19
9. Cleaning and storing the instrument
•
Clean your device and cuff carefully only with a slightly mois-
tened cloth.
•
Do not use detergents or solvents.
•
On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
•
When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff tube
should not have any sharp kinks
10. Rectifying faults
E
rror messages can occur when
•
you move or talk during the measurement (ERR 1),
•
the cuff tube is not properly inserted (ERR 2),
•
when changing the batteries, the start/stop button was
pressed,
•
inflation takes longer than 20 seconds (ERR 2),
•
the inflation pressure is higher than 300 mmHg (ERR 300).
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do not
move or talk. Re-insert the batteries if necessary, or else re-
place them.
11. Specifications
Model No. BM 60
Method of Measurement Oscillometric, non-invasive blood
pressure measurement on the upper
arm
Range of Measurement Pressure 0 – 300 mmHg,
pulse 40 – 160 beats/minute
Accuracy of the Systolic ± 3 mmHg / diastolic
pressure reading ± 3 mmHg / pulse ± 5 % of the dis-
played value
Measuring uncertainty Max. permissible standard deviation
according to clinical testing:
systolic: 8 mmHg / diastolic: 8 mmHg
Sets of Memory 2 x 60 memory places
Unit Dimension L 101 x W 130 x H 54 mm
Unit Weight Approximately 260 g, excluding bat-
teries
Cuff Size Cuff for upper arm circumference of
22 – 32 cm
Operating Temperature +10 °C to +40 °C, ≤ 85 % RH
Storage-Temperature -5 °C to +50 °C, ≤ 85 % RH
Power Supply 4 x 1,5 V AA batteries (alkaline type
LR6)
Battery Life For approx. 300 measurements,
depending on blood pressure level or
inflation pressure
Accessories With storage pouch, instruction
manu al, 4 pcs “AA” battery LR6
Classification Application part Type BF
Protection class Internal power supply, IPXO, no AP
or APG, continuous operation
Key to symbols Application part Type BF
Important! Read the operating
instruction!
These specification are subject to change without notice for
purpose of improvement.