field is blank: PI :_____!
R low, Continuity
If a voltage of higher than 10 V (AC or DC) is detected between test terminals, the
measurement will not be performed.
Parallel loops may influence on test results.
Earth, Earth 2 clamp, Ro
If voltage between test terminals is higher than 10 V (Earth, Earth 2 clamps) or 30 V (Ro)
the measurement will not be performed.
Contactless earthing resistance measurement (using two current clamps) enables
simple testing of individual earthing rods in large earthing system. It is especially suitable
for use in urban areas because there is usually no possibility to place the test probes.
For two clamps earth resistance measurement clamps A 1018 and A 1019 should be
used. Clamps A 1391 are not supported. The distance between clamps should be at
least 30 cm.
For specific earth resistance measurements Adaptor A 1199 should be used.
RCD t, RCD I, RCD Uc, RCD Auto
Parameters set in one function are also kept for other RCD functions!
Selective (time-delayed) RCDs have delayed response characteristics. As the contact
voltage pre-test or other RCD tests influence the time delayed RCD it takes a certain
period to recover into normal state. Therefore a time delay of 30 s is inserted before
performing trip-out test by default.
Portable RCDs (PRCD, PRCD-K and PRCD-S) are tested as general (non-delayed)
RCDs. Trip-out times, trip-out currents and contact voltage limits are equal to limits of
general (non-delayed) RCDs.
The a.c. part of MI and EV RCDs is tested as general (non-delayed) RCDs.
The d.c part of MI and EV RCDs is tested with a d.c. test current. The Pass limit is
between 0.5 and 1.0 IdNDC.
The Zs rcd function takes longer to complete but offers much better accuracy of fault
loop resistance (in comparison to the RL sub-result in Contact voltage function).
Auto test is finished without x5 tests in case of testing the RCD types A, F, B and B+ with
rated residual currents of IdN = 300 mA, 500 mA, and 1000 mA or testing the RCD type
AC with rated residual current of IdN = 1000 mA . In this case Auto test result passes if all
other results pass, and indications for x5 are omitted.
Auto test is finished without x1 tests in case of testing the RCD types B and B+ with
rated residual currents of IdN = 1000 mA. In this case Auto test result passes if all other
results pass, and indications for x1 are omitted (MI 3152 only).
Tests for sensitivity Idn(+) and Idn(-) are omitted for selective type RCD.
10

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Metrel MI 3152 - EurotestXC wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Metrel MI 3152 - EurotestXC

Metrel MI 3152 - EurotestXC Bedienungsanleitung - Deutsch - 213 seiten

Metrel MI 3152 - EurotestXC Bedienungsanleitung - Holländisch - 85 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info