5.3 Measurements with the Guard Cable (with Accessory Guard 5000A)
The measurement of very high impedance resistances presupposes
extremely small measurement currents, and can be rendered problematic
by influences such as electromagnetic fields, moisture or surface currents.
For this reason, the measurement set-up must be implemented in an
accurate fashion.
For measurements within a range of 100 G (10 G) ... 1 T a guard
cable must be used in order to prevent the distortion of measurement
results caused by surface currents. The guard rings prevent current at the
surface of the insulation material from flowing from the +measurement
cable to the measurement cable, instead of through the insulation mate-
rial itself.
Ð Insert the guard cable plug into the appropriate jack at the tester.
Ð Plug the crocodile clip onto the guard cable test prod.
Ð Clip the crocodile clip onto the guard ring located between the two
measuring points of the insulation material to be measured.
Ð See Chapter 5.1, page 6 for measurement procedures.
Note
The following materials can be used as guard rings:
aluminum foil, copper foil or metal tubing clamps.
6 Characteristic Values
Insulation Resistance
ShortCircuit Current I
K
1.3 mA
Direct and Alternating Voltage
Protective Devices
1)
with reference to scale length 97.5 mm (100 M range) or 109.8 mm (1 T range)
2)
PTC thermistor cool-down period until start of new measurement:
at least 2 minutes must be observed!
+Measurement Cable
– Meas. Cable
Guard Cable
Conductor
Insulation
Material
Guard Rings
Contact Ring
Measuring
Range
Nominal
Range of Use
Nominal/Test
Voltage
U
N
/ U
T
Nom./Test
Current
I
N
/ I
T
Intrinsic Un-
certainty
1)
Measuring Un-
certainty
100 k...
100 M
100 k...
10 M
100 V
250 V
500 V
1000 V
1mA
2.5%
30%
of measured
value
10 k... 1 T
100 k...
100 G
100/1500 V
250/
2000 V
500/
2500 V
1000/5000 V
1mA/0.7mA
1mA/0.5mA
1mA/0.4mA
1mA/0.1mA
5%
Measuring Range Frequency Internal
Resistance
Max. Allowable
Voltage
Intrinsic
Uncertainty
1)
0 ... 2000 V DC/AC 15 ... 500 Hz 5 M
2200 V DC/AC
max. 10 s
5%
Terminal Internal
Resistance
Max. Allowable
Voltage
Protective Device
–Measurement
Cable
to +meas. cable/
to guard cable:
2000 V DC/AC
max. 10 s
via grounded
damping diodes
+Measurement
Cable
Insulation
Measurement
to –meas. cable/
to guard cable:
2000 V DC/AC
max. 10 s
diodes in high-voltage
cascade, PTC thermistor
2)
and series resistors
Guard Cable
between guard and
meas. cables 90 k
to meas. cable
2000 V DC/AC
max. 10 s
PTC thermistor
2)
and
series resistors
Battery 10 V
pole protection
with diodes voltage limiting
in battery charger (optional)
8

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt METRISO PRIME wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gossen Metrawatt METRISO PRIME

Gossen Metrawatt METRISO PRIME Bedienungsanleitung - Deutsch - 14 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info