6 GMC-I Messtechnik GmbH
3.4 Analog Display
Logarithmic representation of the upper resistance scale allows for quick
recognition of the magnitude involved.
In order to achieve the required accuracy for testing of protective mea-
sures, the RANGE key allows for switching to the expanded lower limit
measuring range of 100 k ... 100 M.
The orange colored LEDs, (4) and (5), at the right end of the scale indicate
which of the two resistance measuring ranges is currently active.
The indicator LED lights up green to confirm correct insulation mea-
surement. If this LED does not light up, test voltage has not been
achieved. In this event a battery test is advisable.
The two lower scales are for voltage measurement and battery testing;
see Chapter 3.2.
4 Direct and Alternating Voltage Measurements
Direct voltages and sinusoidal alternating voltages with frequencies rang-
ing from 15 to 500 Hz can be measured with the tester. Deflection of the
needle at the instrument is always positive for direct voltage measure-
ments, regardless of polarity at the connections. Alternating voltage is
indicated as an effective value.
Voltage measurements are used to test for the absence of voltage prior to
insulation resistance measurements, as well as for the automatic dis-
charge of capacitive devices under test. The voltage drop can be
observed at the display.
Note
Voltage measurement can always be performed with the function
selector switch in the “V” position, regardless of the position of the
range selector switch (even without batteries).
Ð Set the function selector switch to the “V” position.
Ð Check to see that the needle points to “0” in the V scale when the test
prods are not connected. Reset the needle as required with the ad-
justor screw for mechanical zero adjustment.
Ð
The position of the range selector switch has no influence on voltage mea-
surements, although we recommend setting it to the OFF/V position.
Ð Contact the measuring points with both test prods.
Ð Read the measurement value at the V scale.
Note
Voltages of greater than 2000 V may not be applied. Input resis-
tance for the voltage measuring range is equal to 5 M
5 Insulation Resistance Measurement
Be certain that the device under test is voltage-free before performing
measurements; see Chapter 4.
For measurements within a range of 100 G (10 G) ... 1 T the guard
cable should be used; see Chapter 5.3, page 8.
5.1 Measurement Procedure
Ð Set the function selector switch to the position, in order to activate
test voltage.
Ð A range of either 10 k ... 1 T or 100 k ... 100 Mcan be pre-se-
lected with the RANGE key.
Ð Select the required nominal voltage of 100 V, 250 V, 500 V, 1000 V,
1500 V, 2000 V, 2500 V or 5000 V with the range selector switch de-
pending upon the nominal voltage of the DUT.
When lit, the LED (4) indicates that the upper limit measuring range of
10 k ... 1 T has been activated.
Ð Contact the measuring point with both test prods and wait until the
needle has come to rest. Depending upon the DUT this might take
only a few, or up to 30 seconds, if, for example, large capacities (long
cables) must be charged.
Ð Read the measurement value from the upper scale.
If the indicator LED (3) lights up green, the insulation measurement
is valid. If this LED does not light up, test voltage has not been
achieved. In this event a battery test is advisable; see Chapter 3.2,
page 5.
Ð In order to achieve the required accuracy for testing of protective
measures, select the high resolution range, 100 k ... 100 M, with
the help of the RANGE key. The LED (5) lights up.
Ð Contact the measuring point with both test prods once again.
Ð Read the measurement value from the lower scale.
6

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt METRISO PRIME wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gossen Metrawatt METRISO PRIME

Gossen Metrawatt METRISO PRIME Bedienungsanleitung - Deutsch - 14 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info