Model: V5
Suc tion Unit
w w w .rossmax.c om
IN3V50000000000XX
W arranty Card
This instrument is cover ed by a 2 year guarantee from the pur chase date. The
guarantee is valid only on presentation of the guarantee car d completed by the dealer
confirming purchase date or the receipt. Suction unit accessories are not included .
Opening or altering the instrument invalidates the guarantee . The guarantee does
not cover damage, accidents or noncompliance with the instruction manual. P lease
contact your local seller/dealer or www.rossmax.com.
Cust omer Name: ____________________________________________________
Address: ___________________________________________________________
T elephone: _________________________________________________________
E-mail address: ______________________________________________________
Product Information
Date of purchase: ___________________________________________________
Store wher e purchased:
____________________________________________________________________
Gasket (12a) Silicone O O X
Over flow valve (13) PP O O X
Rubber valve (13a) Silicone O O X
Float (13a) PP O O X
T ubing depository
(9) PC O O X
Connecting tubing
(Ø7mm,180cm) (19) PVC O X X
Connecting tubing
(Ø8mm,14cm) (17) Silicone O O X
O: Applicable, X: Not applicable
Note:
1. The collection canister and tubing after disinfecting should be dated and stor ed
in clean and dr y areas to a void recontamination.
2. Do not use any cleaners or disinfectants that contain ammonia, benzene and/or
acetone to clean the unit.
3. Dispose of the solution following the instructions pro vided by the disinfectant
manufacturer .
4. Dispose of the suction catheter in accordance with the local laws and r egulations.
Collection canister – 1400 ML
Filt er
Silicone and PV C tubing
Maintenance
Rules for returning and r epairing
Microbial contamina tion
Ac cessories (optional)
T echnical Specications
T r oubleshooting
Electromagnetic Compatibility Informa tion
Symbols
The collection canister consists of an overow valve, a bottle in PC (polycarbonate)
material, a lid with switch button on handle and must be r eplaced if there are
visible signs of the damage.
Note:
1. K eep the device vertical to allow the over ow valve t o function cor rectly .
2. Do not overtur n the collection canister during use, in order to pr event the
inter vention of the ov er ow valve; should this occur , switch the aspirat or o
and detach the tubing connected to the antibacterial lter .
3. T o accelerate aspiration and simplify cleaning, it is recommended t o put water
in the collection canister before using .
4. Suction at the patient is automatically obstructed when the liquid level r eaches
the overow valve located on the underside of the canister lid.
5. Contents of the container should be emptied when automatic o ver ow valve
is activated. F urther suc tioning could cause damage to the vacuum pump .
The antibacterial lter supplied with Rossmax aspirators must be strictly replaced
for each new patient, or if ov er ow occurs or the lter is saturated. Should
the aspirator be used on the same patient, the lter f or single -patient use is
recommended to be r eplaced after ever y application. I t should also be replaced
if the unit is used with a patient whose risk of contamination is unk nown. The
lter cannot be sanitized , disinfected, or sterilized .
Note:
1. Do not use the aspirator without the antibacterial lter , the unit could be
damaged. F rom a bacteriological point of view , it becomes dangerous f or the
patient.
2. Do not substitute any other material for this antibacterial lter . Substitution
may lead to contamination or poor performance; use only Rossmax lters.
The tubing is recommended t o be single use, but the number of cycles of
disinfection and/or cleaning is strictly link ed to the employment of the said tubing .
Theref ore, after each cleaning cycle, it is up to the nal user t o verify whether the
tubing is suitable for r euse. The component must be r eplaced if there are visible
signs of the decay of the tubing .
The suction equipment does not need maintenance or lubrication. I t is, how ever
to inspect the unit and accessories before each use .
DANGER:
Electric shock hazard. Do not attempt to open or r emove cabinet, ther e are no
user-serviceable internal components. I f service is required, r eturn the device to
your local distributor or authorized Rossmax ser vice center . Opening or tampering
with the unit will void warranty .
Rossmax requires the f ollowing procedur es to be carried out to protect
the instrument and the safety of all who c ome in contact with it.
Befor e returning an instrument for repair , the ex ternal sur faces and all accessories
MUST be carefully disinf ected with a cloth soak ed in methylated spirits or
hypochlorite-based solution. The instrument and accessories should then
be placed in a bag with a note outlining the disinfection undertak en. F ailur e
to follo w this procedure will r esult in the instrument being returned to the
purchaser unrepair ed. Instruments returned for r epair MUST be accompanied by
a description of the problem. Rossmax will not be responsible f or damage caused
In the presence of pathologies with the risk of infection and microbial contamination,
we recommend personal use of the accessories, collection canister , connec ting
tubing and antibacterial lter and all accessories should be thoroughly cleaned
and disinfected after each use.
Note: Use only with original Rossmax ac cessories.
Name Model No.
