T echnical data
Name: COMP AIR
Model: NE-C28P (NE-C28P-E)
T ype: Compressor Nebulizer
Rating: 220-240V , 50/60Hz
Power Consumption: 192V A
Operating T emperature/
Humidity:
+10°C to +40°C / 30% to 85%RH
Storage and T ransport
T emperature/Humidity/
Air Pressure:
-20°C to +60°C / 10% to 95%RH /
700 - 1060hPa
Weight: Approx. 1.9kg (compressor only)
Dimensions: Approx. 170 (W) × 103 (H) ×182 (D) mm
Contents:
Compressor , Nebulizer Kit, Air T ube (PVC,
200cm), Mouthpiece, Nosepiece, Adult Mask
(PVC), Child Mask (PVC), 5 pcs Spare Air
Filters, Carring Bag, Instruction Manual,
W arranty Card.
Classification: Class ll equipment, T ype B applied part
= Class ll
equipment
= T ype B
applied part
Read the instruction
manual carefully
=
Power off
=
Power on
T echnical data for the OMRON NE-C28P Compressor with the V .V .T .
Nebulizer Kit N-01:
Particle Size: **MMAD Approx. 3 m
MMAD = Mass Median Aerodynamic Diameter
Medication T ank Capacity: 7ml maximum
Appropriate Medication
Quantities:
2ml minimum - 7ml maximum
Sound: *Noise level (at 1 m distance) 60dB
Nebulization Rate: *Approx. 0.5ml/min (by weight loss)
Aerosol Output: **0.4ml (2ml, 1%NaF)
Aerosol Output Rate: **0.09ml/min (2ml, 1%NaF)
Result of cascade impactor measurements for particle size**
*
Measured by OMRON HEAL THCARE Co., Ltd.
**
Independently measured by Prof. Dr . Hiroshi T akano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering, Doshisha
University , Kyoto, Japan, according to EN 13544-1:2007+A1:2009.
Notes:
• Subject to technical modification without prior notice.
• This OMRON product is produced under the strict quality system of OMRON
HEAL THCARE Co., Ltd., Japan.
• The device may not work if the temperature and voltage conditions are different to
those defined in the specifications.
• The device fulfils the provisions of the EC directive 93/42/EEC (Medical Device
Directive) and the European Standard EN13544-1:2007+A1:2009, Respiratory
therapy equipment - Part1: Nebulizing systems and their components.
• Performance may vary with drugs such as suspensions or high viscosity . See drug
supplier ’ s data sheet for further details.
• See website of OMRON HEAL THCARE EUROPE to update technical information.
URL: www .omron-healthcare.com
Removing condensation from the air tube
If there is moisture or liquid remaining in the air tube, be sure to follow the
procedure below to remove any moisture from within the air tube.
1) Make sure that the air tube is connected to the air connector on the
compressor .
2) Remove the air tube from the nebulizer kit.
3) T urn on the compressor and pump air through the air tube to expel the
moisture.
Changing the air filter
If the air filter has been used for more than 60 days, replace it with a new
one.
1. Pull the air filter cover off the compressor .
2. Change the air filter.
Remove the old air filter with a sharp object
and insert the new air filter .
3. Put the air filter cover back in place.
Make sure it is attached properly .
Notes:
• Use only OMRON air filters designed for this device. Do not operate without a filter .
• Do not wash or clean the filter . If the air filter becomes wet, replace it. Damp air filters
can cause blockages.
• There is no front or back orientation for the air filters.
• Check that air filters are clean and free of dust before inserting them.
Cleaning and daily disinfecting
Cleaning
Clean the parts after each use to remove residual medication. This will avoid
inefficient nebulization or infection.
■
W ashable parts
• Nebulizer Kit, Mask (PVC)/(SEBS), Mouthpiece, Nosepiece,
Air T ube (Silicone)
W ash them in warm water and mild detergent (neutral detergent). Rinse them
thoroughly with clean hot tap water and allow to air dry in a clean place.
■
Non-washable parts
• Compressor , Air T ube (PVC)
Firstly , make sure that the power plug is unplugged from the power outlet.
Wipe clean with a soft cloth moistened with water or mild detergent
(neutral detergent).
• Air Filter
Do not wash or clean the air filter . If the air filter becomes wet, replace it.
Damp air filters can cause blockages.
Disinfecting
Always disinfect the parts after the last treatment of the day . If the parts are
heavily stained, replace them with new ones.
