2 Safety Features and Precautions
This instrument fulfills the requirements of the applicable European and
national EC guidelines. We confirm this with the CE marking. The relevant
declaration of conformity can be obtained from GMC-I Messtechnik
GmbH.
The test instrument has been manufactured and tested in accordance
with the following safety regulations:
IEC 61010-1 / DIN EN 61010-1 / VDE 0411-1, DIN VDE 0404, DIN VDE
0413 Part 2 and 4 and DIN VDE 0104 (Feature F02 or
SECULIFE ST
HV
)
Only when used for its intended purpose, the safety of the user, the test
instrument and the device under test (electrical equipment or electrical
medical devices) is assured.
Read the operating instructions carefully and completely before placing your test
instrument into service, and follow all instructions contained therein. Make sure
that the operating instructions are available to all users of the instrument.
Tests may only be performed by a qualified electrician or under the super-
vision of a qualified electrician. The user must be instructed by a qualified
electrician concerning performance and evaluation of the test.
Note
+
Manufacturers and importers of electrical medical devices must
provide documentation for the performance of maintenance by
trained personnel.
Observe the following safety precautions:
The instrument may only be connected to electrical systems with
230 V/240 V
which conform to the applicable safety regulations (e.g.
IEC 60346, VDE 0100) and are protected with a fuse or circuit breaker
with a maximum rating of 16 A.
Measurements within electrical systems are prohibited.
Be prepared for the occurrence of unexpected voltages at devices
under test (for example, capacitors can be dangerously charged).
Make certain that the measurement cables are in flawless condition,
e.g. no damage to insulation, no interruptions in cables or plugs etc.
When using a probe with coil cable (SK2W):
Grip the test probe firmly, for example after insertion into a jack socket.
Tensioning at the coil cord may otherwise cause the test probe to snap
back resulting in possible injury.
Measurement of Insulation Resistance and Equivalent Leakage Current
This test is performed with a maximum voltage of 500 V, with a current
limit having been set (I < 3.5 mA). However, contacting the terminals
(3 or 2) causes an electric shock which, in turn, may result in accidents.
Leakage Current Measurement
During leakage current measurement it is imperative to ensure that the
device under test is operated at line voltage. Exposed conductive parts
may be charged with hazardous contact voltage during the test and
may consequently not be touched under any circumstances. (There is
a power shutdown as soon as the leakage current is higher than ap-
prox. 10 mA).
Attention!
!
The function test may only be performed after the DUT has
successfully passed the safety test!
Switching loads on and off
When switching the DUT on or off under load it is imperative that you ad-
here to the order indicated below. This helps to prevent excessive wear
and tear of the mains relays at the test instrument.
Before the measurement:
1) DUT: Switch the DUT off at the proprietary switch.
2) SECULIFE ST or SECULIFE ST
HV:
Apply line voltage to the test socket .
3) DUT: Switch on the DUT at the proprietary switch.
After the measurement:
4) DUT: Switch the DUT off at the proprietary switch.
5) SECULIFE ST or SECULIFE ST
HV:
Disconnect the test socket from the line .
The measuring and test instrument may not be used:
If it demonstrates visible damage
With damaged connector cables, measuring cables or patient ports
If it no longer functions properly
After extraordinary stresses due to transport
In such cases, the instrument must be removed from operation and
secured against unintentional use.
9

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info