2 Applications
The METRISO 1000A insulation tester complies with DIN VDE 0413
“Regulations for Devices for the Testing of Protective Measures in Electri-
cal Systems”
Part 1: insulation testers, and
Part 4: ohmmeters.
It is suitable for the measurement of insulation resistance at voltage-free
devices and systems with nominal voltages of up to 1000 V, as well as for
testing resistance at earthing conductors, protective conductors and
bonding conductors including their connections and terminals.
The instrument has also been equipped with a 1000 V measuring range
for direct and alternating voltage. This is especially advantageous for test-
ing for the absence of voltage at devices under test, and for discharging
capacitive devices under test.
3 Initial Start-Up
3.1 Battery Installation
Attention!
!
Before opening the housing base make, certain that the function
selector switch has been set to the “1000 V” position, and that
the instrument has been fully disconnected from all external
electrical circuits.
Unscrew the housing base.
Insert 6 commercially available, type IEC R20 or IEC LR20 1.5 V
batteries (single cell) with plus and minus poles aligned in accordance
with the symbols in the battery compartment.
Replace the housing base and screw into place.
3.2 Testing the Batteries
Each time the measurement key is activated, the instrument automatically
tests the batteries. Momentary drain on the batteries resulting from the
function selector switch position and the pre-selected nominal voltage are
taken into consideration during this test. Battery condition is indicated by
the color of the indicator lamp at the left-hand portion of the scale area.
Green Lamp: Batteries are OK.
Red Lamp: The batteries are too weak for the currently selected
measurement function.
3.3 Switching the Tester On and Off
The tester remains activated as long as the measurement key in the test
probe is depressed. We recommend setting the function selector switch
to the 1000 V position during transport and maintenance, in order to
prevent the tester from being switched on inadvertently.
3.4 Analog Display
Measurement values are plainly displayed at analog scales. Each of the
function selector switch positions has its own scale. A battery status
indicator lamp is located at the left-hand portion of the scale area.
3.5 Fusing
The instrument is protected with a single common fuse during low-
impedance and insulation resistance measurements. The fuse blows if an
extraneous voltage is inadvertently applied when the instrument is set to
the resistance measuring range.
Voltage measurement is possible regardless of the condition of the fuse.
3.5.1 Checking the Fuse
The Ω and MΩ ranges are no longer functional after the fuse has blown.
Low-Impedance Measuring Range:
The needle remains at the mechanically set zero point when the
measurement key is activated.
Insulation Resistance Measuring Range:
With short-circuited test probes, the needle remains at the mechanically
set zero point when the measurement key is activated.
See chapter 8.2 on page 9, “Replacing the Fuse”.
4

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt METRISO 1000A wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info