9 Measuring Protective Conductor Resistance
Definition
Protective conductor resistance
is the sum of the following
resistances:
Connector cable or device
connector cable resistance
Contact resistance at plug
and terminal connections
Extension cable resistance if
utilized
Resistance is measured:
Between each conductive
part of the housing and the earthing contacts at the mains and the de-
vice plug (if a removable mains connector cable is used), or the protec-
tive conductor terminal for permanently installed devices.
as 4-pole measurement
Between the earthing contacts at the mains plug and the earthing
contacts at the device plug for device connector cables
Between the earthing contacts at the mains plug and the earthing
contacts at the coupling socket for extension cables
Connecting Safety Class I Devices to the Test Socket
When the DUT is connected, resistance is measured between the protec-
tive conductor terminal at the test socket or at the PE jack and the probe
connection at the DUT (contact with conductive parts of the housing).
õ In order to measure protective conductor resistance, contact a
conductive part of the housing with the probe, which is connected to
the protective conductor.
During measurement, the connector cable must only be moved in as far as
it is accessible during repair, modification or testing.
If a change in resistance occurs during the manual test step of the conti-
nuity test, it must be assumed that the protective conductor is damaged,
or that one of the connector contacts is no longer in flawless condition.
Testing Extension Cables
See test sequence in chapter 15.7 on page 46.
Note
+
“Connection of the DUT: SC I/II” is not displayed when the test is
performed individually, but rather only during the automatic test
sequence.
Selecting Test Current and Polarity
Test current amperage (200 mA DC or 10 A AC (Feature G00 or SECULIFE ST) or
200 mA DC, 25 A AC (Feature G01 or SECULIFE ST
HV)) as well as polarity can
both be changed by pressing the or the key.
Testing with 10 A Test Current (Feature G00 or SECULIFE ST)
or 25 A (Feature G01 or SECULIFE ST
HV)
The test has a maximum duration of 30 s (fixed value) if 10 A or 25 A test
current is used. After this time period has elapsed, the last measured
value is frozen and “data hold, measurement stopped” appears at the dis-
play. If the test instrument becomes excessively warm, testing cannot be
repeated until after a waiting period of 1 minute. When testing with 10 A or
25 A, the last measurement can be repeated if the test results in failure.
PE
Ω
19

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info