T o select a method for disinfection, refer to the table below .
o
: applicable
X
: not applicable
Disinfecting
method
Parts
Mouthpiece
Nosepiece
Nebulizer Kit
Air Tube
(Silicone, 100cm)
(optional)
Adult Mask (SEBS)
(optional)
Child Mask (SEBS)
(optional)
Adult Mask (PVC)
Child Mask (PVC)
Air Tube
(PVC, 200cm)
Air Filter
Materials
PP
PP
PP
(V aporiser Head: PC)
Tube: Silicone
Connector: PP
Mask: SEBS
Band: Rubber
(non-latex)
Adapter: PP
Mask: PVC
(Phthalate free)
Band: Rubber
(non-latex)
PVC
(Phthalate free)
Polyester
A
Alcohol
oo o o o o
XX
Disinfecting
ethanol
Sodium
hypochlorite
oo o o o o
XX
Milton*
Quaternary
ammonium
oo o o o o
XX
Osvan*
Chlorhexidine
oo o o o o
XX
Hibitane*
Amphoteric
Surfactant
oo o o o o
XX
T ego*
B
Boiling
oo o o o
XX X
*
an example of commercially available disinfectant.
A. Use a commercially available disinfectant. Follow the instructions
provided by the disinfectant manufacturer .
Note: Never clean with benzene, thinner or a flammable chemical.
B. Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake
off excess water and allow to air dry in a clean
environment.
Caution: Handling the V aporiser Head
• Always wash the vaporiser head after each use.
• Do not use a brush or pin to clean the vaporiser head.
• When disinfecting the parts by boiling, be sure to boil them in plenty of water .
• Do not boil vaporiser head together with objects other than applicable nebulizer
accessories.
How to use
W arning:
• Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using
them when as follows:
- the first time after purchase
- if the device has not been used for a long period of time
- if more than one person uses the same device
For the instructions, refer to “Cleaning and disinfecting” section.
1. Make sure that the power switch is in the off (
) position.
2. Plug the power plug into a power outlet.
3. Remove the cap and the inhalation top from
the medication tank.
4. Add the correct amount of prescribed
medication to the medication tank.
5. Make sure that the vaporiser head is securely
attached to the inhalation top.
6. Put the inhalation top back inside the
medication tank.
7. Attach the cap and mask, mouthpiece
or nosepiece to the nebulizer kit tightly .
Note:
It is not necessary to attach the cap when
nebulizing with Adult mask. If the amount of
aerosol is too great, attach the cap.
8. Attach the air tube.
T wist the air tube plug slightly and push it firmly into
the air tube connector .
9. Hold the nebulizer kit as
indicated on the right.
Follow the instructions of your
doctor or respiratory therapist.
Caution:
Do not tilt the nebulizer kit at an
angle of greater than 45 degrees
in all directions. Medication may
flow into the mouth or it may result
in not nebulizing effectively .
10. Press the power switch to the on (
) position. As the compressor
starts, nebulization begins and aerosol is generated.
Inhale the medication.
1 1. When treatment is complete, turn the power off and unplug the
compressor from the power outlet.
T roubleshooting
Check the following if the device should fail during operation.
Problem Cause Remedy
Nothing happens when the
power switch is pressed.
Is the power plug plugged into
a power outlet?
Check that the plug is inserted
in a power outlet. Unplug then
reinsert the plug if necessary .
No nebulization or low
nebulization rate, when the
power is on.
Is there medication in the
medication tank?
Add the correct amount of
medication to the medication tank.
Is there too much/little
medication in the medication
tank?
Is the vaporiser head missing
or not assembled correctly?
Attach the vaporiser head
correctly .
Is the nebulizer kit assembled
correctly?
Assemble the nebulizer kit
correctly .
Is the nozzle blocked? Make sure that the nozzle is free
of blockages.
Is the nebulizer kit tilted at a
sharp angle?
Make sure that the nebulizer kit is
not tilted at an angle of more than
45 degrees.
Is the air tube connected
correctly?
Make sure that the air tube
is correctly connected to the
compressor and nebulizer kit.
Is the air tube folded or
damaged?
Make sure that the air tube does
not contain kinks.
Is the air tube blocked? Make sure that the air tube is free
of blockages.
Is the air filter dirty? Replace the air filter with a new
one.
The compressor is
abnormally loud.
Is the air filter cover attached
correctly?
Attach the air filter cover correctly .
The compressor is very hot. Is the compressor covered?
Do not cover the compressor with
any type of cover during use.
Are the ventilation slots
blocked?
Do not block the ventilation slots.
Note:
If the suggested remedy does not solve the problem, do not try to repair the device
- no parts of the device are user serviceable. Return the device to an authorized
OMRON retail outlet or distributor .
1
2
2
1
2
1
2
Right angle
0
10
20
30
40
50
60
70
80
90
100
100 10 1 0.1
Cumulative % particle mass of sodium fluoride undersize
Cumulative Undersize%
Particle Size Dp ( m)
Individual tests
Mean
MMAD: 3 m