Memory function
Your ExactFit™ 1 blood pressure monitor will automatically store the
latest measurement. Storing is done automatically after the completed
measurement.
To recall the reading, press and hold the power button
for at least
2 seconds while the unit is switched off. When all the display elements
are shown, release the power button and the screen will show the last
reading with an
symbol.
Battery charge indicator
Batteries almost discharged
When the batteries are approximately 75%
used, a battery warning appears. This icon
blinks 3 times first when the device is turned
on, then goes to measuring mode. The
monitor will continue to measure reliably, but
new batteries should be obtained.
Batteries discharged—replacements required
When the batteries are fully discharged,
the battery symbol will flash empty along
with 3 short beep sounds as soon as the
device is switched on. You cannot take any
further measurement and must replace
the batteries. Refer to the section called
“Inserting batteries”.
Storage and cleaning
Do not expose device to either extreme
temperatures, humidity, dust or direct
sunlight.
The cuff contains a sensitive air-tight
inflatable bladder. Handle this carefully
and avoid all types of straining through
twisting or buckling.
Clean the product with a soft, slightly
damp cloth. Stains on the cuff can be
carefully removed with a damp cloth. Do
not use gasoline, thinners or similar
solvents. Do not wash or dry clean the
cuff. Do not store the product in a place
exposed to direct sunlight or high
humidity.
Do not drop the device or treat it roughly
in any way. Avoid strong vibrations.
!
NEVER open the device! This will make
the manufacturer’s warranty invalid and
could adversely affect the devices
accuracy!
Calibration
This product has been calibrated at the time of manufacture. If used
according to the instructions for use, periodic recalibration is not
required. If at any time you question the measuring accuracy, please
contact our service representative (see last page for contact information).
What to do if ...
Error Description Potential cause and remedy
Batteries are low Get ready to replace batteries soon with
new ones.
Batteries are fully
discharged
Insert new batteries or check that
batteries are correctly inserted.
«Err 1» Signal too weak The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«Err 2» Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping the arm still.
«Err 3» No pressure in the
cuff
An adequate pressure cannot be
generated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«Err 5» Abnormal result The measuring signals are inaccurate
and no result can therefore be
displayed. Read through the checklist
for performing reliable measurements
and then repeat the measurement.*
«HI» Pulse or cuff
pressure
too high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for
5 minutes and repeat the measurement.*
«Lo» Pulse too low The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
* If this or any other problem occurs repeatedly, please consult your
doctor.
English
5

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Braun BUA5000 - ExactFit 1 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Braun BUA5000 - ExactFit 1

Braun BUA5000 - ExactFit 1 Bedienungsanleitung - Holländisch - 11 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info