590775
17
Zoom out
Zoom in
Vorherige Seite
1/72
Nächste Seite
GMC-I Messtechnik GmbH 17
7 Connecting the Device Under Test
õ Connect the DUT in accordance with the schematic diagrams included
in the online help function.
Connection of the DUT to the test instrument depends upon:
The type of device under test:
electrical equipment or not, with or without application parts
The type of connector included with the DUT:
– With plug (“to test socket” parameter), applies to EL1 adapter as well
– Without plug, single or multi-phase connection (“to jacks” parameter)
– No connection to tester (“permanent connection” parameter),
see also chapter 3.1
Whether or not an adapter is used:
– Adapter to socket (customer specific adapter)
– AT3-II S to socket, adapter for devices which are equipped with
5-pole, 16 A CEE plugs
– AT3-III E to socket, adapter for devices which are equipped with
5-pole, 32 A CEE plugs, see AT3-III operating instructions for test
sequence.
The DUT’s safety class (I, II or III)
Note
+
The DUT must be switched on for all tests. Switches, relays,
temperature regulators etc. must all be taken into consideration.
The test instrument automatically recognizes whether or not the DUT is
connected to jacks 1 through 3. The instrument also recognizes whether
or not the DUT has been connected to the test socket. As a default set-
ting, the program sequence assumes that the plug from the DUT has
been connected to the test socket.
Note
+
Safety Class II Devices with Safety Class I Mains Plugs
If the device under test is equipped with a safety class I plug
although it complies with safety class II, safety class I is recog-
nized by the test instrument. If this is the case, switch from safety
class I to safety class II in the initial menu.
If the test instrument is unable to automatically recognize how the DUT
has been connected, the recommended connection setup should be
double checked and determined manually if necessary.
õ Position the cursor at the third line in the start menu for the test se-
quence.
õ A selection of possible connection setups can be displayed by
activating the key.
õ Select the desired connection setup with the cursor and
acknowledge with the key.
For omitting the protective conducter test in the case of fully insulated devices see
page 65.
Protective Conductor and Insulation Resistance Measurements for Permanently
Installed Devices Under Test
Attention!
!
Deactivate the electrical system which supplies power to the
device under test before connecting the test instrument!
õ Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
Measuring Contact Current (absence of voltage)
Make sure that the contacted parts are not grounded.
High-Voltage Test (Feature F02 or SECULIFE ST
HV)
Attention!
!
The KS13 cable set, or similar cable sets, may not be used
for the high-voltage test. The high-voltage test may only be
performed directly via the test socket!
õ Connect the device under test to the test socket.
õ Safety class II only: Connect the probe to jacks 4 and 5.
Attention!
!
Make sure that the application parts are not connected during
the high-voltage test!
17

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info