80-0067-00-MO-RevJ 2019-01-30 SunTech CT40 User Manual | 26
CAUTION: Inaccurate measurements may be caused by:
Incorrect sensor application or use. Proper sensor uses, and placement is critical for good performance.
Significant levels of dysfunctional hemoglobin. (e.g., carboxyhemoglobin or methemoglobin)
Intravascular dyes such as indocyanine green or methylene blue.
Dyes or any substance containing dyes that affect light absorption may cause erroneous readings.
Some nail polish colors (particularly dark shades) or artificial fingernails may reduce light transmission and
affect pulse oximetry accuracy. Remove any nail polish or artificial fingernails before using the SpO2 sensor.
Exposure to excessive illumination, such as surgical lamps (especially ones with a xenon light source), bilirubin
lamps, fluorescent lights, infrared heating lamps, or direct sunlight (exposure to excessive illumination can be
corrected by covering the sensor with a dark or opaque material)
Excessive patient movement.
Abnormal pulse rhythm.
Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line. Always place
the SpO2 sensor on the arm without a blood pressure cuff, arterial catheter, or intravascular line.
CAUTION: Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore may not
detect certain arrhythmias. The pulse oximeter should not be used as a replacement or substitute for ECG based
arrhythmia analysis.
CAUTION: SpO2 is empirically calibrated to functional arterial oxygen saturation in healthy adult volunteers with
normal levels of carboxyhemoglobin (COHb) and methemoglobin (MetHb). A pulse oximeter cannot measure
elevated levels of COHb or MetHb. Increases in either COHb or MetHb will affect the accuracy of the SpO2
measurement.
CAUTION: COHb levels above normal tend to increase the level of SpO2. The level of increase is approximately equal
to the amount of COHb that is present.
CAUTION: High levels of COHb may occur with a seemingly normal SpO2. When elevated levels of COHb are
suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
CAUTION: For increased MetHb, the SpO2 may be decreased by levels of MetHb of up to approximately 10% to 15%.
At higher levels of MetHb, the SpO2 may tend to read in the low to mid 80s. When elevated levels of MetHb are
suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
CAUTION: Venous congestion may cause under reading of actual arterial oxygen saturation. Therefore, assure proper
venous outflow from monitored site. Sensor should not be below heart level (e.g. sensor on hand of a patient in a bed
with arm dangling to the floor).
CAUTION: Venous pulsations may cause erroneous low readings (e.g. tricuspid value regurgitation).
CAUTION: The pulsations from intra-aortic balloon support can be additive to the pulse rate on the oximeter pulse
rate display. Be sure to verify patient's pulse rate.
CAUTION: Elevated levels of Total Bilirubin may lead to inaccurate SpO2, measurements.
CAUTION: With very low perfusion at the monitored site, the readings may read lower than core arterial oxygen
saturation.
CAUTION: Do not immerse the sensor or patient cable in water or, solvents, or cleaning solutions (The sensors and
connectors are not waterproof).
CAUTION: Loss of pulse signal can occur in any of the following situations:
26

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für SunTech CT40 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von SunTech CT40

SunTech CT40 Bedienungsanleitung - Deutsch - 49 seiten

SunTech CT40 Bedienungsanleitung - Holländisch - 52 seiten

SunTech CT40 Bedienungsanleitung - Französisch - 50 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info