Connecting PVC tubing (Ø7mm-length 180cm) SU_AC_001_00
Collection canister (1400ml) SU_AC_002_00
Antibacterial lter k it (Antibacterial lter + 2 Silicone tubing
Ø8mm-length 14cm) SU_AC_004_00
Antibacterial lter SU_AC_005_00
Connecting silicone tubing (Ø8mm-length 14cm) SU_AC_006_00
T ubing depository SU_AC_008_00
AC power cor d (EU plug) SU_AD_004_00
AC power cor d (US plug) SU_AD_005_00
AC power cor d (UK plug) SU_AD_006_00
AC power cor d ( JET plug) SU_AD_007_00
Suction class: High V acuum / High Flow
P ower source: AC100-120V , 50/60Hz or AC220-240V , 50/60Hz
Max. vacuum @ pump inlet: -80 kP a/ -600 mmHg (mmHg for ref erence only)
Max air ow @ pump inlet: 30L/min
P ower Consumption: ≤ 55V A
Collection canister: 1,400 ml
W eight: Approx. 3.2 (Kgs)
Machine dimension (with-
out collection canister): 363 ( W ) × 175 (D) × 254.5 (H) mm (Incl. rubber
pad)
Operating conditions: T emperature min.: 10°C; max. 40°C;
Relative humidity : ≤90% non- condensing
Atmospheric pressure: Min. 700 hP a; max. 1060 hP a
Storage and transportation
conditions:
T emperature min.: -20°C; max. 60°C;
Relative humidity : ≤90% non- condensing
Atmospheric pressure: Min. 700 hP a; max. 1060 hP a
IP Classication: IP21: Prot ection against har mful ingr ess of water
and par ticulate matter
If, after checking the described conditions below the device is still not work ing
properly , we recommend that y ou consult your local distributor or authorized
Rossmax ser vice center .
Pr oblem P ossible cause Solution
The device does not
work
The power cor d has not
been correctly inser ted
into the socket of the
device or the power socket
Correctly inser t the power
cord into the sockets
P ower switch has not been
activated.
Check if the power switch
is turned to the “I” (ON)
position
Lack of suction
The collection canister
lid has not been correctly
positioned on the canister
Correctly position the
collection canister lid
Lid gasket (rubber O -ring)
not in place Correctly position the gasket
on the lid
Leaks or cracks in canist er
assembly Replace the canister
F ilter blocked Replace the filter
Blocked float Incrustation on the float
Remove the canister lid
and overflow valve and
extrac t the float. Continue
by carr ying out the cleaning
procedures as described in
the paragraph “Sanitization,
Disinfection operations ”
V acuum power poor
and/or non-existent
V acuum flow adjustment
knob/ vacuum regulator
completely open
F ully close the adjustment
knob and check the vacuum
power
F ilter blocked Replace the filter
Connecting tubing to
the device or collection
canister is clogged, bent,
disconnected or broken; or
suction catheter is twisted
or blocked
Check the condition of the
tubing, r eplace it if blocked
or broken and correctly
connect it as per the
Assembling P rocedures of
this manual
Canister lid overflow valve
closed or blocked Unblock the overflow valve,
keep the device upright
Dir ty , blocked or damaged
pump
T ak e the device to your local
distributor or authorized
Rossmax ser vice center
Class II device
T ype BF applied par t
Attention: check the instructions befor e using this device
+More vacuum
SN Device serial number
Risk : Electrocution.
Consequence: Death.
Do not use the device while tak ing a bath or a shower
ΙSwitch on
ΟSwitch o
Alternating current
1639 CE Mark ing medical ref . Dir 93/42/EEC amended by 2007/47/EC
Date of manufacture
Manufacturer
Attention: Consult accompan ying documents
W ARNING: The symbol on this product means that it’ s an electronic
product and following the E uropean directive 2012/19/EU the elec-
tronic products hav e to be disposed on your local recycling centre f or
safe tr eatment.
through improper use . T o avoid such damage, please r ead the instruction carefully .
Rossmax will not be responsible for damage of accessories . Accessories will be
replaced at the customer’ s expense.
OBM_IB_SU-V5_
EN_SW_ver2030
1. This device needs to be installed and put into service in accordance with the information provided
in the user manual.
2. W ARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the V5,
including cables specied by the manufacturer . Other wise, degradation of the performance of this
device could result.
If higher IMMUNIT Y TEST LEVELS than those specied in T able 9 are used, the minimum separation
distance may be lower ed. Lower minimum separation distances shall be calculat ed using the
equation specied in 8.10.
Manufacturer ’ s declaration-electromagnetic immunity
The V5 is intended for use in the electromagnetic en vironment specied below . The customer or
the user of the V5 should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic envir onment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80 MH z
6 Vrms: in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80
MHz
6 Vrms: in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the V5 including cables, than the
recommended separation distance calcu-
lated from the equation applicable to the
frequency of the transmitter .
Recommended separation distance:
d = 1,2 √P , d = 1,2 √P 80MH z to 800 MHz, d
= 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power rat -
ing of the transmitter in watts ( W ) accord-
ing to the transmitter manufacturer and d
is the recommended separation distance in
metres (m).
Inter ference ma y occur in the vicinity of
equipment marked with the following sym-
bol:
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GH z
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GH z
80 % AM at 1 kHz
NOTE1: A t 80 MH z and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines ma y not apply in all situations. Electromagnetic propagation is a ected by
absorption and reection from structures, objects and people.
More information on EMC compliance of the device can be obtained fr om Rossmax website: www.
rossmax.com.
Rossmax Swiss GmbH,
W idnauerstrasse 1, CH-9435 Heerbrugg , S witzerland