G B
W ARNING
Read the instruction manual carefully before
u s i n g t h i s d e v i c e , e s p e c i a l l y t h e s a f e t y
instructions, and keep the instruction manual
for future use. Should you give this device to
another person, it is vital that you also pass on
these instructions for use.
Safety Instructions
Explanation of symbols
IMPORT ANT
Follow the instructions for use!
Non-observance of these instructions
can result in serious injury or damage
to the device.
W ARNING
These warning notes must be observed
to prevent any injury to the user.
CAUTION
These notes must be observed to prevent
any damage to the device.
NOTE
These notes give you useful additional
information on the installation or
operation.
Device classification: type BF applied part
LOT number
Manufacturer
Date of manufacture
Common factors of wrong measurements
Blood Pressure
Monitor
Instruction Manual
Please read carefully!
•
Please keep quiet for 5-10 minutes and avoid eating,
drinking alcohol, smoking, exercising and bathing
before taking a measurement. All these factors will
influence the measurement result.
Remove watches or other spruce from your wrist
before measurement.
Always measure on the same wrist (normally left).
Measurements should be taken regularly at the
same time of each day , as the blood pressure varies
even during the day .
Any effort to support the arm during measurement
may increase the measured blood pressure.
Make sure, you are in a comfortable, relaxed posi-
tion and do not move or constrict your muscles
during measurement. Use a cushion to support your
arm if necessary .
If the wrist is not at heart level, a false reading may
be obtained.
A loose or open cuff causes false readings.
With repeated measurements, blood accumulates
in the wrist, what may lead to false readings.
Consecutive blood pressure measurements should
be carried out after a 1-minute pause or after the
arm has been held up in order to allow the accu-
mulated blood to flow normally .
•
•
•
•
•
•
•
•
SAFETY NOTES FOR BA TTERIES
Do not disassemble batteries! i
Replace all batteries if the unit display shows the low battery symbol. i
Never leave any low battery in the battery compartment since it may leak and
cause damage to the unit. i
Increased risk of leakage! Avoid contact with skin, eyes and mucous mem-
branes! If battery acid comes in contact with any of these parts, rinse the
affected area with copious amounts of fresh water and seek medical attention
immediately! i
If a battery has been swallowed, seek medical attention immediately! i
Replace all of the batteries simultaneously! i
Only replace with batteries of the same type, never use dif ferent types of
batteries together or used batteries with new ones! i
Insert the batteries correctly , observing the polarity! i
Remove the batteries from the device if it is not going to be used for an exten-
ded period of 3 months or more! i
Keep batteries out of children's reach! i
Do not attempt to recharge batteries! There is a danger of explosion! i
Do not short circuit! There is a danger of explosion! i
Do not throw into a fire! There is a danger of explosion! i
Do not throw used batteries into the household refuse; put them in a
hazardous waste container or take them to a battery collection point, at the
shop where they were purchased!
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Blood pressure classification WHO
These values have been specified by the World Health Organisation (WHO)
without making any allowance for age.
Low blood pressure systolic <100 diastolic <60
Normal blood pressure (green display area / )
systolic 100 - 139 diastolic 60 - 89
Forms of high blood pressure / hypertension
Mild hypertension (yellow display area / )
systolic 140 – 159 diastolic 90 – 99
Moderate hypertension (orange display area / )
systolic 160 – 179 diastolic 100 – 109
Severe hypertension (red display area / )
systolic ≥ 180 diastolic ≥ 1 10
Blood pressure that is too low represents just as great a health
risk as blood pressure that is too high! Fits of dizziness may
lead to dangerous situations arising (e.g. on stairs or in traffic)!
Influencing and evaluating readings
•
Measure your blood pressure several times, then record and compare the
results. Do not draw any conclusions from a single reading. i
Y our blood pressure readings should always be evaluated by a doctor who is
also familiar with your personal medical history . When using the unit regularly
and recording the values for your doctor, you should visit the doctor from time
to time to keep him updated. i
When taking readings, remember that the daily values are influenced by se-
veral factors. Smoking, consumption of alcohol, drugs and physical exertion
influence the measured values in various ways. i
Measure your blood pressure before meals. i
Before taking readings, allow yourself at least 5-10 minutes rest. i
If the systolic and diastolic readings seem unusual (too high or too low) on
several occasions, despite correct use of the unit, please inform your doctor .
