BENNING ST 750 A
10. Carrying Out Manual Measurements / Tests
First of all, each test begins with an inspection of the test object. Moreover, it has to be determined
which protection class (protective measure) the test sample corresponds to and which tests / mea-
surements (VDE regulation, manufacturer's accompanying documents, depending on the type of
the test sample) have to be carried out. The order of the individual measurements (tests) has to be
observed absolutely, because the measurement has to be stopped immediately in case of a failed
test. Only after troubleshooting, the test can be repeated right from the beginning.
Order of the individual test steps:
- Visual inspection
- Protective conductor test (if the protective conductor exists)
- Insulation measurement (if technically possible)
- Measurement of leakage current, protective conductor current and / or contact current
- Moreover: safe isolation; proof of further protective measures; inscriptions; functional test;
analysis, evaluation and documentation
If it is not directly possible to isolate the test sample from the electrical systems, proceed as follows:
- visual inspection
- protective conductor resistance measurement for devices of protection class I and
- contact current measurement on accessible protective parts of devices of protection class II
as well as on accessible parts of devices of protection class I which are not connected to the
protective conductor!
The respective test steps and measuring types depend on the type of the test
sample!
Visual inspection
Important parts of the device contributing to its safety must neither be visibly damaged nor be obvi-
ously inappropriate for the device. This particularly applies to insulation and insulating parts which
areaccessibleduringrepair,modicationortestaswellasforhousingsofdeviceswithprotective
insulation (protection class II).
Examples for the application of visual inspections are:
- Melt inserts of fuses must comply with the parameters requested by the manufacturer (e.g.
nominal current intensity and fusing characteristics);
- Airlters(ifrequested)mustbebuiltin;
- Cooling apertures must not be blocked;
- Protective covers requested by the manufacturer must be used and must be attached in proper
condition and in due form;
- Coolingnsmustnotbeblocked;
- The markings provided by the manufacturer, particularly those contributing to safety (warning
symbols, characteristic data on fuse holders, switch positions on isolator switches and similar),
must be available and perfectly legible. If necessary, markings have to be corrected after modi-
cations.
- Mains connection cables (also enclosed device connecting cables) must not show any external
defects between the connecting point of the device and the connecting point of the mains;
- Strainreliefandbendingprotection(ifapplicable)ofthemainsconnectioncablemustbermly
attachedandefcient;
- There must not be any signs of overload or improper use;
- Itisnotallowedtomakeanyimproperinterventionsandmodications;
- Any contamination and corrosion affecting safety must be removed by cleaning.
At the beginning of a test (after requesting the corresponding VDE regulation and protection class),
the"VisualInspection"menuappearsonthedisplay.Conrmwith<OK> to continue with the test!
The tester is provided with a switchable test socket to contact the test sample. The additional test
sockets"L/N"areonlyconnectedin"TestSocket"mode.The"PE"testsocketisrmlyconnected
with the earthing contact of the test socket. For measurements to earth (test samples of protection
classes II) and for direct measurement of the leakage current, a measuring shunt in the milliohm
range is integrated into the cable for PE connection. Due to a correspondingly required measuring
arrangement,the"Sonde/Probe"jackisintendedformeasuring/testingtheabsenceofvoltageor
the contact current etc. according to DIN VDE 0701-0702.
For measuring device leakage current or for the patient leakage current etc. according to DIN VDE
0751,thespeciedmeasuringarrangement"MD"(DINVDE0751-1)isconnectedtothesamejack
("Sonde/Probe").
For this reason, it is absolutely necessary to preselect the correct VDE regulation
for the measurements / tests!
Connection test
Forthersttest/measurement,thetesterstartsaconnectiontesttodetermine,whetherdangerous
contact voltage is applied, whether a test sample is connected and that there is no short-circuit. After
a positive connection test, the measurement is started. In case of a negative connection test (e.g.
open switch), an information message is displayed. Another connection test will only be started
whenthe"VisualInspection"menuhasbeenpassedthroughonceagain.
Attention! Pressing the <Pause> button will stop the measurement. During the pause,
the test socket will remain live!
10.1 Protective Conductor Test (Protective Conductor Resistance) R
PE
This function is intended to measure the protective conductor resistance R
PE
(devices of protection
class I). Measurement is made with a testing current of +/- 200mA DC according to DIN VDE 0701-
0702, VDE 0751-1 or a testing current of 10A AC (formerly: DIN VDE 0701 T260). For measuring,
the test cable (probe) is connected to accessible metal parts of the test sample and to the earthing
contactofthemainsplugbymeansofthealligatorclip.Connectthemainsplugtothe"testsocket"
oftheappliancetester.Ifnecessary,usethemeasuringjacks"PE"and"L/N"whichareconnected
inparalleltothe"testsocket".
18

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Benning ST 750A wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Benning ST 750A

Benning ST 750A Bedienungsanleitung - Deutsch - 68 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info