CARE AND MAINTENANCE
Battery Replacement
A battery wiring diagram is printed on a decal located on
the front cover. See the specification table for correct battery
specifications.
Battery posts, terminals, and related accessories
contain lead and lead compounds. Wear goggles and
gloves when handling batteries and wash hands after
handling.
The batteries in your power chair should only be
serviced or replaced by a Quantum Rehab Provider or a
qualified technician.
Do not replace batteries when seat is occupied.
Power chair batteries are heavy. See specifications
table. If you are unable to lift that much weight, be sure
to get help. Use proper lifting techniques and avoid
lifting beyond your capacity.
Do not mix old and new batteries. Always replace both
batteries at the same time.
Keep tools and other metal objects away from the battery
terminals. Contact with tools can cause electrical shock.
To replace the batteries:
1. Turn off the power to the controller.
2. Make sure that the power chair is in drive mode. See
“Your Power Chair.
3. Remove the front cover by squeezing the release
handles (see figure 24) and lifting the front cover up
and away from the power base.
4. Disconnect the quick release battery connectors. See
figure 24.
5. Remove the batteries.
6. Disconnect the battery harness from each battery.
7. Connect the battery harness to the new front battery
according to the battery wiring diagram. See figure 24.
Make sure you tighten the fasteners so that the
connections are secure.
8. Connect the battery harness to the new rear battery
according to the battery wiring diagram. See figure 24.
9. Install the new batteries and connect the quick release
battery connectors.
10. Reinstall the front cover.
11. Charge the batteries. See “Batteries and Charging.
PROHIBITED!
When to See Your Quantum Rehab Provider for Service
The following symptoms could indicate a serious problem
with your power chair. If necessary, contact your Quantum
Rehab Provider. When calling, have the model number,
serial number, nature of the problem, and the error code if
available.
Motor noise
Frayed harnesses
Cracked or broken connectors
Uneven wear on any of the tires
Jerky motion
Pulling to one side
Bent or broken wheel assemblies
Does not power up
Powers up, but does not move
Loose seat or seating components
Corrective Maintenance
If the battery condition meter does not light up when you turn
on the power:
Check the harness connections. Make sure they are
tight.
Check the circuit breaker. Reset it if necessary.
Check the battery connections.
If the above conditions prove normal, you can load test the
batteries with a battery load tester. These testers are available
at automotive parts stores. Disconnect both batteries before
load testing and follow the directions that come with the load
tester. If either one of the batteries fails the load test, replace
both of them. If your power chair still does not power up,
contact your Quantum Rehab Provider.
WARNING!
WARNING!
MANDATORY!
25

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Quantum Q6 Edge 2.0 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Quantum Q6 Edge 2.0

Quantum Q6 Edge 2.0 Bedienungsanleitung - Deutsch - 32 seiten

Quantum Q6 Edge 2.0 Bedienungsanleitung - Holländisch - 28 seiten

Quantum Q6 Edge 2.0 Bedienungsanleitung - Holländisch - 44 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info