This also applies to the rare occasions when an irregular or very weak pulse
prevents you from taking readings.
•
•
•
•
•
Starting up
Insert / replace batteries
W ARNING
Do not take any therapeutic measures on the basis of a self
measurement. Never alter the dose of a medicine prescribed
by a doctor .
5.
6.
The readings are automatically saved in the selected memory ( or ).
Up to 60 measured values with date and time can be stored in each memo-
ry . i
The measurement readings remain on the display . The unit switches off
automatically after approx. 2 minutes if no buttons are pressed. The unit
may also be switched off using the - button .
Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for any reason
(eg. the patient feels unwell) the - button can be pressed at any time.
The device immediately decrease the cuff pressure automatically .
Display stored values
This unit features 2 separate memories, each with a capacity of 60 memory
slots. Results are automatically stored in the memory selected by the user .
Press the - button when power off to power on the device. T o call up the
measured values stored, press the - button to select the user ( or )
and then the +/M - button . The average values of the last 3 measurements
appear on the display , together with "A VG" . Pressing the +/M - button i
again, the last measurement value will be shown. Repeatedly pressing the
+/M - button displays the respective values measured previously . If you have
reached the final entry and do not press a button, the unit automatically switch-
es itself off after approximately 2 min. in memory recall mode. Memory recall
mode can be exited at any time by pressing the - button , which will also
switch the unit off. The oldest value is deleted if 60 measured values have
been stored in the memory and a new value is saved.
Delete saved values
If you want to delete the record of a saved measurement, first follow the ins-
tructions under "Display stored values " to display the respective record. Now
press and hold the +/M - button , until CL and "-- " appears. Press the -
button to confirm the deletion. i
If you are sure that you want to permanently remove all stored memories,
follow the instructions under "Display stored values " to display the average
values of last 3 measurements ("A VG" lights up). Press and hold the +/M -
button , until CL and "-- " appears. Press the - button to confirm the
deletion. After each successful deletion CL and "00 " appears on the display .
Error messages and error remedying
Error indicators
The following symbol will appear on the display when measuring abnormal:
Please contact your supplier or the service centre in case of a claim under the
warranty . If you have to return the unit, please enclose a copy of your receipt and
state what the defect is.
The following warranty terms apply:
1. The warranty period for ecomed products is two years from date of purchase.
In case of a warranty claim, the date of purchase has to be proven by means
of the sales receipt or invoice.
2. Defects in material or workmanship will be removed free of charge within the
warranty period.
3. Repairs under warranty do not extend the warranty period either for the unit
or for the replacement parts.
4. The following is excluded under the warranty:
a. All damage which has arisen due to improper treatment, e.g.
nonobservance of the user instructions.
b. All damage which is due to repairs or tampering by the customer or
unauthorised third parties.
c. Damage which has arisen during transport from the manufacturer to the
consumer or during transport to the service centre.
d. Accessories which are subject to normal wear and tear as batteries etc..
5. Liability for direct or indirect consequential losses caused by the unit are
excluded even if the damage to the unit is accepted as a warranty claim.
For service, accessories and spare parts, please contact:
Guidelines / Standards
This blood pressure monitor meets the requirements of the EU standard for
non-invasive blood pressure monitors. It is certified in accordance with EC
Guidelines and carries the CE symbol (conformity symbol) “CE 0297”. i
The blood pressure monitor corresponds to European standards EN 1060-1 and
EN 1060-3. i
The specifications of EU Guideline “93/42/EEC of the Council Directive dated 14
June 1993 concerning medical devices” are met. i
Electromagnetic compatibility: The device complies with the EN 60601-1-2
standard for electromagnetic compatibility . Y ou can find details on this measure-
ment data on the seperate enclosure.
W arranty and repair terms
MEDISANA AG, 41468 NEUSS, GERMANY .
ECOMED is a registered trademark of MEDISANA AG.
MEDHEAL TH SUPPLIES L TD.
