Onyx Vantage 9590 Care, Maintenance, and Cleaning
The advanced digital circuitry within the device requires no calibration or periodic maintenance other than battery
replacement. The device’s expected service life is 5 years. Field repair of the 9590 circuitry is not possible. Do not attempt
to open the case or repair the electronics. Opening the case will damage the device and void the warranty. Do not open the
9590 more than 90°, and do not twist or pull on the device when cleaning.
Cleaning the Onyx Vantage 9590
CAUTIONS:
!
Clean the device before applying it to a patient.
Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids onto the device.
Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium chloride or isopropyl alcohol.
1. To clean, wipe the surfaces with a soft cloth dampened with mild detergent or a 10% bleach solution or equivalent.
(household bleach [5.25% sodium hypochlorite]). Do not use undiluted bleach or any cleaning solution other than those
recommended here, as permanent damage could result.
2. Dry with a soft cloth, or allow to air dry. Ensure that all surfaces are completely dry.
Equipment Response Time
If the signal from the sensor is inadequate, the last measured SpO
2
and pulse rate values
freeze for 10 seconds and are then replaced with dashes.
Example: SpO
2
Exponential Averaging
SpO
2
decreases 0.75% per second; pulse rate = 75 BPM
The response of the 4-beat average is 1.5 seconds.
Testing Summary
SpO
2
accuracy and low perfusion testing was conducted by Nonin Medical, Inc. as described below.
SpO
2
Accuracy Testing
At an independent research laboratory, SpO
2
accuracy testing is conducted during induced hypoxia studies on healthy, male
and female, non-smoking, light-to-dark-skinned subjects that are aged 18 years and older. The measured arterial hemoglobin
saturation value (SpO
2
) of the device is compared to arterial hemoglobin oxygen (SaO
2
) value, determined from blood
samples with a laboratory co-oximeter. The accuracy of the device is in comparison to the co-oximeter samples measured
over the SpO
2
range of 70–100%. Accuracy data is calculated using the root-mean-squared (A
rms
value) for all subjects, per
ISO 80601-2-61 and ISO 9919, Standard Specification for Pulse Oximeters for Accuracy.
Low Perfusion Testing
This test uses an SpO
2
Simulator to provide a simulated pulse rate, with adjustable amplitude settings of various SpO
2
levels.
The device must maintain accuracy in accordance with ISO 80601-2-61 and ISO 9919 for pulse rate and SpO
2
at the lowest
obtainable pulse amplitude (0.3% modulation).
Performance in Motion
Motion artifact simulation introduced by a pulse oximeter tester determines whether the oximeter meets the criteria of
ISO 80601-2-61 and ISO 9919 for pulse rate during simulated movement, tremor, and spike motions.
Contact regulatory@nonin.com for more information regarding motion testing.
Principles of Operation
Pulse oximetry is a non-invasive method that passes red and infrared light through perfused tissue and detects the fluctuating
signals caused by arterial pulses. Well-oxygenated blood is bright red, while poorly oxygenated blood is dark red. The pulse
oximeter determines functional oxygen saturation of arterial hemoglobin (SpO
2
) from this color difference by measuring the
ratio of absorbed red and infrared light as volume fluctuates with each pulse.
SpO
2
Values Average Latency
Standard/Fast Averages SpO
2
4 beat exponential 2 beats
Pulse Rate Values Response Latency
Standard/Fast Averages Pulse Rate 4 beat exponential 2 beats
Equipment Delays Delay
Display Update Delay 1.5 seconds
4

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Nonin Onyx Vantage 9590 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Nonin Onyx Vantage 9590

Nonin Onyx Vantage 9590 Bedienungsanleitung - Deutsch - 9 seiten

Nonin Onyx Vantage 9590 Bedienungsanleitung - Holländisch - 9 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info