Unit 6, Dock offices
Surrey Quays Road
London
SE16 2XU
T el.: + 44 / 207 - 237 88 99
Fax: + 44 / 207 - 231 13 00
eMail: info@medhealthcare.co.uk
Internet: www .medisana.co.uk
GB Unit and LC-Display
2 x LR03 , 1.5 V , A AA
approx. 1 cm
1
2
3
4
a
b
c
Proper use
Contraindications
•
•
The device is not suitable for measuring the
blood pressure of children. Ask your doctor
before using it on older children.
Persons who suffer from arrhythmia, diabetes,
cardiovascular problems or who have had a stroke
should consult their doctor before using the device.
•
This fully automatic electronic blood pressure
monitor is intended for measuring the blood
pressure at home. It is a non-invasive blood
pressure measurement system to measure the
diastolic and systolic blood pressure and pulse of
an adult using a non-invasive technique by means
of a cuff, which needs to be fitted on the wrist.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
The unit is intended only for use in the home. Consult your doctor prior to
using the blood pressure monitor if you are concerned about health matters.
Only use the unit for its intended purpose in accordance with the instructions
for use. i
W arranty claims become void if the unit is misused. i
Please consult your doctor prior to using the unit if you suffer from illnesses
such as arterial occlusive disease. i
The unit may not be used to check the heart rate of a pacemaker . i
Pregnant women should take the necessary precautions and pay attention to
their individual circumstances. Please consult your doctor if necessary . i
If you experience discomfort during a measurement, such as pain in the wrist
or other complaints, press the - button to release the air immediately
from the cuff. Loosen the cuff and remove it from your wrist. i
Blood pressure measurements in high repetition rates may lead to unwanted
side effects, e.g. nerval compression or blood clots. i
Blood pressure measurements - especially if executed frequently - may lead to
temporary marks on the skin. These marks may be visible for several days in
individual cases. If necessary , contact your physician to get more information.
This device is not designed to be used by persons with limited physical, sen-
sory or mental abilities, or by persons with insufficient experience and/or know-
ledge, unless under observation by a person responsible for their safety , or
unless they have been instructed in the use of the device. i
This device is intended to be used by adults. The unit is unsuitable for children
or babies. Contact your doctor , if you want to use this device on young persons.
Do not allow children to use the unit. Medical products are not toys! i
Please ensure that the unit is kept away from the reach of children. i
The swallowing of small parts like packaging bag, battery , battery cover and
so on may cause suffocation. i
The user needs to check that the equipment functions properly and that it is in
proper working condition before being used. i
The device may not be operated in rooms with high levels of radiation or in
the vicinity of high-radiation devices, such as radio transmitters, mobile tele-
phones or microwaves, as this could cause functional defects or incorrect
measurements. i
Do not use the equipment where flammable gas (such as anaesthetic gas,
oxygen or hydrogen) or flammable liquids (such as alcohol) are present. i
Do not take measurements, if any other equipment is applied to the same limb
for monitoring. This could lead to disturbances or failures. i
Never apply the cuff over hurt skin or in case of an inserted catheter . i
Before applying the cuff, make sure there are no wrinkles in the cuff, as these
could cause bruises. i
No modification of this unit/device is allowed. i
Please do not attempt to repair the unit yourself in the event of malfunctions,
since this will void all warranty claims. Only have repairs carried out by
authorised service centres. i
Protect the unit against moisture. If liquid does manage to penetrate into the
unit, remove the batteries immediately and do not continue to use the unit. In
this case contact your specialist dealer or notify us directly . i
If the device is shared with other persons, the cuff should be disinfected before
each use. T o disinfect, UV light can be used (in this case the LC-display needs
to be protected) or you can disinfect with mild soapy water . Before use, all
parts of the device have to be completely dry . i
Please do not use a dilution agent, alcohol or petrol to clean the unit. i
Please do not hit the device heavily or let if fall down from a high place. i
If the unit is not going to be used for a long period, please remove the batteries.
•
•
•
•
•
•
•
Unit and LCD display
Lid of the Battery Compartment Blood Pressure Indicator (green -
yellow - orange - red) - button (ST ART/ST OP) LC-display
- button (Clock/Date) - button (User Setting) +/M - button
(Memory Recall) Wrist cuf f Change Battery symbol
Memory Location number Average value Display of Systolic
Pressure Display of Diastolic Pressure Display of Pulse Rate
Pulse symbol & Arrhythmia display (arrhythmia cordis) User memory 2
User memory 1 Inflation / Deflation symbol Blood Pressure Indi-
cator Display of T ime Display of Date
Scope of supply
Please check first of all that the unit is complete.
The following parts are included as standard:
• 1 ecomed blood pressure monitor BW-82E
• 2 batteries (type AAA, LR03) 1.5 V
• 1 instruction manual
Please contact your supplier if you notice any transport damage on unpacking
the unit.
What is blood pressure?
Blood pressure is the pressure created in the blood vessels by each heart beat.
When the heart contracts (= systole) and pumps blood into the arteries, it brings
about a rise in pressure. The highest value is known as the systolic pressure
and is the first value to be measured. When the cardiac muscle relaxes to take
in fresh blood, the pressure in the arteries also falls. Once the arteries are re-
laxed, a second reading is taken – the diastolic pressure.
How is blood pressure measured?
The ecomed BW-82E is a blood pressure unit which is used to measure blood
pressure from the wrist. The measurement is carried out by a microprocessor ,
which, via a pressure sensor , measures the vibrations resulting from the infla-
tion and deflation of the cuff over the artery .
1. Adjustment of the year:
2. Month / date setting:
The month figure starts to flash for setting. Keep pressing the +/M - button i
until the required month appears. Press - button to confirm the month
setting. Afterwards, the day needs to be set. As with the month setting, keep
pressing the +/M - button until the required day appears. Press - button
to confirm the day setting. Afterwards, you will enter the time setting mode.
3. Time setting:
Y ou must insert the batteries provided before you can use your unit. The lid of the
battery compartment is located on the left hand side of the unit. Open it by
pushing the cover outward and remove it. Insert the 2 AAA LR03 type 1.5 V batter-
ies supplied. Ensure correct polarity when inserting the batteries (as marked on
the battery compartment cover). i
Close the battery compartment. Replace the batteries immediately , if the partial
change battery symbol flashes on the display and "E6 " is appearing
at the same time or if nothing appears on the display after the unit has been
switched on. If the partial change battery symbol appears on the display
constantly together with "E6 ", the batteries are already in poor condition. If the
full change battery symbol appears, the batteries are in good condition.
Settings
After inserting the batteries, time and date need to be adjusted correctly . The
device automatically switches to settings mode after insertion of the batteries.
If you want to get to settings mode manually , press and hold the - button i
for approx. 3 seconds after switching on the unit.
The hour figure starts to flash for setting. Keep pressing the +/M - button un-
til the required hour appears. Press - button to confirm the hour setting
and proceed in the same way for minute setting. i
After this, all settings are completed. The display shows the actual time and
date. Press - button to switch off the device. i
If the batteries are changed, the set information is lost and needs to be enter-
ed again.
Setting up the user memory
The ecomed wrist blood pressure monitor BW-82E gives you the opportunity to
assign readings to one of two user memories, each of which provides 60 me-
mory slots. After switching on the device (press - button to switch the
device on), press - button to select between user 1 and user 2 .
Y our selection will be saved by the device and will be considered in next blood
pressure measurement respectively when the measurement result is saved.
Y our selection is valid until you make another selection. i
After selecting the user , press - button to switch off the device.
The year figure starts to flash for setting. Keep pressing the +/M - button until
the required year appears. Press - button to confirm the year setting. i
Afterwards, you will enter the month/date setting mode.
Fitting the cuff
1.
2.
3.
Attach the clean cuff to the left wrist which is free from clothing with the palm
of the hand facing upwards (fig.1 ). i
There should be a gap of approx. 1 cm between the cuff and the palm of the
hand (fig.2 ). i
Press the V elcro tape firmly around the wrist ensuring that it is not so tight that
it could falsify the measured result (fig.3 ).
Correct position for taking readings
Carry out the measurement whilst seated. i
Relax your arm and place it comfortably , on a table for instance. i
Raise the wrist so that the cuff of the monitor is at heart level (fig. 4 , position
a = too high, b = correct, c = too low). i
Keep still during the measurement: Do not move and do not talk as this may
alter the measurements.
•
•
•
•
T aking a blood pressure measurement
After the cuff has been appropriately positioned,the measurement can begin:
1.
2.
3.
4.
Switch the unit on by pressing the - button . The measurement will also
start immediately . i
The unit is ready for measurement and the number 0 shortly appears. It auto-
matically inflates the cuff slowly in order to measure your blood pressure.
The rising pressure in the cuff is shown on the display . i
The unit inflates the cuff until sufficient pressure has built up for a measure-
ment. Then the unit slowly releases air from the cuff and carries out the
measurement. When the device detects the signal, the heart symbol i
starts to flash on the display . i
When the measurement is finished, the cuff is deflated completely . The
systolic and diastolic blood pressure and the pulse value appear on the
display . The blood pressure indicator is shown next to the relevant
coloured bar depending on the WHO blood pressure classification. If the unit
has detected an irregular heartbeat, the arrhythmia display will also
be shown.
Symbol Cause Correction
E-1
E-2
E-3
E-4
E-5
E-6
-
Cuff not fitted
properly
Movement or
speaking during
measurement
Error during inflation
process
Measurement can
not be proceeded
or measurement
failed
T oo high cuff
pressure
Weak batteries
Device constantly
re-pumps the cuff
Put on the cuff correctly .
Repeat the measurement in the
correct way .
Repeat the measurement in the
correct way after relaxing 30 minutes.
Do not speak or move during the
measurement.
Put on the cuff correctly .
Repeat the measurement.
A weak pulse signal may be the cause
for measurement problems.
Repeat the measurement after relaxing
30 minutes.
If you get unusual readings for 3 times,
please contact your doctor .
Pressure above 300 mmHg. Contact
your doctor .
See chapter "Insert / replace batteries "
Remove the batteries and insert new
ones. Repeat the measurement.
Please contact the manufacturer resp. the service centre if you can’t solve the
problem. Do not disassemble the unit by yourself.
Cleaning and care
Remove the batteries before cleaning the unit. Clean the unit and the cuff using
a soft cloth lightly moistened with a mild soapy solution. Never use abrasive
cleaning agents, alcohol, naphta, thinner or gasoline etc.. Never immerse the
unit or any component in water . Be cautiious not to get any moisture in the main
unit. Do not use the unit again until it is completely dry . i
Only inflate the cuff when it is in position around the wrist. i
Do not expose the unit to direct sunlight; protect it against dirt and moisture. i
Do not subject the unit to extreme hot or cold temperature. Keep the unit in the
original packing when not in use. Store the unit in a clean and dry location.
Disposal
This product must not be disposed of together with domestic waste. All
users are obliged to hand in all electrical or electronic devices, regardless
of whether or not they contain toxic substances, at a municipal or commer-
cial collection point so that they can be disposed of in an environmentally
acceptable manner .
Remove the batteries before disposing of the device/unit. Do not dispose of old
batteries with your household waste, but at a battery collection station at a re-
cycling site or in a shop. Consult your local authority or your supplier for infor-
mation about disposal.
BW -82E
Name and model
Display system
Memory slots
Measuring method
Power supply
Blood pressure measuring range
Pulse measuring range
Maximum error tolerance for
static pressure
Maximum error tolerance for
pulse rate
Pressure generation
Deflation
Automatic switch-off
Operating conditions
Storage conditions
Dimensions (main unit)
Weight
Item number
EAN number
ecomed Blood pressure monitor BW-82E
Digital display
2 x 60 measurement values
Oscillometric
3 V=, 2 x 1.5V batteries AAA LR03
30 – 250 mmHg
40 – 180 beats/min.
± 3 mmHg or 2% of the value
± 5 % of the value
Automatic with pump
Automatic
After approx. 2 minutes
+10 °C to +40 °C,
15 to 85 % max. relative humidity
-20 °C to +60 °C,
10 to 85 % max. relative humidity
Approx. 70 x 80 x 32 mm (L x W x H)
Approx. 142 g without batteries
23212
40 15588 23212 1
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
T echnical specifications
In accordance with our policy of continual product improvement, we
reserve the right to make technical and visual changes without notice.
23212 04/2014